Central and Midperipheral Corneal Thickness Measured with Scheimpflug Imaging and Optical Coherence Tomography
PURPOSE: To compare corneal thickness measurements using Pentacam (Oculus, Germany), Sirius (CSO, Italy), Galilei (Ziemer, Switzerland), and RTVue-100 OCT (Optovue Inc., USA). METHODS: Sixty-six eyes of 66 healthy volunteers were enrolled. Three consecutive measurements were performed with each devi...
Autores principales: | , , , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Public Library of Science
2014
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4031212/ https://www.ncbi.nlm.nih.gov/pubmed/24854348 http://dx.doi.org/10.1371/journal.pone.0098316 |
_version_ | 1782317503051988992 |
---|---|
author | Huang, Jinhai Ding, Xixia Savini, Giacomo Jiang, Zhengxuan Pan, Chao Hua, Yanjun Wu, Fan Feng, Yifan Yu, Ye Wang, Qinmei |
author_facet | Huang, Jinhai Ding, Xixia Savini, Giacomo Jiang, Zhengxuan Pan, Chao Hua, Yanjun Wu, Fan Feng, Yifan Yu, Ye Wang, Qinmei |
author_sort | Huang, Jinhai |
collection | PubMed |
description | PURPOSE: To compare corneal thickness measurements using Pentacam (Oculus, Germany), Sirius (CSO, Italy), Galilei (Ziemer, Switzerland), and RTVue-100 OCT (Optovue Inc., USA). METHODS: Sixty-six eyes of 66 healthy volunteers were enrolled. Three consecutive measurements were performed with each device. The mean value of the three measurements was used for subsequent analysis. Central corneal thickness (CCT), thinnest corneal thickness (TCT), and midperipheral corneal thickness (MPCT; measured at superior, inferior, nasal, and temporal locations with a distance of 1 mm (CT(2mm)) or 2.5 mm (CT(5mm)) from the corneal apex) were analyzed. Differences and agreement between measurements were assessed using the repeated-measures analysis of variance (ANOVA) and Bland-Altman analyses, respectively. RESULTS: Statistically significant differences (p<0.001) among the four devices were revealed in CCT, TCT and CT(2mm)measurements. The CCT, TCT, and CT(2mm) values were ranked from the thickest to the thinnest as follows: Galilei>Sirius>Pentacam>RTVue OCT. For these measurements, agreement between measurements by Sirius and Pentacam was good, whereas Galilei overestimated and RTVue underestimated corneal thickness compared to Sirius and Pentacam. As regards CT(5mm) measurements, Pentacam provided the largest values, whereas RTVue OCT yielded the smallest values. Agreement of the CT(5mm) measurements was good between the Pentacam, Sirius, moderate between Galilei and the other two Scheimpflug systems, and poor between the RTVue OCT and the remaining devices. CONCLUSIONS: The Pentacam and Sirius can be used interchangeably for CCT measurements, while the Galilei and RTVue systematically over- and underestimate CCT, respectively. The three Scheimpflug cameras, but not the RTVue, may be used interchangeably for MPCT measurements. |
format | Online Article Text |
id | pubmed-4031212 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2014 |
publisher | Public Library of Science |
record_format | MEDLINE/PubMed |
spelling | pubmed-40312122014-05-28 Central and Midperipheral Corneal Thickness Measured with Scheimpflug Imaging and Optical Coherence Tomography Huang, Jinhai Ding, Xixia Savini, Giacomo Jiang, Zhengxuan Pan, Chao Hua, Yanjun Wu, Fan Feng, Yifan Yu, Ye Wang, Qinmei PLoS One Research Article PURPOSE: To compare corneal thickness measurements using Pentacam (Oculus, Germany), Sirius (CSO, Italy), Galilei (Ziemer, Switzerland), and RTVue-100 OCT (Optovue Inc., USA). METHODS: Sixty-six eyes of 66 healthy volunteers were enrolled. Three consecutive measurements were performed with each device. The mean value of the three measurements was used for subsequent analysis. Central corneal thickness (CCT), thinnest corneal thickness (TCT), and midperipheral corneal thickness (MPCT; measured at superior, inferior, nasal, and temporal locations with a distance of 1 mm (CT(2mm)) or 2.5 mm (CT(5mm)) from the corneal apex) were analyzed. Differences and agreement between measurements were assessed using the repeated-measures analysis of variance (ANOVA) and Bland-Altman analyses, respectively. RESULTS: Statistically significant differences (p<0.001) among the four devices were revealed in CCT, TCT and CT(2mm)measurements. The CCT, TCT, and CT(2mm) values were ranked from the thickest to the thinnest as follows: Galilei>Sirius>Pentacam>RTVue OCT. For these measurements, agreement between measurements by Sirius and Pentacam was good, whereas Galilei overestimated and RTVue underestimated corneal thickness compared to Sirius and Pentacam. As regards CT(5mm) measurements, Pentacam provided the largest values, whereas RTVue OCT yielded the smallest values. Agreement of the CT(5mm) measurements was good between the Pentacam, Sirius, moderate between Galilei and the other two Scheimpflug systems, and poor between the RTVue OCT and the remaining devices. CONCLUSIONS: The Pentacam and Sirius can be used interchangeably for CCT measurements, while the Galilei and RTVue systematically over- and underestimate CCT, respectively. The three Scheimpflug cameras, but not the RTVue, may be used interchangeably for MPCT measurements. Public Library of Science 2014-05-22 /pmc/articles/PMC4031212/ /pubmed/24854348 http://dx.doi.org/10.1371/journal.pone.0098316 Text en © 2014 Huang et al http://creativecommons.org/licenses/by/4.0/ This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are properly credited. |
spellingShingle | Research Article Huang, Jinhai Ding, Xixia Savini, Giacomo Jiang, Zhengxuan Pan, Chao Hua, Yanjun Wu, Fan Feng, Yifan Yu, Ye Wang, Qinmei Central and Midperipheral Corneal Thickness Measured with Scheimpflug Imaging and Optical Coherence Tomography |
title | Central and Midperipheral Corneal Thickness Measured with Scheimpflug Imaging and Optical Coherence Tomography |
title_full | Central and Midperipheral Corneal Thickness Measured with Scheimpflug Imaging and Optical Coherence Tomography |
title_fullStr | Central and Midperipheral Corneal Thickness Measured with Scheimpflug Imaging and Optical Coherence Tomography |
title_full_unstemmed | Central and Midperipheral Corneal Thickness Measured with Scheimpflug Imaging and Optical Coherence Tomography |
title_short | Central and Midperipheral Corneal Thickness Measured with Scheimpflug Imaging and Optical Coherence Tomography |
title_sort | central and midperipheral corneal thickness measured with scheimpflug imaging and optical coherence tomography |
topic | Research Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4031212/ https://www.ncbi.nlm.nih.gov/pubmed/24854348 http://dx.doi.org/10.1371/journal.pone.0098316 |
work_keys_str_mv | AT huangjinhai centralandmidperipheralcornealthicknessmeasuredwithscheimpflugimagingandopticalcoherencetomography AT dingxixia centralandmidperipheralcornealthicknessmeasuredwithscheimpflugimagingandopticalcoherencetomography AT savinigiacomo centralandmidperipheralcornealthicknessmeasuredwithscheimpflugimagingandopticalcoherencetomography AT jiangzhengxuan centralandmidperipheralcornealthicknessmeasuredwithscheimpflugimagingandopticalcoherencetomography AT panchao centralandmidperipheralcornealthicknessmeasuredwithscheimpflugimagingandopticalcoherencetomography AT huayanjun centralandmidperipheralcornealthicknessmeasuredwithscheimpflugimagingandopticalcoherencetomography AT wufan centralandmidperipheralcornealthicknessmeasuredwithscheimpflugimagingandopticalcoherencetomography AT fengyifan centralandmidperipheralcornealthicknessmeasuredwithscheimpflugimagingandopticalcoherencetomography AT yuye centralandmidperipheralcornealthicknessmeasuredwithscheimpflugimagingandopticalcoherencetomography AT wangqinmei centralandmidperipheralcornealthicknessmeasuredwithscheimpflugimagingandopticalcoherencetomography |