Validation of the integration of technology that measures additional “vascular” indices into an ambulatory blood pressure monitoring system
BACKGROUND: The objective of this study was to validate the novel integration of oscillometric (Vasotens(®)) technology into a BPLab(®) ambulatory blood pressure (BP) monitoring system to measure central BP, the aortic augmentation index, and pulse wave velocity (PWV) compared with the recommended a...
Autores principales: | , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Dove Medical Press
2014
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4014382/ https://www.ncbi.nlm.nih.gov/pubmed/24833924 http://dx.doi.org/10.2147/MDER.S61839 |
Sumario: | BACKGROUND: The objective of this study was to validate the novel integration of oscillometric (Vasotens(®)) technology into a BPLab(®) ambulatory blood pressure (BP) monitoring system to measure central BP, the aortic augmentation index, and pulse wave velocity (PWV) compared with the recommended and widely accepted tonometric method. METHODS: The ARTERY Society guidelines for comparison of PWV measurement techniques were used as the basis for recruitment of 99 individuals (mean age 44±19 years, 52 males). The standard for comparison was the conventional “classic” SphygmoCor device. RESULTS: Accordance of the two methods was satisfactory (r=0.98, mean difference of 2.9±3.5 mmHg for central systolic BP; r=0.98, mean difference of −1.1±2.3 mmHg for central diastolic BP; r=0.83, mean difference of −2.6%±13% for aortic augmentation index; r=0.85, mean difference of 0.69±1.4 for PWV). CONCLUSION: The performance of Vasotens algorithms using an oscillometric ambulatory BP monitoring system is feasible for accurate diagnosis, risk assessment, and evaluation of the effects of antihypertensive drugs. |
---|