Enhanced Photoacoustic Gas Analyser Response Time and Impact on Accuracy at Fast Ventilation Rates during Multiple Breath Washout

BACKGROUND: The Innocor device contains a highly sensitive photoacoustic gas analyser that has been used to perform multiple breath washout (MBW) measurements using very low concentrations of the tracer gas SF(6). Use in smaller subjects has been restricted by the requirement for a gas analyser resp...

Descripción completa

Detalles Bibliográficos
Autores principales: Horsley, Alex, Macleod, Kenneth, Gupta, Ruchi, Goddard, Nick, Bell, Nicholas
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Public Library of Science 2014
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4043749/
https://www.ncbi.nlm.nih.gov/pubmed/24892522
http://dx.doi.org/10.1371/journal.pone.0098487
_version_ 1782318981227479040
author Horsley, Alex
Macleod, Kenneth
Gupta, Ruchi
Goddard, Nick
Bell, Nicholas
author_facet Horsley, Alex
Macleod, Kenneth
Gupta, Ruchi
Goddard, Nick
Bell, Nicholas
author_sort Horsley, Alex
collection PubMed
description BACKGROUND: The Innocor device contains a highly sensitive photoacoustic gas analyser that has been used to perform multiple breath washout (MBW) measurements using very low concentrations of the tracer gas SF(6). Use in smaller subjects has been restricted by the requirement for a gas analyser response time of <100 ms, in order to ensure accurate estimation of lung volumes at rapid ventilation rates. METHODS: A series of previously reported and novel enhancements were made to the gas analyser to produce a clinically practical system with a reduced response time. An enhanced lung model system, capable of delivering highly accurate ventilation rates and volumes, was used to assess in vitro accuracy of functional residual capacity (FRC) volume calculation and the effects of flow and gas signal alignment on this. RESULTS: 10–90% rise time was reduced from 154 to 88 ms. In an adult/child lung model, accuracy of volume calculation was −0.9 to 2.9% for all measurements, including those with ventilation rate of 30/min and FRC of 0.5 L; for the un-enhanced system, accuracy deteriorated at higher ventilation rates and smaller FRC. In a separate smaller lung model (ventilation rate 60/min, FRC 250 ml, tidal volume 100 ml), mean accuracy of FRC measurement for the enhanced system was minus 0.95% (range −3.8 to 2.0%). Error sensitivity to flow and gas signal alignment was increased by ventilation rate, smaller FRC and slower analyser response time. CONCLUSION: The Innocor analyser can be enhanced to reliably generate highly accurate FRC measurements down at volumes as low as those simulating infant lung settings. Signal alignment is a critical factor. With these enhancements, the Innocor analyser exceeds key technical component recommendations for MBW apparatus.
format Online
Article
Text
id pubmed-4043749
institution National Center for Biotechnology Information
language English
publishDate 2014
publisher Public Library of Science
record_format MEDLINE/PubMed
spelling pubmed-40437492014-06-09 Enhanced Photoacoustic Gas Analyser Response Time and Impact on Accuracy at Fast Ventilation Rates during Multiple Breath Washout Horsley, Alex Macleod, Kenneth Gupta, Ruchi Goddard, Nick Bell, Nicholas PLoS One Research Article BACKGROUND: The Innocor device contains a highly sensitive photoacoustic gas analyser that has been used to perform multiple breath washout (MBW) measurements using very low concentrations of the tracer gas SF(6). Use in smaller subjects has been restricted by the requirement for a gas analyser response time of <100 ms, in order to ensure accurate estimation of lung volumes at rapid ventilation rates. METHODS: A series of previously reported and novel enhancements were made to the gas analyser to produce a clinically practical system with a reduced response time. An enhanced lung model system, capable of delivering highly accurate ventilation rates and volumes, was used to assess in vitro accuracy of functional residual capacity (FRC) volume calculation and the effects of flow and gas signal alignment on this. RESULTS: 10–90% rise time was reduced from 154 to 88 ms. In an adult/child lung model, accuracy of volume calculation was −0.9 to 2.9% for all measurements, including those with ventilation rate of 30/min and FRC of 0.5 L; for the un-enhanced system, accuracy deteriorated at higher ventilation rates and smaller FRC. In a separate smaller lung model (ventilation rate 60/min, FRC 250 ml, tidal volume 100 ml), mean accuracy of FRC measurement for the enhanced system was minus 0.95% (range −3.8 to 2.0%). Error sensitivity to flow and gas signal alignment was increased by ventilation rate, smaller FRC and slower analyser response time. CONCLUSION: The Innocor analyser can be enhanced to reliably generate highly accurate FRC measurements down at volumes as low as those simulating infant lung settings. Signal alignment is a critical factor. With these enhancements, the Innocor analyser exceeds key technical component recommendations for MBW apparatus. Public Library of Science 2014-06-03 /pmc/articles/PMC4043749/ /pubmed/24892522 http://dx.doi.org/10.1371/journal.pone.0098487 Text en © 2014 Horsley et al http://creativecommons.org/licenses/by/4.0/ This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are properly credited.
spellingShingle Research Article
Horsley, Alex
Macleod, Kenneth
Gupta, Ruchi
Goddard, Nick
Bell, Nicholas
Enhanced Photoacoustic Gas Analyser Response Time and Impact on Accuracy at Fast Ventilation Rates during Multiple Breath Washout
title Enhanced Photoacoustic Gas Analyser Response Time and Impact on Accuracy at Fast Ventilation Rates during Multiple Breath Washout
title_full Enhanced Photoacoustic Gas Analyser Response Time and Impact on Accuracy at Fast Ventilation Rates during Multiple Breath Washout
title_fullStr Enhanced Photoacoustic Gas Analyser Response Time and Impact on Accuracy at Fast Ventilation Rates during Multiple Breath Washout
title_full_unstemmed Enhanced Photoacoustic Gas Analyser Response Time and Impact on Accuracy at Fast Ventilation Rates during Multiple Breath Washout
title_short Enhanced Photoacoustic Gas Analyser Response Time and Impact on Accuracy at Fast Ventilation Rates during Multiple Breath Washout
title_sort enhanced photoacoustic gas analyser response time and impact on accuracy at fast ventilation rates during multiple breath washout
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4043749/
https://www.ncbi.nlm.nih.gov/pubmed/24892522
http://dx.doi.org/10.1371/journal.pone.0098487
work_keys_str_mv AT horsleyalex enhancedphotoacousticgasanalyserresponsetimeandimpactonaccuracyatfastventilationratesduringmultiplebreathwashout
AT macleodkenneth enhancedphotoacousticgasanalyserresponsetimeandimpactonaccuracyatfastventilationratesduringmultiplebreathwashout
AT guptaruchi enhancedphotoacousticgasanalyserresponsetimeandimpactonaccuracyatfastventilationratesduringmultiplebreathwashout
AT goddardnick enhancedphotoacousticgasanalyserresponsetimeandimpactonaccuracyatfastventilationratesduringmultiplebreathwashout
AT bellnicholas enhancedphotoacousticgasanalyserresponsetimeandimpactonaccuracyatfastventilationratesduringmultiplebreathwashout