Cargando…
Evaluation of the Steelex M600H coagulometer prothrombin time-international normalized ratio assay with Steelex test reagents
INTRODUCTION: The aim of the present study was to validate prothrombin time (PT) international normalized ratio (INR) results obtained using Steelex test reagents and a Steelex coagulometer (Steelex Scientific Instrument Company, Beijing, China), in comparison with use of a well-established standard...
Autores principales: | , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Croatian Society of Medical Biochemistry and Laboratory Medicine
2012
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4062317/ https://www.ncbi.nlm.nih.gov/pubmed/22384527 |
_version_ | 1782321627736834048 |
---|---|
author | Budak, Yasemin U. Huysal, Kağan Polat, Murat Tarakçı, Gülsevil Uçar, Hakan |
author_facet | Budak, Yasemin U. Huysal, Kağan Polat, Murat Tarakçı, Gülsevil Uçar, Hakan |
author_sort | Budak, Yasemin U. |
collection | PubMed |
description | INTRODUCTION: The aim of the present study was to validate prothrombin time (PT) international normalized ratio (INR) results obtained using Steelex test reagents and a Steelex coagulometer (Steelex Scientific Instrument Company, Beijing, China), in comparison with use of a well-established standard test employing Pacific Hemostasis reagents (Fisher Diagnostics, Middletown, VA, USA) and Teco Coatron A4 coagulometer (Teco Medical Instruments GmbH, Neufahrn, Germany). MATERIALS AND METHODS: Between- and within-day coefficients of variation (CVs) of both assays were calculated using control samples provided by the test manufacturers. Samples from 90 subjects were collected and INR values were determined in a double-blind parallel manner employing both systems. RESULTS: The within-day coefficients of variation (CVs) in INR estimates ranged from 2.6% (INR = 1.12) to 3.1% (INR = 2.51) for the Steelex system and from 2.1% (INR = 1.09) to 1.8% (INR = 2.8) for the Pacific test; the between-day values ran from 3.4% (INR = 1.16) to 7.9% (INR = 2.64) and from 3.3% (INR = 1.1) to 2.3% (INR = 2.7), respectively. Passing-Bablok fit of the of the Steelex and Pacific methods yielded the equation: Steelex INR = 0.85 (0.79–0.91) × Pacific INR + 0.12 (−0.02–0.21), whereas the CUSUM linearity P value was < 0.01. The mean bias as determined by the Bland-Altman test was −0.156 (−0.912–0.600). CONCLUSION: The results obtained using Steelex reagents and the M600H coagulometer are not equivalent to those obtained using Pacific Hemostasis reagents and a Teco Coatron A4 coagulometer, at least in the therapeutic range. |
format | Online Article Text |
id | pubmed-4062317 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2012 |
publisher | Croatian Society of Medical Biochemistry and Laboratory Medicine |
record_format | MEDLINE/PubMed |
spelling | pubmed-40623172014-06-23 Evaluation of the Steelex M600H coagulometer prothrombin time-international normalized ratio assay with Steelex test reagents Budak, Yasemin U. Huysal, Kağan Polat, Murat Tarakçı, Gülsevil Uçar, Hakan Biochem Med (Zagreb) Original Article INTRODUCTION: The aim of the present study was to validate prothrombin time (PT) international normalized ratio (INR) results obtained using Steelex test reagents and a Steelex coagulometer (Steelex Scientific Instrument Company, Beijing, China), in comparison with use of a well-established standard test employing Pacific Hemostasis reagents (Fisher Diagnostics, Middletown, VA, USA) and Teco Coatron A4 coagulometer (Teco Medical Instruments GmbH, Neufahrn, Germany). MATERIALS AND METHODS: Between- and within-day coefficients of variation (CVs) of both assays were calculated using control samples provided by the test manufacturers. Samples from 90 subjects were collected and INR values were determined in a double-blind parallel manner employing both systems. RESULTS: The within-day coefficients of variation (CVs) in INR estimates ranged from 2.6% (INR = 1.12) to 3.1% (INR = 2.51) for the Steelex system and from 2.1% (INR = 1.09) to 1.8% (INR = 2.8) for the Pacific test; the between-day values ran from 3.4% (INR = 1.16) to 7.9% (INR = 2.64) and from 3.3% (INR = 1.1) to 2.3% (INR = 2.7), respectively. Passing-Bablok fit of the of the Steelex and Pacific methods yielded the equation: Steelex INR = 0.85 (0.79–0.91) × Pacific INR + 0.12 (−0.02–0.21), whereas the CUSUM linearity P value was < 0.01. The mean bias as determined by the Bland-Altman test was −0.156 (−0.912–0.600). CONCLUSION: The results obtained using Steelex reagents and the M600H coagulometer are not equivalent to those obtained using Pacific Hemostasis reagents and a Teco Coatron A4 coagulometer, at least in the therapeutic range. Croatian Society of Medical Biochemistry and Laboratory Medicine 2012-02-15 /pmc/articles/PMC4062317/ /pubmed/22384527 Text en © Copyright by Croatian Society of Medical Biochemistry and Laboratory Medicine. This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc-nd/3.0/) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Original Article Budak, Yasemin U. Huysal, Kağan Polat, Murat Tarakçı, Gülsevil Uçar, Hakan Evaluation of the Steelex M600H coagulometer prothrombin time-international normalized ratio assay with Steelex test reagents |
title | Evaluation of the Steelex M600H coagulometer prothrombin time-international normalized ratio assay with Steelex test reagents |
title_full | Evaluation of the Steelex M600H coagulometer prothrombin time-international normalized ratio assay with Steelex test reagents |
title_fullStr | Evaluation of the Steelex M600H coagulometer prothrombin time-international normalized ratio assay with Steelex test reagents |
title_full_unstemmed | Evaluation of the Steelex M600H coagulometer prothrombin time-international normalized ratio assay with Steelex test reagents |
title_short | Evaluation of the Steelex M600H coagulometer prothrombin time-international normalized ratio assay with Steelex test reagents |
title_sort | evaluation of the steelex m600h coagulometer prothrombin time-international normalized ratio assay with steelex test reagents |
topic | Original Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4062317/ https://www.ncbi.nlm.nih.gov/pubmed/22384527 |
work_keys_str_mv | AT budakyaseminu evaluationofthesteelexm600hcoagulometerprothrombintimeinternationalnormalizedratioassaywithsteelextestreagents AT huysalkagan evaluationofthesteelexm600hcoagulometerprothrombintimeinternationalnormalizedratioassaywithsteelextestreagents AT polatmurat evaluationofthesteelexm600hcoagulometerprothrombintimeinternationalnormalizedratioassaywithsteelextestreagents AT tarakcıgulsevil evaluationofthesteelexm600hcoagulometerprothrombintimeinternationalnormalizedratioassaywithsteelextestreagents AT ucarhakan evaluationofthesteelexm600hcoagulometerprothrombintimeinternationalnormalizedratioassaywithsteelextestreagents |