Repeatability of Ocular Measurements with a Dual-Scheimpflug Analyzer in Healthy Eyes

Purpose. To assess the repeatability of the Galilei dual Scheimpflug analyzer (GDSA) in anterior segment examination. Methods. Fifty-two eyes from 52 healthy volunteers were prospectively and consecutively recruited. Anatomic, axial, refractive, and instantaneous parameters were measured with GDSA t...

Descripción completa

Detalles Bibliográficos
Autores principales: Lopez de la Fuente, Carmen, Sanchez-Cano, Ana, Segura, Francisco, Fuentes-Broto, Lorena, Pinilla, Isabel
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Hindawi Publishing Corporation 2014
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4017806/
https://www.ncbi.nlm.nih.gov/pubmed/24868548
http://dx.doi.org/10.1155/2014/808646
_version_ 1782480008859615232
author Lopez de la Fuente, Carmen
Sanchez-Cano, Ana
Segura, Francisco
Fuentes-Broto, Lorena
Pinilla, Isabel
author_facet Lopez de la Fuente, Carmen
Sanchez-Cano, Ana
Segura, Francisco
Fuentes-Broto, Lorena
Pinilla, Isabel
author_sort Lopez de la Fuente, Carmen
collection PubMed
description Purpose. To assess the repeatability of the Galilei dual Scheimpflug analyzer (GDSA) in anterior segment examination. Methods. Fifty-two eyes from 52 healthy volunteers were prospectively and consecutively recruited. Anatomic, axial, refractive, and instantaneous parameters were measured with GDSA to provide a complete characterization of the anterior segment. Repeatability was assessed calculating intraclass correlation coefficient (ICC), and coefficient of variation (COV). Results. Correlation among repeated measurements showed almost perfect reliability (ICC > 0.81) for all parameters except thinnest central corneal thickness (CCT) (0.78), corneal thickness average out (0.79), and posterior axial curvature average out (0.60). Repeatability was excellent (COV < 10%) for all parameters except anterior chamber volume and, superior iridocorneal angle and eccentricities. In these last three parameters, repeatability limits were excessively high compared to the mean. Conclusions. GDSA in healthy young persons had an almost perfect correlation in measuring anatomic, axial, instantaneous, and refractive parameters with greater variability for peripheral terms. Repeatability of anatomical parameters like pachymetry, anterior chamber, or iridocorneal angle and eccentricity were limited. In healthy young persons, the other evaluated parameters had very good repeatability and their limits of agreement showed excellent clinical results for this device.
format Online
Article
Text
id pubmed-4017806
institution National Center for Biotechnology Information
language English
publishDate 2014
publisher Hindawi Publishing Corporation
record_format MEDLINE/PubMed
spelling pubmed-40178062014-05-27 Repeatability of Ocular Measurements with a Dual-Scheimpflug Analyzer in Healthy Eyes Lopez de la Fuente, Carmen Sanchez-Cano, Ana Segura, Francisco Fuentes-Broto, Lorena Pinilla, Isabel Biomed Res Int Research Article Purpose. To assess the repeatability of the Galilei dual Scheimpflug analyzer (GDSA) in anterior segment examination. Methods. Fifty-two eyes from 52 healthy volunteers were prospectively and consecutively recruited. Anatomic, axial, refractive, and instantaneous parameters were measured with GDSA to provide a complete characterization of the anterior segment. Repeatability was assessed calculating intraclass correlation coefficient (ICC), and coefficient of variation (COV). Results. Correlation among repeated measurements showed almost perfect reliability (ICC > 0.81) for all parameters except thinnest central corneal thickness (CCT) (0.78), corneal thickness average out (0.79), and posterior axial curvature average out (0.60). Repeatability was excellent (COV < 10%) for all parameters except anterior chamber volume and, superior iridocorneal angle and eccentricities. In these last three parameters, repeatability limits were excessively high compared to the mean. Conclusions. GDSA in healthy young persons had an almost perfect correlation in measuring anatomic, axial, instantaneous, and refractive parameters with greater variability for peripheral terms. Repeatability of anatomical parameters like pachymetry, anterior chamber, or iridocorneal angle and eccentricity were limited. In healthy young persons, the other evaluated parameters had very good repeatability and their limits of agreement showed excellent clinical results for this device. Hindawi Publishing Corporation 2014 2014-04-24 /pmc/articles/PMC4017806/ /pubmed/24868548 http://dx.doi.org/10.1155/2014/808646 Text en Copyright © 2014 Carmen Lopez de la Fuente et al. https://creativecommons.org/licenses/by/3.0/ This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Research Article
Lopez de la Fuente, Carmen
Sanchez-Cano, Ana
Segura, Francisco
Fuentes-Broto, Lorena
Pinilla, Isabel
Repeatability of Ocular Measurements with a Dual-Scheimpflug Analyzer in Healthy Eyes
title Repeatability of Ocular Measurements with a Dual-Scheimpflug Analyzer in Healthy Eyes
title_full Repeatability of Ocular Measurements with a Dual-Scheimpflug Analyzer in Healthy Eyes
title_fullStr Repeatability of Ocular Measurements with a Dual-Scheimpflug Analyzer in Healthy Eyes
title_full_unstemmed Repeatability of Ocular Measurements with a Dual-Scheimpflug Analyzer in Healthy Eyes
title_short Repeatability of Ocular Measurements with a Dual-Scheimpflug Analyzer in Healthy Eyes
title_sort repeatability of ocular measurements with a dual-scheimpflug analyzer in healthy eyes
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4017806/
https://www.ncbi.nlm.nih.gov/pubmed/24868548
http://dx.doi.org/10.1155/2014/808646
work_keys_str_mv AT lopezdelafuentecarmen repeatabilityofocularmeasurementswithadualscheimpfluganalyzerinhealthyeyes
AT sanchezcanoana repeatabilityofocularmeasurementswithadualscheimpfluganalyzerinhealthyeyes
AT segurafrancisco repeatabilityofocularmeasurementswithadualscheimpfluganalyzerinhealthyeyes
AT fuentesbrotolorena repeatabilityofocularmeasurementswithadualscheimpfluganalyzerinhealthyeyes
AT pinillaisabel repeatabilityofocularmeasurementswithadualscheimpfluganalyzerinhealthyeyes