Agreement of Two Different Spectral Domain Optical Coherence Tomography Instruments for Retinal Nerve Fiber Layer Measurements
PURPOSE: To determine the agreement between Spectralis and Cirrus spectral domain optical coherence tomography (SD-OCT) measurements of peripapillary retinal nerve fiber layer (RNFL) thickness. METHODS: Suspected or confirmed cases of glaucoma who met the inclusion criteria underwent peripapillary R...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Ophthalmic Research Center
2014
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4074471/ https://www.ncbi.nlm.nih.gov/pubmed/24982729 |
Sumario: | PURPOSE: To determine the agreement between Spectralis and Cirrus spectral domain optical coherence tomography (SD-OCT) measurements of peripapillary retinal nerve fiber layer (RNFL) thickness. METHODS: Suspected or confirmed cases of glaucoma who met the inclusion criteria underwent peripapillary RNFL thickness measurement using both the Spectralis and Cirrus on the same day within a few minutes. RESULTS: Measurements were performed on 103 eyes of 103 patients with mean age of 50.4±17.7 years. Mean RNFL thickness was 89.22±15.87 versus 84.54±13.68 µm using Spectralis and Cirrus, respectively. The difference between measurements and the average of paired measurements with the two devices showed a significant linear relationship. Bland-Altman plots demonstrated that Spectralis thickness values were systematically larger than that of Cirrus. CONCLUSION: Spectralis OCT generates higher peripapillary RNFL thickness readings as compared to Cirrus OCT; this should be kept in mind when values obtained with different instruments are compared during follow-up. |
---|