Left and Right Ventricle Leads Switch as a Solution for T Wave Oversensing - How a Good Idea Turned Out Bad
A 50-year-old male with a CRT defibrillator received inappropriate ICD shocks due to T-wave oversensing. Decreasing the sensitivity to avoid T wave oversensing was not an option due to a suboptimal R-wave sensing amplitude. We decided to re-plug the LV lead in the RV port and the RV lead in the LV p...