Cargando…

Effect that an educational program for cystic fibrosis patients and caregivers has on the contamination of home nebulizers

OBJECTIVE: To describe the pathogens found in home nebulizers and in respiratory samples of cystic fibrosis (CF) patients, and to evaluate the effect that a standardized instruction regarding cleaning and disinfection of nebulizers has on the frequency of nebulizer contamination. METHODS: We include...

Descripción completa

Detalles Bibliográficos
Autores principales: Zuana, Adriana Della, Garcia, Doroti de Oliveira, Juliani, Regina Célia Turola Passos, da Silva, Luiz Vicente Ribeiro Ferreira
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Sociedade Brasileira de Pneumologia e Tisiologia 2014
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4083641/
https://www.ncbi.nlm.nih.gov/pubmed/24831395
http://dx.doi.org/10.1590/S1806-37132014000200004
Descripción
Sumario:OBJECTIVE: To describe the pathogens found in home nebulizers and in respiratory samples of cystic fibrosis (CF) patients, and to evaluate the effect that a standardized instruction regarding cleaning and disinfection of nebulizers has on the frequency of nebulizer contamination. METHODS: We included 40 CF patients (22 males), all of whom used the same model of nebulizer. The median patient age was 11.2 ± 3.74 years. We collected samples from the nebulizer mouthpiece and cup, using a sterile swab moistened with sterile saline. Respiratory samples were collected by asking patients to expectorate into a sterile container or with oropharyngeal swabs after cough stimulation. Cultures were performed on selective media, and bacteria were identified by classical biochemical tests. Patients received oral and written instructions regarding the cleaning and disinfection of nebulizers. All determinations were repeated an average of two months later. RESULTS: Contamination of the nebulizer (any part) was detected in 23 cases (57.5%). The nebulizer mouthpiece and cup were found to be contaminated in 16 (40.0%) and 19 (47.5%), respectively. After the standardized instruction had been given, there was a significant decrease in the proportion of contaminated nebulizers (43.5%). CONCLUSIONS: In our sample of CF patients, nebulizer contamination was common, indicating the need for improvement in patient practices regarding the cleaning and disinfection of their nebulizers. A one-time educational intervention could have a significant positive impact.