Repeatability of Ocular Measurements with a Dual-Scheimpflug Analyzer in Healthy Eyes
Purpose. To assess the repeatability of the Galilei dual Scheimpflug analyzer (GDSA) in anterior segment examination. Methods. Fifty-two eyes from 52 healthy volunteers were prospectively and consecutively recruited. Anatomic, axial, refractive, and instantaneous parameters were measured with GDSA t...
Autores principales: | , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Hindawi Publishing Corporation
2014
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4017806/ https://www.ncbi.nlm.nih.gov/pubmed/24868548 http://dx.doi.org/10.1155/2014/808646 |
_version_ | 1782480008859615232 |
---|---|
author | Lopez de la Fuente, Carmen Sanchez-Cano, Ana Segura, Francisco Fuentes-Broto, Lorena Pinilla, Isabel |
author_facet | Lopez de la Fuente, Carmen Sanchez-Cano, Ana Segura, Francisco Fuentes-Broto, Lorena Pinilla, Isabel |
author_sort | Lopez de la Fuente, Carmen |
collection | PubMed |
description | Purpose. To assess the repeatability of the Galilei dual Scheimpflug analyzer (GDSA) in anterior segment examination. Methods. Fifty-two eyes from 52 healthy volunteers were prospectively and consecutively recruited. Anatomic, axial, refractive, and instantaneous parameters were measured with GDSA to provide a complete characterization of the anterior segment. Repeatability was assessed calculating intraclass correlation coefficient (ICC), and coefficient of variation (COV). Results. Correlation among repeated measurements showed almost perfect reliability (ICC > 0.81) for all parameters except thinnest central corneal thickness (CCT) (0.78), corneal thickness average out (0.79), and posterior axial curvature average out (0.60). Repeatability was excellent (COV < 10%) for all parameters except anterior chamber volume and, superior iridocorneal angle and eccentricities. In these last three parameters, repeatability limits were excessively high compared to the mean. Conclusions. GDSA in healthy young persons had an almost perfect correlation in measuring anatomic, axial, instantaneous, and refractive parameters with greater variability for peripheral terms. Repeatability of anatomical parameters like pachymetry, anterior chamber, or iridocorneal angle and eccentricity were limited. In healthy young persons, the other evaluated parameters had very good repeatability and their limits of agreement showed excellent clinical results for this device. |
format | Online Article Text |
id | pubmed-4017806 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2014 |
publisher | Hindawi Publishing Corporation |
record_format | MEDLINE/PubMed |
spelling | pubmed-40178062014-05-27 Repeatability of Ocular Measurements with a Dual-Scheimpflug Analyzer in Healthy Eyes Lopez de la Fuente, Carmen Sanchez-Cano, Ana Segura, Francisco Fuentes-Broto, Lorena Pinilla, Isabel Biomed Res Int Research Article Purpose. To assess the repeatability of the Galilei dual Scheimpflug analyzer (GDSA) in anterior segment examination. Methods. Fifty-two eyes from 52 healthy volunteers were prospectively and consecutively recruited. Anatomic, axial, refractive, and instantaneous parameters were measured with GDSA to provide a complete characterization of the anterior segment. Repeatability was assessed calculating intraclass correlation coefficient (ICC), and coefficient of variation (COV). Results. Correlation among repeated measurements showed almost perfect reliability (ICC > 0.81) for all parameters except thinnest central corneal thickness (CCT) (0.78), corneal thickness average out (0.79), and posterior axial curvature average out (0.60). Repeatability was excellent (COV < 10%) for all parameters except anterior chamber volume and, superior iridocorneal angle and eccentricities. In these last three parameters, repeatability limits were excessively high compared to the mean. Conclusions. GDSA in healthy young persons had an almost perfect correlation in measuring anatomic, axial, instantaneous, and refractive parameters with greater variability for peripheral terms. Repeatability of anatomical parameters like pachymetry, anterior chamber, or iridocorneal angle and eccentricity were limited. In healthy young persons, the other evaluated parameters had very good repeatability and their limits of agreement showed excellent clinical results for this device. Hindawi Publishing Corporation 2014 2014-04-24 /pmc/articles/PMC4017806/ /pubmed/24868548 http://dx.doi.org/10.1155/2014/808646 Text en Copyright © 2014 Carmen Lopez de la Fuente et al. https://creativecommons.org/licenses/by/3.0/ This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Research Article Lopez de la Fuente, Carmen Sanchez-Cano, Ana Segura, Francisco Fuentes-Broto, Lorena Pinilla, Isabel Repeatability of Ocular Measurements with a Dual-Scheimpflug Analyzer in Healthy Eyes |
title | Repeatability of Ocular Measurements with a Dual-Scheimpflug Analyzer in Healthy Eyes |
title_full | Repeatability of Ocular Measurements with a Dual-Scheimpflug Analyzer in Healthy Eyes |
title_fullStr | Repeatability of Ocular Measurements with a Dual-Scheimpflug Analyzer in Healthy Eyes |
title_full_unstemmed | Repeatability of Ocular Measurements with a Dual-Scheimpflug Analyzer in Healthy Eyes |
title_short | Repeatability of Ocular Measurements with a Dual-Scheimpflug Analyzer in Healthy Eyes |
title_sort | repeatability of ocular measurements with a dual-scheimpflug analyzer in healthy eyes |
topic | Research Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4017806/ https://www.ncbi.nlm.nih.gov/pubmed/24868548 http://dx.doi.org/10.1155/2014/808646 |
work_keys_str_mv | AT lopezdelafuentecarmen repeatabilityofocularmeasurementswithadualscheimpfluganalyzerinhealthyeyes AT sanchezcanoana repeatabilityofocularmeasurementswithadualscheimpfluganalyzerinhealthyeyes AT segurafrancisco repeatabilityofocularmeasurementswithadualscheimpfluganalyzerinhealthyeyes AT fuentesbrotolorena repeatabilityofocularmeasurementswithadualscheimpfluganalyzerinhealthyeyes AT pinillaisabel repeatabilityofocularmeasurementswithadualscheimpfluganalyzerinhealthyeyes |