Evaluation of the Steelex M600H coagulometer prothrombin time-international normalized ratio assay with Steelex test reagents

INTRODUCTION: The aim of the present study was to validate prothrombin time (PT) international normalized ratio (INR) results obtained using Steelex test reagents and a Steelex coagulometer (Steelex Scientific Instrument Company, Beijing, China), in comparison with use of a well-established standard...

Descripción completa

Detalles Bibliográficos
Autores principales: Budak, Yasemin U., Huysal, Kağan, Polat, Murat, Tarakçı, Gülsevil, Uçar, Hakan
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Croatian Society of Medical Biochemistry and Laboratory Medicine 2012
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4062317/
https://www.ncbi.nlm.nih.gov/pubmed/22384527
_version_ 1782321627736834048
author Budak, Yasemin U.
Huysal, Kağan
Polat, Murat
Tarakçı, Gülsevil
Uçar, Hakan
author_facet Budak, Yasemin U.
Huysal, Kağan
Polat, Murat
Tarakçı, Gülsevil
Uçar, Hakan
author_sort Budak, Yasemin U.
collection PubMed
description INTRODUCTION: The aim of the present study was to validate prothrombin time (PT) international normalized ratio (INR) results obtained using Steelex test reagents and a Steelex coagulometer (Steelex Scientific Instrument Company, Beijing, China), in comparison with use of a well-established standard test employing Pacific Hemostasis reagents (Fisher Diagnostics, Middletown, VA, USA) and Teco Coatron A4 coagulometer (Teco Medical Instruments GmbH, Neufahrn, Germany). MATERIALS AND METHODS: Between- and within-day coefficients of variation (CVs) of both assays were calculated using control samples provided by the test manufacturers. Samples from 90 subjects were collected and INR values were determined in a double-blind parallel manner employing both systems. RESULTS: The within-day coefficients of variation (CVs) in INR estimates ranged from 2.6% (INR = 1.12) to 3.1% (INR = 2.51) for the Steelex system and from 2.1% (INR = 1.09) to 1.8% (INR = 2.8) for the Pacific test; the between-day values ran from 3.4% (INR = 1.16) to 7.9% (INR = 2.64) and from 3.3% (INR = 1.1) to 2.3% (INR = 2.7), respectively. Passing-Bablok fit of the of the Steelex and Pacific methods yielded the equation: Steelex INR = 0.85 (0.79–0.91) × Pacific INR + 0.12 (−0.02–0.21), whereas the CUSUM linearity P value was < 0.01. The mean bias as determined by the Bland-Altman test was −0.156 (−0.912–0.600). CONCLUSION: The results obtained using Steelex reagents and the M600H coagulometer are not equivalent to those obtained using Pacific Hemostasis reagents and a Teco Coatron A4 coagulometer, at least in the therapeutic range.
format Online
Article
Text
id pubmed-4062317
institution National Center for Biotechnology Information
language English
publishDate 2012
publisher Croatian Society of Medical Biochemistry and Laboratory Medicine
record_format MEDLINE/PubMed
spelling pubmed-40623172014-06-23 Evaluation of the Steelex M600H coagulometer prothrombin time-international normalized ratio assay with Steelex test reagents Budak, Yasemin U. Huysal, Kağan Polat, Murat Tarakçı, Gülsevil Uçar, Hakan Biochem Med (Zagreb) Original Article INTRODUCTION: The aim of the present study was to validate prothrombin time (PT) international normalized ratio (INR) results obtained using Steelex test reagents and a Steelex coagulometer (Steelex Scientific Instrument Company, Beijing, China), in comparison with use of a well-established standard test employing Pacific Hemostasis reagents (Fisher Diagnostics, Middletown, VA, USA) and Teco Coatron A4 coagulometer (Teco Medical Instruments GmbH, Neufahrn, Germany). MATERIALS AND METHODS: Between- and within-day coefficients of variation (CVs) of both assays were calculated using control samples provided by the test manufacturers. Samples from 90 subjects were collected and INR values were determined in a double-blind parallel manner employing both systems. RESULTS: The within-day coefficients of variation (CVs) in INR estimates ranged from 2.6% (INR = 1.12) to 3.1% (INR = 2.51) for the Steelex system and from 2.1% (INR = 1.09) to 1.8% (INR = 2.8) for the Pacific test; the between-day values ran from 3.4% (INR = 1.16) to 7.9% (INR = 2.64) and from 3.3% (INR = 1.1) to 2.3% (INR = 2.7), respectively. Passing-Bablok fit of the of the Steelex and Pacific methods yielded the equation: Steelex INR = 0.85 (0.79–0.91) × Pacific INR + 0.12 (−0.02–0.21), whereas the CUSUM linearity P value was < 0.01. The mean bias as determined by the Bland-Altman test was −0.156 (−0.912–0.600). CONCLUSION: The results obtained using Steelex reagents and the M600H coagulometer are not equivalent to those obtained using Pacific Hemostasis reagents and a Teco Coatron A4 coagulometer, at least in the therapeutic range. Croatian Society of Medical Biochemistry and Laboratory Medicine 2012-02-15 /pmc/articles/PMC4062317/ /pubmed/22384527 Text en © Copyright by Croatian Society of Medical Biochemistry and Laboratory Medicine. This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc-nd/3.0/) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Original Article
Budak, Yasemin U.
Huysal, Kağan
Polat, Murat
Tarakçı, Gülsevil
Uçar, Hakan
Evaluation of the Steelex M600H coagulometer prothrombin time-international normalized ratio assay with Steelex test reagents
title Evaluation of the Steelex M600H coagulometer prothrombin time-international normalized ratio assay with Steelex test reagents
title_full Evaluation of the Steelex M600H coagulometer prothrombin time-international normalized ratio assay with Steelex test reagents
title_fullStr Evaluation of the Steelex M600H coagulometer prothrombin time-international normalized ratio assay with Steelex test reagents
title_full_unstemmed Evaluation of the Steelex M600H coagulometer prothrombin time-international normalized ratio assay with Steelex test reagents
title_short Evaluation of the Steelex M600H coagulometer prothrombin time-international normalized ratio assay with Steelex test reagents
title_sort evaluation of the steelex m600h coagulometer prothrombin time-international normalized ratio assay with steelex test reagents
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4062317/
https://www.ncbi.nlm.nih.gov/pubmed/22384527
work_keys_str_mv AT budakyaseminu evaluationofthesteelexm600hcoagulometerprothrombintimeinternationalnormalizedratioassaywithsteelextestreagents
AT huysalkagan evaluationofthesteelexm600hcoagulometerprothrombintimeinternationalnormalizedratioassaywithsteelextestreagents
AT polatmurat evaluationofthesteelexm600hcoagulometerprothrombintimeinternationalnormalizedratioassaywithsteelextestreagents
AT tarakcıgulsevil evaluationofthesteelexm600hcoagulometerprothrombintimeinternationalnormalizedratioassaywithsteelextestreagents
AT ucarhakan evaluationofthesteelexm600hcoagulometerprothrombintimeinternationalnormalizedratioassaywithsteelextestreagents