Limit of blank and limit of detection of Plasmodium falciparum thick blood smear microscopy in a routine setting in Central Africa
BACKGROUND: Proper malaria diagnosis depends on the detection of asexual forms of Plasmodium spp. in the blood. Thick blood smear microscopy is the accepted gold standard of malaria diagnosis and is widely implemented. Surprisingly, diagnostic performance of this method is not well investigated and...
Autores principales: | , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
BioMed Central
2014
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4069274/ https://www.ncbi.nlm.nih.gov/pubmed/24929248 http://dx.doi.org/10.1186/1475-2875-13-234 |
_version_ | 1782322531982639104 |
---|---|
author | Joanny, Fanny Löhr, Sascha JZ Engleitner, Thomas Lell, Bertrand Mordmüller, Benjamin |
author_facet | Joanny, Fanny Löhr, Sascha JZ Engleitner, Thomas Lell, Bertrand Mordmüller, Benjamin |
author_sort | Joanny, Fanny |
collection | PubMed |
description | BACKGROUND: Proper malaria diagnosis depends on the detection of asexual forms of Plasmodium spp. in the blood. Thick blood smear microscopy is the accepted gold standard of malaria diagnosis and is widely implemented. Surprisingly, diagnostic performance of this method is not well investigated and many clinicians in African routine settings base treatment decisions independent of microscopy results. This leads to overtreatment and poor management of other febrile diseases. Implementation of quality control programmes is recommended, but requires sustained funding, external logistic support and constant training and supervision of the staff. This study describes an easily applicable method to assess the performance of thick blood smear microscopy by determining the limit of blank and limit of detection. These two values are representative of the diagnostic quality and allow the correct discrimination between positive and negative samples. METHODS: Standard-conform methodology was applied and adapted to determine the limit of blank and the limit of detection of two thick blood smear microscopy methods (WHO and Lambaréné method) in a research centre in Lambaréné, Gabon. Duplicates of negative and low parasitaemia thick blood smears were read by several microscopists. The mean and standard deviation of the results were used to calculate the limit of blank and subsequently the limit of detection. RESULTS: The limit of blank was 0 parasites/μL for both methods. The limit of detection was 62 and 88 parasites/μL for the Lambaréné and WHO method, respectively. CONCLUSION: With a simple, back-of-the-envelope calculation, the performance of two malaria microscopy methods can be measured. These results are specific for each diagnostic unit and cannot be generalized but implementation of a system to control microscopy performance can improve confidence in parasitological results and thereby strengthen malaria control. |
format | Online Article Text |
id | pubmed-4069274 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2014 |
publisher | BioMed Central |
record_format | MEDLINE/PubMed |
spelling | pubmed-40692742014-06-27 Limit of blank and limit of detection of Plasmodium falciparum thick blood smear microscopy in a routine setting in Central Africa Joanny, Fanny Löhr, Sascha JZ Engleitner, Thomas Lell, Bertrand Mordmüller, Benjamin Malar J Methodology BACKGROUND: Proper malaria diagnosis depends on the detection of asexual forms of Plasmodium spp. in the blood. Thick blood smear microscopy is the accepted gold standard of malaria diagnosis and is widely implemented. Surprisingly, diagnostic performance of this method is not well investigated and many clinicians in African routine settings base treatment decisions independent of microscopy results. This leads to overtreatment and poor management of other febrile diseases. Implementation of quality control programmes is recommended, but requires sustained funding, external logistic support and constant training and supervision of the staff. This study describes an easily applicable method to assess the performance of thick blood smear microscopy by determining the limit of blank and limit of detection. These two values are representative of the diagnostic quality and allow the correct discrimination between positive and negative samples. METHODS: Standard-conform methodology was applied and adapted to determine the limit of blank and the limit of detection of two thick blood smear microscopy methods (WHO and Lambaréné method) in a research centre in Lambaréné, Gabon. Duplicates of negative and low parasitaemia thick blood smears were read by several microscopists. The mean and standard deviation of the results were used to calculate the limit of blank and subsequently the limit of detection. RESULTS: The limit of blank was 0 parasites/μL for both methods. The limit of detection was 62 and 88 parasites/μL for the Lambaréné and WHO method, respectively. CONCLUSION: With a simple, back-of-the-envelope calculation, the performance of two malaria microscopy methods can be measured. These results are specific for each diagnostic unit and cannot be generalized but implementation of a system to control microscopy performance can improve confidence in parasitological results and thereby strengthen malaria control. BioMed Central 2014-06-14 /pmc/articles/PMC4069274/ /pubmed/24929248 http://dx.doi.org/10.1186/1475-2875-13-234 Text en Copyright © 2014 Joanny et al.; licensee BioMed Central Ltd. http://creativecommons.org/licenses/by/4.0 This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly credited. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated. |
spellingShingle | Methodology Joanny, Fanny Löhr, Sascha JZ Engleitner, Thomas Lell, Bertrand Mordmüller, Benjamin Limit of blank and limit of detection of Plasmodium falciparum thick blood smear microscopy in a routine setting in Central Africa |
title | Limit of blank and limit of detection of Plasmodium falciparum thick blood smear microscopy in a routine setting in Central Africa |
title_full | Limit of blank and limit of detection of Plasmodium falciparum thick blood smear microscopy in a routine setting in Central Africa |
title_fullStr | Limit of blank and limit of detection of Plasmodium falciparum thick blood smear microscopy in a routine setting in Central Africa |
title_full_unstemmed | Limit of blank and limit of detection of Plasmodium falciparum thick blood smear microscopy in a routine setting in Central Africa |
title_short | Limit of blank and limit of detection of Plasmodium falciparum thick blood smear microscopy in a routine setting in Central Africa |
title_sort | limit of blank and limit of detection of plasmodium falciparum thick blood smear microscopy in a routine setting in central africa |
topic | Methodology |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4069274/ https://www.ncbi.nlm.nih.gov/pubmed/24929248 http://dx.doi.org/10.1186/1475-2875-13-234 |
work_keys_str_mv | AT joannyfanny limitofblankandlimitofdetectionofplasmodiumfalciparumthickbloodsmearmicroscopyinaroutinesettingincentralafrica AT lohrsaschajz limitofblankandlimitofdetectionofplasmodiumfalciparumthickbloodsmearmicroscopyinaroutinesettingincentralafrica AT engleitnerthomas limitofblankandlimitofdetectionofplasmodiumfalciparumthickbloodsmearmicroscopyinaroutinesettingincentralafrica AT lellbertrand limitofblankandlimitofdetectionofplasmodiumfalciparumthickbloodsmearmicroscopyinaroutinesettingincentralafrica AT mordmullerbenjamin limitofblankandlimitofdetectionofplasmodiumfalciparumthickbloodsmearmicroscopyinaroutinesettingincentralafrica |