Immediate and 12 months follow up of function and lead integrity after cranial MRI in 356 patients with conventional cardiac pacemakers
BACKGROUND: Conventional cardiac pacemakers are still often regarded as a contraindication to magnetic resonance imaging (MRI). We conducted this study to support the hypothesis that it is safe to scan patients with cardiac pacemakers in a 1.5 Tesla MRI, if close supervision and monitoring as well a...
Autores principales: | , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
BioMed Central
2014
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4053551/ https://www.ncbi.nlm.nih.gov/pubmed/24903354 http://dx.doi.org/10.1186/1532-429X-16-39 |
Sumario: | BACKGROUND: Conventional cardiac pacemakers are still often regarded as a contraindication to magnetic resonance imaging (MRI). We conducted this study to support the hypothesis that it is safe to scan patients with cardiac pacemakers in a 1.5 Tesla MRI, if close supervision and monitoring as well as adequate pre- and postscan programming is provided. METHODS: We followed up 356 patients (age 61.3 ± 9.1 yrs., 229 men) with single (n = 132) or dual chamber (n = 224) cardiac pacemakers and urgent indication for a cranial MRI for 12 months. The scans were performed at 1.5T. During the scan patients were monitored with a 3-lead ECG and pulse oximetry. Prior to the scan pacemakers were programmed according to our own protocol. RESULTS: All 356 scans were completed without complications. No arrhythmias were induced, programmed parameters remained unchanged. No pacemaker dysfunction was identified. Follow-up examinations were performed immediately, 2 weeks, 2, 6, and 12 months after the scan. There was no significant change of pacing capture threshold (ventricular 0.9 ± 0.4 V@0.4 ms, atrial 0.9 ± 0.3 V@0.4 ms) immediately (ventricular 1.0 ± 0.3 V@0.4 ms, atrial 0.9 ± 0.4 V@0.4 ms) or at 12 months follow-up examinations (ventricular 0.9 ± 0.2 V@0.4 ms, atrial 0.9 ± 0.3 V@0.4 ms). There was no significant change in sensing threshold (8.0 ± 4.0 mV vs. 8.1 ± 4.2 mV ventricular lead, 2.0 ± 0.9 mV vs. 2.1 ± 1.0 mV atrial lead) or lead impedance (ventricular 584 ± 179 Ω vs. 578 ± 188 Ω, atrial 534 ± 176 Ω vs. 532 ± 169 Ω) after 12 months. CONCLUSIONS: This supports the evidence that patients with conventional pacemakers can safely undergo cranial MRI in a 1.5T system with suitable preparation, supervision and precautions. Long term follow-up did not reveal significant changes in pacing capture nor sensing threshold. |
---|