Analysis of Variation in Total Airborne Bacteria Concentration to Assess the Performance of Biological Safety Cabinets in Microbial Laboratories

BACKGROUND: The purpose of this study was to compare the concentration of total airborne bacteria (TAB) in biosafety cabinets (BSCs) at universities and hospital microbial laboratories to assess the performance of BSCs. METHODS: TAB was determined by using the single-stage Anderson sampler (BioStage...

Descripción completa

Detalles Bibliográficos
Autores principales: Hwang, Sung Ho, Park, Hyun Hee, Yoon, Chung Sik
Formato: Online Artículo Texto
Lenguaje:English
Publicado: 2014
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4048000/
https://www.ncbi.nlm.nih.gov/pubmed/24932416
http://dx.doi.org/10.1016/j.shaw.2014.01.001
_version_ 1782480474109640704
author Hwang, Sung Ho
Park, Hyun Hee
Yoon, Chung Sik
author_facet Hwang, Sung Ho
Park, Hyun Hee
Yoon, Chung Sik
author_sort Hwang, Sung Ho
collection PubMed
description BACKGROUND: The purpose of this study was to compare the concentration of total airborne bacteria (TAB) in biosafety cabinets (BSCs) at universities and hospital microbial laboratories to assess the performance of BSCs. METHODS: TAB was determined by using the single-stage Anderson sampler (BioStage Viable Cascade Impactor). The samples were obtained three times (with the BSC turned off and the shield open; with the BSC turned off and the shield closed; and with the BSC tuned on and operating) from the areas in front of 11 BSCs. RESULTS: TAB concentrations of accredited and nonaccredited BSCs were determined. No significant differences were observed in the TAB concentrations of the accredited BSCs and the nonaccredited BSCs for the areas outside the BSCs in the laboratories (p > 0.05). TAB concentrations for the BSCs sampled with the shield open and the instrument turned off showed differences based on the sampling site outside the BSC in each laboratory. CONCLUSION: These results imply that TAB concentration is not altered by the performance of the BSCs or TAB itself and/or concentration of TAB outside the BSC is not a good index of BSC performance.
format Online
Article
Text
id pubmed-4048000
institution National Center for Biotechnology Information
language English
publishDate 2014
record_format MEDLINE/PubMed
spelling pubmed-40480002014-06-13 Analysis of Variation in Total Airborne Bacteria Concentration to Assess the Performance of Biological Safety Cabinets in Microbial Laboratories Hwang, Sung Ho Park, Hyun Hee Yoon, Chung Sik Saf Health Work Original Article BACKGROUND: The purpose of this study was to compare the concentration of total airborne bacteria (TAB) in biosafety cabinets (BSCs) at universities and hospital microbial laboratories to assess the performance of BSCs. METHODS: TAB was determined by using the single-stage Anderson sampler (BioStage Viable Cascade Impactor). The samples were obtained three times (with the BSC turned off and the shield open; with the BSC turned off and the shield closed; and with the BSC tuned on and operating) from the areas in front of 11 BSCs. RESULTS: TAB concentrations of accredited and nonaccredited BSCs were determined. No significant differences were observed in the TAB concentrations of the accredited BSCs and the nonaccredited BSCs for the areas outside the BSCs in the laboratories (p > 0.05). TAB concentrations for the BSCs sampled with the shield open and the instrument turned off showed differences based on the sampling site outside the BSC in each laboratory. CONCLUSION: These results imply that TAB concentration is not altered by the performance of the BSCs or TAB itself and/or concentration of TAB outside the BSC is not a good index of BSC performance. 2014-01-24 2014-03 /pmc/articles/PMC4048000/ /pubmed/24932416 http://dx.doi.org/10.1016/j.shaw.2014.01.001 Text en © 2014 Published by Elsevier B.V. on behalf of Occupational Safety and Health Research Institute. http://creativecommons.org/licenses/by-nc/3.0/ This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/3.0) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Original Article
Hwang, Sung Ho
Park, Hyun Hee
Yoon, Chung Sik
Analysis of Variation in Total Airborne Bacteria Concentration to Assess the Performance of Biological Safety Cabinets in Microbial Laboratories
title Analysis of Variation in Total Airborne Bacteria Concentration to Assess the Performance of Biological Safety Cabinets in Microbial Laboratories
title_full Analysis of Variation in Total Airborne Bacteria Concentration to Assess the Performance of Biological Safety Cabinets in Microbial Laboratories
title_fullStr Analysis of Variation in Total Airborne Bacteria Concentration to Assess the Performance of Biological Safety Cabinets in Microbial Laboratories
title_full_unstemmed Analysis of Variation in Total Airborne Bacteria Concentration to Assess the Performance of Biological Safety Cabinets in Microbial Laboratories
title_short Analysis of Variation in Total Airborne Bacteria Concentration to Assess the Performance of Biological Safety Cabinets in Microbial Laboratories
title_sort analysis of variation in total airborne bacteria concentration to assess the performance of biological safety cabinets in microbial laboratories
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4048000/
https://www.ncbi.nlm.nih.gov/pubmed/24932416
http://dx.doi.org/10.1016/j.shaw.2014.01.001
work_keys_str_mv AT hwangsungho analysisofvariationintotalairbornebacteriaconcentrationtoassesstheperformanceofbiologicalsafetycabinetsinmicrobiallaboratories
AT parkhyunhee analysisofvariationintotalairbornebacteriaconcentrationtoassesstheperformanceofbiologicalsafetycabinetsinmicrobiallaboratories
AT yoonchungsik analysisofvariationintotalairbornebacteriaconcentrationtoassesstheperformanceofbiologicalsafetycabinetsinmicrobiallaboratories