Cargando…

Percutaneous Retrieval and Redeployment of an Atrial Septal Occluder under Three-Dimensional Transesophageal Echocardiographic Guidance: A Case Report

Percutaneous device closure for secundum atrial septal defects (ASDs) has been performed commonly and safely with high success rates. However, it is still challenging to close ASDs that are surrounded with deficient or hypermobile rims and could be compromised with an unexpected migration of device....

Descripción completa

Detalles Bibliográficos
Autores principales: Ahn, Jinhee, Kim, Jin Hee, Choi, Jung Hyun, Oh, Jun-Hyok
Formato: Online Artículo Texto
Lenguaje:English
Publicado: The Korean Academy of Medical Sciences 2014
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4055824/
https://www.ncbi.nlm.nih.gov/pubmed/24932092
http://dx.doi.org/10.3346/jkms.2014.29.6.871
Descripción
Sumario:Percutaneous device closure for secundum atrial septal defects (ASDs) has been performed commonly and safely with high success rates. However, it is still challenging to close ASDs that are surrounded with deficient or hypermobile rims and could be compromised with an unexpected migration of device. We report a case of percutaneous Amplazter Septal Occluder (ASO; St. Jude Medical Inc., St. Paul, Minnesota, USA) device closure for an ASD with a thin and floppy interatrial septum, which immediately migrated into the right atrium and was not pulled back into the delivery sheath. To our knowledge, this is the first report on a successful percutaneous retrieval and redeployment of the device in such a situation, preventing any vascular injury or unplanned emergency open heart surgery. GRAPHICAL ABSTRACT: [Image: see text]