Comparison of Anterior Corneal Curvature Measurements Using a Galilei Dual Scheimpflug Analyzer and Topcon Auto Kerato-Refractometer
Purpose. To compare anterior corneal keratometry (K) measurements taken by a dual Scheimpflug analyzer and an auto kerato-refractometer. Methods. Sixty-four normal eyes underwent keratometric measurements with both devices. The repeatability of the auto kerato-refractometer measurements was assessed...
Autores principales: | , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Hindawi Publishing Corporation
2014
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4055652/ https://www.ncbi.nlm.nih.gov/pubmed/24967095 http://dx.doi.org/10.1155/2014/140628 |
_version_ | 1782320697373097984 |
---|---|
author | Wang, Xiaogang Dong, Jing Wu, Qiang |
author_facet | Wang, Xiaogang Dong, Jing Wu, Qiang |
author_sort | Wang, Xiaogang |
collection | PubMed |
description | Purpose. To compare anterior corneal keratometry (K) measurements taken by a dual Scheimpflug analyzer and an auto kerato-refractometer. Methods. Sixty-four normal eyes underwent keratometric measurements with both devices. The repeatability of the auto kerato-refractometer measurements was assessed by calculating the coefficient of variation (COV). The interdevice agreement was evaluated using the Bland-Altman analysis, Pearson correlation coefficient, and paired two-tailed t-test. Results. The COV of the flat K and steep K measurements taken by the auto kerato-refractometer were 0.21% and 0.29%, respectively. There were no significant differences between the steep K and average K measurements for the Topcon and Galilei devices (P = 0.475, and P = 0.137, resp.). The Galilei flat K values were lower than those of the Topcon (P = 0.002). Both of the instruments showed good agreement for all anterior corneal keratometric values. There was a significant linear correlation between the Galilei and Topcon devices for the flat K (r = 0.989, P < 0.0001), steep K (r = 0.987, P < 0.0001), and average K values (r = 0.994, P < 0.0001). Conclusions. The anterior corneal flat keratometric values were not interchangeable between the Galilei and Topcon devices. However, the measurements of the two instruments showed significant linear correlation with each other. |
format | Online Article Text |
id | pubmed-4055652 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2014 |
publisher | Hindawi Publishing Corporation |
record_format | MEDLINE/PubMed |
spelling | pubmed-40556522014-06-25 Comparison of Anterior Corneal Curvature Measurements Using a Galilei Dual Scheimpflug Analyzer and Topcon Auto Kerato-Refractometer Wang, Xiaogang Dong, Jing Wu, Qiang J Ophthalmol Research Article Purpose. To compare anterior corneal keratometry (K) measurements taken by a dual Scheimpflug analyzer and an auto kerato-refractometer. Methods. Sixty-four normal eyes underwent keratometric measurements with both devices. The repeatability of the auto kerato-refractometer measurements was assessed by calculating the coefficient of variation (COV). The interdevice agreement was evaluated using the Bland-Altman analysis, Pearson correlation coefficient, and paired two-tailed t-test. Results. The COV of the flat K and steep K measurements taken by the auto kerato-refractometer were 0.21% and 0.29%, respectively. There were no significant differences between the steep K and average K measurements for the Topcon and Galilei devices (P = 0.475, and P = 0.137, resp.). The Galilei flat K values were lower than those of the Topcon (P = 0.002). Both of the instruments showed good agreement for all anterior corneal keratometric values. There was a significant linear correlation between the Galilei and Topcon devices for the flat K (r = 0.989, P < 0.0001), steep K (r = 0.987, P < 0.0001), and average K values (r = 0.994, P < 0.0001). Conclusions. The anterior corneal flat keratometric values were not interchangeable between the Galilei and Topcon devices. However, the measurements of the two instruments showed significant linear correlation with each other. Hindawi Publishing Corporation 2014 2014-05-22 /pmc/articles/PMC4055652/ /pubmed/24967095 http://dx.doi.org/10.1155/2014/140628 Text en Copyright © 2014 Xiaogang Wang et al. https://creativecommons.org/licenses/by/3.0/ This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Research Article Wang, Xiaogang Dong, Jing Wu, Qiang Comparison of Anterior Corneal Curvature Measurements Using a Galilei Dual Scheimpflug Analyzer and Topcon Auto Kerato-Refractometer |
title | Comparison of Anterior Corneal Curvature Measurements Using a Galilei Dual Scheimpflug Analyzer and Topcon Auto Kerato-Refractometer |
title_full | Comparison of Anterior Corneal Curvature Measurements Using a Galilei Dual Scheimpflug Analyzer and Topcon Auto Kerato-Refractometer |
title_fullStr | Comparison of Anterior Corneal Curvature Measurements Using a Galilei Dual Scheimpflug Analyzer and Topcon Auto Kerato-Refractometer |
title_full_unstemmed | Comparison of Anterior Corneal Curvature Measurements Using a Galilei Dual Scheimpflug Analyzer and Topcon Auto Kerato-Refractometer |
title_short | Comparison of Anterior Corneal Curvature Measurements Using a Galilei Dual Scheimpflug Analyzer and Topcon Auto Kerato-Refractometer |
title_sort | comparison of anterior corneal curvature measurements using a galilei dual scheimpflug analyzer and topcon auto kerato-refractometer |
topic | Research Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4055652/ https://www.ncbi.nlm.nih.gov/pubmed/24967095 http://dx.doi.org/10.1155/2014/140628 |
work_keys_str_mv | AT wangxiaogang comparisonofanteriorcornealcurvaturemeasurementsusingagalileidualscheimpfluganalyzerandtopconautokeratorefractometer AT dongjing comparisonofanteriorcornealcurvaturemeasurementsusingagalileidualscheimpfluganalyzerandtopconautokeratorefractometer AT wuqiang comparisonofanteriorcornealcurvaturemeasurementsusingagalileidualscheimpfluganalyzerandtopconautokeratorefractometer |