Effects of Calibration Approaches on the Accuracy for LC–MS Targeted Quantification of Therapeutic Protein
[Image: see text] LC–MS provides a promising alternative to ligand-binding assays for quantification of therapeutic proteins and biomarkers. As LC–MS methodology is based on the analysis of proteolytic peptides, calibration approaches utilizing various calibrators and internal standards (I.S.) have...
Autores principales: | , , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
American
Chemical
Society
2014
|
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3982980/ https://www.ncbi.nlm.nih.gov/pubmed/24611550 http://dx.doi.org/10.1021/ac5001477 |
_version_ | 1782311234913173504 |
---|---|
author | Nouri-Nigjeh, Eslam Zhang, Ming Ji, Tao Yu, Haoying An, Bo Duan, Xiaotao Balthasar, Joseph Johnson, Robert W. Qu, Jun |
author_facet | Nouri-Nigjeh, Eslam Zhang, Ming Ji, Tao Yu, Haoying An, Bo Duan, Xiaotao Balthasar, Joseph Johnson, Robert W. Qu, Jun |
author_sort | Nouri-Nigjeh, Eslam |
collection | PubMed |
description | [Image: see text] LC–MS provides a promising alternative to ligand-binding assays for quantification of therapeutic proteins and biomarkers. As LC–MS methodology is based on the analysis of proteolytic peptides, calibration approaches utilizing various calibrators and internal standards (I.S.) have been developed. A comprehensive assessment of the accuracy and reliability of these approaches is essential but has yet been reported. Here we performed a well-controlled and systematic comparative study using quantification of monoclonal-antibody in plasma as the model system. Method development utilized a high-throughput orthogonal-array-optimization, and two sensitive and stable signature-peptides (SP) from different domains were selected based on extensive evaluations in plasma matrix. With the purities of all protein/peptide standards corrected by quantitative amino acid analysis (AAA), five calibration approaches using stable-isotope-labeled (SIL) I.S. were thoroughly compared, including those at peptide, extended-peptide, and protein levels and two “hybrid” approaches (i.e., protein calibrator with SIL-peptide or SIL-extended-peptide I.S.). These approaches were further evaluated in parallel for a 15 time point, preclinical pharmacokinetic study. All methods showed good precision (CV% < 20%). When examined with protein-spiked plasma QC, peptide-level calibration exhibited severe negative biases (−23 to −62%), highly discordant results between the two SP (deviations of 38–56%), and misleading pharmacokinetics assessments. Extended-peptide calibration showed significant improvements but still with unacceptable accuracy. Conversely, protein-level and the two hybrid calibrations achieved good quantitative accuracy (error < 10%), concordant results by two SP (deviations < 15%), and correct pharmacokinetic parameters. Hybrid approaches were found to provide a cost-effective means for accurate quantification without the costly SIL-protein. Other key findings include (i) using two SP provides a versatile gauge for method reliability; (ii) evaluation of peptide stability in the matrix before SP selection is critical; and (iii) using AAA to verify purities of protein/peptide calibrators ensures accurate quantitation. These results address fundamental calibration issues that have not been adequately investigated in published studies and will provide valuable guidelines for the “fit for purpose” development of accurate LC–MS assays for therapeutic proteins and biomarkers in biological matrices. |
format | Online Article Text |
id | pubmed-3982980 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2014 |
publisher | American
Chemical
Society |
record_format | MEDLINE/PubMed |
spelling | pubmed-39829802015-03-10 Effects of Calibration Approaches on the Accuracy for LC–MS Targeted Quantification of Therapeutic Protein Nouri-Nigjeh, Eslam Zhang, Ming Ji, Tao Yu, Haoying An, Bo Duan, Xiaotao Balthasar, Joseph Johnson, Robert W. Qu, Jun Anal Chem [Image: see text] LC–MS provides a promising alternative to ligand-binding assays for quantification of therapeutic proteins and biomarkers. As LC–MS methodology is based on the analysis of proteolytic peptides, calibration approaches utilizing various calibrators and internal standards (I.S.) have been developed. A comprehensive assessment of the accuracy and reliability of these approaches is essential but has yet been reported. Here we performed a well-controlled and systematic comparative study using quantification of monoclonal-antibody in plasma as the model system. Method development utilized a high-throughput orthogonal-array-optimization, and two sensitive and stable signature-peptides (SP) from different domains were selected based on extensive evaluations in plasma matrix. With the purities of all protein/peptide standards corrected by quantitative amino acid analysis (AAA), five calibration approaches using stable-isotope-labeled (SIL) I.S. were thoroughly compared, including those at peptide, extended-peptide, and protein levels and two “hybrid” approaches (i.e., protein calibrator with SIL-peptide or SIL-extended-peptide I.S.). These approaches were further evaluated in parallel for a 15 time point, preclinical pharmacokinetic study. All methods showed good precision (CV% < 20%). When examined with protein-spiked plasma QC, peptide-level calibration exhibited severe negative biases (−23 to −62%), highly discordant results between the two SP (deviations of 38–56%), and misleading pharmacokinetics assessments. Extended-peptide calibration showed significant improvements but still with unacceptable accuracy. Conversely, protein-level and the two hybrid calibrations achieved good quantitative accuracy (error < 10%), concordant results by two SP (deviations < 15%), and correct pharmacokinetic parameters. Hybrid approaches were found to provide a cost-effective means for accurate quantification without the costly SIL-protein. Other key findings include (i) using two SP provides a versatile gauge for method reliability; (ii) evaluation of peptide stability in the matrix before SP selection is critical; and (iii) using AAA to verify purities of protein/peptide calibrators ensures accurate quantitation. These results address fundamental calibration issues that have not been adequately investigated in published studies and will provide valuable guidelines for the “fit for purpose” development of accurate LC–MS assays for therapeutic proteins and biomarkers in biological matrices. American Chemical Society 2014-03-10 2014-04-01 /pmc/articles/PMC3982980/ /pubmed/24611550 http://dx.doi.org/10.1021/ac5001477 Text en Copyright © 2014 American Chemical Society |
spellingShingle | Nouri-Nigjeh, Eslam Zhang, Ming Ji, Tao Yu, Haoying An, Bo Duan, Xiaotao Balthasar, Joseph Johnson, Robert W. Qu, Jun Effects of Calibration Approaches on the Accuracy for LC–MS Targeted Quantification of Therapeutic Protein |
title | Effects of Calibration Approaches on the Accuracy
for LC–MS Targeted Quantification of Therapeutic Protein |
title_full | Effects of Calibration Approaches on the Accuracy
for LC–MS Targeted Quantification of Therapeutic Protein |
title_fullStr | Effects of Calibration Approaches on the Accuracy
for LC–MS Targeted Quantification of Therapeutic Protein |
title_full_unstemmed | Effects of Calibration Approaches on the Accuracy
for LC–MS Targeted Quantification of Therapeutic Protein |
title_short | Effects of Calibration Approaches on the Accuracy
for LC–MS Targeted Quantification of Therapeutic Protein |
title_sort | effects of calibration approaches on the accuracy
for lc–ms targeted quantification of therapeutic protein |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3982980/ https://www.ncbi.nlm.nih.gov/pubmed/24611550 http://dx.doi.org/10.1021/ac5001477 |
work_keys_str_mv | AT nourinigjeheslam effectsofcalibrationapproachesontheaccuracyforlcmstargetedquantificationoftherapeuticprotein AT zhangming effectsofcalibrationapproachesontheaccuracyforlcmstargetedquantificationoftherapeuticprotein AT jitao effectsofcalibrationapproachesontheaccuracyforlcmstargetedquantificationoftherapeuticprotein AT yuhaoying effectsofcalibrationapproachesontheaccuracyforlcmstargetedquantificationoftherapeuticprotein AT anbo effectsofcalibrationapproachesontheaccuracyforlcmstargetedquantificationoftherapeuticprotein AT duanxiaotao effectsofcalibrationapproachesontheaccuracyforlcmstargetedquantificationoftherapeuticprotein AT balthasarjoseph effectsofcalibrationapproachesontheaccuracyforlcmstargetedquantificationoftherapeuticprotein AT johnsonrobertw effectsofcalibrationapproachesontheaccuracyforlcmstargetedquantificationoftherapeuticprotein AT qujun effectsofcalibrationapproachesontheaccuracyforlcmstargetedquantificationoftherapeuticprotein |