Comparison of glucometers used in hospitals and in outpatient settings with the laboratory reference method in a tertiary care hospital in Mumbai

BACKGROUND: Glucometers allow self-monitoring of blood glucose in a convenient manner. With the availability of various glucometers, there is a persistent attempt to improve the accuracy and the precision of these glucometer readings, so as to match the laboratory values of blood glucose. OBJECTIVE:...

Descripción completa

Detalles Bibliográficos
Autores principales: Ullal, Aarti, Parmar, Girish M., Chauhan, Phulrenu H.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Medknow Publications & Media Pvt Ltd 2013
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4046589/
https://www.ncbi.nlm.nih.gov/pubmed/24910840
http://dx.doi.org/10.4103/2230-8210.123569
_version_ 1782480280520491008
author Ullal, Aarti
Parmar, Girish M.
Chauhan, Phulrenu H.
author_facet Ullal, Aarti
Parmar, Girish M.
Chauhan, Phulrenu H.
author_sort Ullal, Aarti
collection PubMed
description BACKGROUND: Glucometers allow self-monitoring of blood glucose in a convenient manner. With the availability of various glucometers, there is a persistent attempt to improve the accuracy and the precision of these glucometer readings, so as to match the laboratory values of blood glucose. OBJECTIVE: We compared the glucometers used in hospital and out-patient settings with the laboratory reference method. MATERIALS AND METHODS: We analyzed a total of 105 blood samples collected from in-patient and out-patient from our tertiary care hospital. Venous blood samples were collected and checked on six glucometers and the same blood sample was sent to the laboratory for glucose estimation. The laboratory value was used as a reference for comparison. The accuracy was evaluated by the ISO criteria. The results were evaluated by Bland Altman graphs, correlation coefficients, scatter plots and Clarke's error grid analysis. RESULTS: We observed good correlation between bed side glucometer and laboratory automated analyzer. Among the in-patient glucometers Breeze 2, Performa and SureStep, the correlation coefficient was 0.97, 0.96 and 0.88 respectively. Among the outpatient glucometers One touch ultra 2, Active and Contour, the correlation coefficient was 0.97, 0.97 and 0.95 respectively. CONCLUSIONS: There is a good correlation between different glucometers and laboratory values especially in the out-patient settings. Among all in-patient glucometers, SureStep by Johnson and Johnson had least correlation coefficient, whereas all three out-patient glucometers correlated well with the laboratory values.
format Online
Article
Text
id pubmed-4046589
institution National Center for Biotechnology Information
language English
publishDate 2013
publisher Medknow Publications & Media Pvt Ltd
record_format MEDLINE/PubMed
spelling pubmed-40465892014-06-06 Comparison of glucometers used in hospitals and in outpatient settings with the laboratory reference method in a tertiary care hospital in Mumbai Ullal, Aarti Parmar, Girish M. Chauhan, Phulrenu H. Indian J Endocrinol Metab Brief Communication BACKGROUND: Glucometers allow self-monitoring of blood glucose in a convenient manner. With the availability of various glucometers, there is a persistent attempt to improve the accuracy and the precision of these glucometer readings, so as to match the laboratory values of blood glucose. OBJECTIVE: We compared the glucometers used in hospital and out-patient settings with the laboratory reference method. MATERIALS AND METHODS: We analyzed a total of 105 blood samples collected from in-patient and out-patient from our tertiary care hospital. Venous blood samples were collected and checked on six glucometers and the same blood sample was sent to the laboratory for glucose estimation. The laboratory value was used as a reference for comparison. The accuracy was evaluated by the ISO criteria. The results were evaluated by Bland Altman graphs, correlation coefficients, scatter plots and Clarke's error grid analysis. RESULTS: We observed good correlation between bed side glucometer and laboratory automated analyzer. Among the in-patient glucometers Breeze 2, Performa and SureStep, the correlation coefficient was 0.97, 0.96 and 0.88 respectively. Among the outpatient glucometers One touch ultra 2, Active and Contour, the correlation coefficient was 0.97, 0.97 and 0.95 respectively. CONCLUSIONS: There is a good correlation between different glucometers and laboratory values especially in the out-patient settings. Among all in-patient glucometers, SureStep by Johnson and Johnson had least correlation coefficient, whereas all three out-patient glucometers correlated well with the laboratory values. Medknow Publications & Media Pvt Ltd 2013-12 /pmc/articles/PMC4046589/ /pubmed/24910840 http://dx.doi.org/10.4103/2230-8210.123569 Text en Copyright: © Indian Journal of Endocrinology and Metabolism http://creativecommons.org/licenses/by-nc-sa/3.0 This is an open-access article distributed under the terms of the Creative Commons Attribution-Noncommercial-Share Alike 3.0 Unported, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Brief Communication
Ullal, Aarti
Parmar, Girish M.
Chauhan, Phulrenu H.
Comparison of glucometers used in hospitals and in outpatient settings with the laboratory reference method in a tertiary care hospital in Mumbai
title Comparison of glucometers used in hospitals and in outpatient settings with the laboratory reference method in a tertiary care hospital in Mumbai
title_full Comparison of glucometers used in hospitals and in outpatient settings with the laboratory reference method in a tertiary care hospital in Mumbai
title_fullStr Comparison of glucometers used in hospitals and in outpatient settings with the laboratory reference method in a tertiary care hospital in Mumbai
title_full_unstemmed Comparison of glucometers used in hospitals and in outpatient settings with the laboratory reference method in a tertiary care hospital in Mumbai
title_short Comparison of glucometers used in hospitals and in outpatient settings with the laboratory reference method in a tertiary care hospital in Mumbai
title_sort comparison of glucometers used in hospitals and in outpatient settings with the laboratory reference method in a tertiary care hospital in mumbai
topic Brief Communication
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4046589/
https://www.ncbi.nlm.nih.gov/pubmed/24910840
http://dx.doi.org/10.4103/2230-8210.123569
work_keys_str_mv AT ullalaarti comparisonofglucometersusedinhospitalsandinoutpatientsettingswiththelaboratoryreferencemethodinatertiarycarehospitalinmumbai
AT parmargirishm comparisonofglucometersusedinhospitalsandinoutpatientsettingswiththelaboratoryreferencemethodinatertiarycarehospitalinmumbai
AT chauhanphulrenuh comparisonofglucometersusedinhospitalsandinoutpatientsettingswiththelaboratoryreferencemethodinatertiarycarehospitalinmumbai