How Reliable Are ATP Bioluminescence Meters in Assessing Decontamination of Environmental Surfaces in Healthcare Settings?
BACKGROUND: Meters based on adenosine triphosphate (ATP) bioluminescence measurements in relative light units (RLU) are often used to rapidly assess the level of cleanliness of environmental surfaces in healthcare and other settings. Can such ATP measurements be adversely affected by factors such as...
Autores principales: | , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Public Library of Science
2014
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4062432/ https://www.ncbi.nlm.nih.gov/pubmed/24940751 http://dx.doi.org/10.1371/journal.pone.0099951 |
_version_ | 1782321647752052736 |
---|---|
author | Omidbakhsh, Navid Ahmadpour, Faraz Kenny, Nicole |
author_facet | Omidbakhsh, Navid Ahmadpour, Faraz Kenny, Nicole |
author_sort | Omidbakhsh, Navid |
collection | PubMed |
description | BACKGROUND: Meters based on adenosine triphosphate (ATP) bioluminescence measurements in relative light units (RLU) are often used to rapidly assess the level of cleanliness of environmental surfaces in healthcare and other settings. Can such ATP measurements be adversely affected by factors such as soil and cleaner-disinfectant chemistry? OBJECTIVE: This study tested a number of leading ATP meters for their sensitivity, linearity of the measurements, correlation of the readings to the actual microbial contamination, and the potential disinfectant chemicals’ interference in their readings. METHODS: First, solutions of pure ATP in various concentrations were used to construct a standard curve and determine linearity and sensitivity. Serial dilutions of a broth culture of Staphylococcus aureus, as a representative nosocomial pathogen, were then used to determine if a given meter’s ATP readings correlated with the actual CFUs. Next, various types of disinfectant chemistries were tested for their potential to interfere with the standard ATP readings. RESULTS: All four ATP meters tested herein demonstrated acceptable linearity and repeatability in their readings. However, there were significant differences in their sensitivity to detect the levels of viable microorganisms on experimentally contaminated surfaces. Further, most disinfectant chemistries tested here quenched the ATP readings variably in different ATP meters evaluated. CONCLUSIONS: Apart from their limited sensitivity in detecting low levels of microbial contamination, the ATP meters tested were also prone to interference by different disinfectant chemistries. |
format | Online Article Text |
id | pubmed-4062432 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2014 |
publisher | Public Library of Science |
record_format | MEDLINE/PubMed |
spelling | pubmed-40624322014-06-24 How Reliable Are ATP Bioluminescence Meters in Assessing Decontamination of Environmental Surfaces in Healthcare Settings? Omidbakhsh, Navid Ahmadpour, Faraz Kenny, Nicole PLoS One Research Article BACKGROUND: Meters based on adenosine triphosphate (ATP) bioluminescence measurements in relative light units (RLU) are often used to rapidly assess the level of cleanliness of environmental surfaces in healthcare and other settings. Can such ATP measurements be adversely affected by factors such as soil and cleaner-disinfectant chemistry? OBJECTIVE: This study tested a number of leading ATP meters for their sensitivity, linearity of the measurements, correlation of the readings to the actual microbial contamination, and the potential disinfectant chemicals’ interference in their readings. METHODS: First, solutions of pure ATP in various concentrations were used to construct a standard curve and determine linearity and sensitivity. Serial dilutions of a broth culture of Staphylococcus aureus, as a representative nosocomial pathogen, were then used to determine if a given meter’s ATP readings correlated with the actual CFUs. Next, various types of disinfectant chemistries were tested for their potential to interfere with the standard ATP readings. RESULTS: All four ATP meters tested herein demonstrated acceptable linearity and repeatability in their readings. However, there were significant differences in their sensitivity to detect the levels of viable microorganisms on experimentally contaminated surfaces. Further, most disinfectant chemistries tested here quenched the ATP readings variably in different ATP meters evaluated. CONCLUSIONS: Apart from their limited sensitivity in detecting low levels of microbial contamination, the ATP meters tested were also prone to interference by different disinfectant chemistries. Public Library of Science 2014-06-18 /pmc/articles/PMC4062432/ /pubmed/24940751 http://dx.doi.org/10.1371/journal.pone.0099951 Text en © 2014 Omidbakhsh et al http://creativecommons.org/licenses/by/4.0/ This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are properly credited. |
spellingShingle | Research Article Omidbakhsh, Navid Ahmadpour, Faraz Kenny, Nicole How Reliable Are ATP Bioluminescence Meters in Assessing Decontamination of Environmental Surfaces in Healthcare Settings? |
title | How Reliable Are ATP Bioluminescence Meters in Assessing Decontamination of Environmental Surfaces in Healthcare Settings? |
title_full | How Reliable Are ATP Bioluminescence Meters in Assessing Decontamination of Environmental Surfaces in Healthcare Settings? |
title_fullStr | How Reliable Are ATP Bioluminescence Meters in Assessing Decontamination of Environmental Surfaces in Healthcare Settings? |
title_full_unstemmed | How Reliable Are ATP Bioluminescence Meters in Assessing Decontamination of Environmental Surfaces in Healthcare Settings? |
title_short | How Reliable Are ATP Bioluminescence Meters in Assessing Decontamination of Environmental Surfaces in Healthcare Settings? |
title_sort | how reliable are atp bioluminescence meters in assessing decontamination of environmental surfaces in healthcare settings? |
topic | Research Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4062432/ https://www.ncbi.nlm.nih.gov/pubmed/24940751 http://dx.doi.org/10.1371/journal.pone.0099951 |
work_keys_str_mv | AT omidbakhshnavid howreliableareatpbioluminescencemetersinassessingdecontaminationofenvironmentalsurfacesinhealthcaresettings AT ahmadpourfaraz howreliableareatpbioluminescencemetersinassessingdecontaminationofenvironmentalsurfacesinhealthcaresettings AT kennynicole howreliableareatpbioluminescencemetersinassessingdecontaminationofenvironmentalsurfacesinhealthcaresettings |