Influence of Vitamin D Binding Protein on Accuracy of 25-Hydroxyvitamin D Measurement Using the ADVIA Centaur Vitamin D Total Assay
Vitamin D status in different populations relies on accurate measurement of total serum 25-hydroxyvitamin D [25(OH)D] concentrations [i.e., 25(OH)D(3) and 25(OH)D(2)]. This study evaluated agreement between the ADVIA Centaur Vitamin D Total assay for 25(OH)D testing (traceable to the NIST-Ghent refe...
Autores principales: | , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Hindawi Publishing Corporation
2014
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4089903/ https://www.ncbi.nlm.nih.gov/pubmed/25045351 http://dx.doi.org/10.1155/2014/691679 |
_version_ | 1782325189633114112 |
---|---|
author | Freeman, James Wilson, Kimberly Spears, Ryan Shalhoub, Victoria Sibley, Paul |
author_facet | Freeman, James Wilson, Kimberly Spears, Ryan Shalhoub, Victoria Sibley, Paul |
author_sort | Freeman, James |
collection | PubMed |
description | Vitamin D status in different populations relies on accurate measurement of total serum 25-hydroxyvitamin D [25(OH)D] concentrations [i.e., 25(OH)D(3) and 25(OH)D(2)]. This study evaluated agreement between the ADVIA Centaur Vitamin D Total assay for 25(OH)D testing (traceable to the NIST-Ghent reference method procedure) and a liquid chromatography tandem mass spectrometry (LC-MS/MS) method for various populations with different levels of vitamin D binding protein (DBP). Total serum 25(OH)D concentrations were measured for 36 pregnant women, 40 hemodialysis patients, and 30 samples (DBP-spiked or not) from healthy subjects. ELISA measured DBP levels. The mean serum DBP concentrations were higher for pregnancy (415 μg/mL) and lower for hemodialysis subjects (198 μg/mL) than for healthy subjects and were highest for spiked serum (545 μg/mL). The average bias between the ADVIA Centaur assay and the LC-MS/MS method was −1.4% (healthy), −6.1% (pregnancy), and 4.4% (hemodialysis). The slightly greater bias for samples from some pregnancy and hemodialysis subjects with serum DBP levels outside of the normal healthy range fell within a clinically acceptable range—reflected by analysis of their low-range (≤136 μg/mL), medium-range (137–559 μg/mL), and high-range (≥560 μg/mL) DBP groups. Thus, the ADVIA Centaur Vitamin D Total assay demonstrates acceptable performance compared with an LC-MS/MS method for populations containing different amounts of DBP. |
format | Online Article Text |
id | pubmed-4089903 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2014 |
publisher | Hindawi Publishing Corporation |
record_format | MEDLINE/PubMed |
spelling | pubmed-40899032014-07-20 Influence of Vitamin D Binding Protein on Accuracy of 25-Hydroxyvitamin D Measurement Using the ADVIA Centaur Vitamin D Total Assay Freeman, James Wilson, Kimberly Spears, Ryan Shalhoub, Victoria Sibley, Paul Int J Endocrinol Research Article Vitamin D status in different populations relies on accurate measurement of total serum 25-hydroxyvitamin D [25(OH)D] concentrations [i.e., 25(OH)D(3) and 25(OH)D(2)]. This study evaluated agreement between the ADVIA Centaur Vitamin D Total assay for 25(OH)D testing (traceable to the NIST-Ghent reference method procedure) and a liquid chromatography tandem mass spectrometry (LC-MS/MS) method for various populations with different levels of vitamin D binding protein (DBP). Total serum 25(OH)D concentrations were measured for 36 pregnant women, 40 hemodialysis patients, and 30 samples (DBP-spiked or not) from healthy subjects. ELISA measured DBP levels. The mean serum DBP concentrations were higher for pregnancy (415 μg/mL) and lower for hemodialysis subjects (198 μg/mL) than for healthy subjects and were highest for spiked serum (545 μg/mL). The average bias between the ADVIA Centaur assay and the LC-MS/MS method was −1.4% (healthy), −6.1% (pregnancy), and 4.4% (hemodialysis). The slightly greater bias for samples from some pregnancy and hemodialysis subjects with serum DBP levels outside of the normal healthy range fell within a clinically acceptable range—reflected by analysis of their low-range (≤136 μg/mL), medium-range (137–559 μg/mL), and high-range (≥560 μg/mL) DBP groups. Thus, the ADVIA Centaur Vitamin D Total assay demonstrates acceptable performance compared with an LC-MS/MS method for populations containing different amounts of DBP. Hindawi Publishing Corporation 2014 2014-06-19 /pmc/articles/PMC4089903/ /pubmed/25045351 http://dx.doi.org/10.1155/2014/691679 Text en Copyright © 2014 James Freeman et al. https://creativecommons.org/licenses/by/3.0/ This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Research Article Freeman, James Wilson, Kimberly Spears, Ryan Shalhoub, Victoria Sibley, Paul Influence of Vitamin D Binding Protein on Accuracy of 25-Hydroxyvitamin D Measurement Using the ADVIA Centaur Vitamin D Total Assay |
title | Influence of Vitamin D Binding Protein on Accuracy of 25-Hydroxyvitamin D Measurement Using the ADVIA Centaur Vitamin D Total Assay |
title_full | Influence of Vitamin D Binding Protein on Accuracy of 25-Hydroxyvitamin D Measurement Using the ADVIA Centaur Vitamin D Total Assay |
title_fullStr | Influence of Vitamin D Binding Protein on Accuracy of 25-Hydroxyvitamin D Measurement Using the ADVIA Centaur Vitamin D Total Assay |
title_full_unstemmed | Influence of Vitamin D Binding Protein on Accuracy of 25-Hydroxyvitamin D Measurement Using the ADVIA Centaur Vitamin D Total Assay |
title_short | Influence of Vitamin D Binding Protein on Accuracy of 25-Hydroxyvitamin D Measurement Using the ADVIA Centaur Vitamin D Total Assay |
title_sort | influence of vitamin d binding protein on accuracy of 25-hydroxyvitamin d measurement using the advia centaur vitamin d total assay |
topic | Research Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4089903/ https://www.ncbi.nlm.nih.gov/pubmed/25045351 http://dx.doi.org/10.1155/2014/691679 |
work_keys_str_mv | AT freemanjames influenceofvitamindbindingproteinonaccuracyof25hydroxyvitamindmeasurementusingtheadviacentaurvitamindtotalassay AT wilsonkimberly influenceofvitamindbindingproteinonaccuracyof25hydroxyvitamindmeasurementusingtheadviacentaurvitamindtotalassay AT spearsryan influenceofvitamindbindingproteinonaccuracyof25hydroxyvitamindmeasurementusingtheadviacentaurvitamindtotalassay AT shalhoubvictoria influenceofvitamindbindingproteinonaccuracyof25hydroxyvitamindmeasurementusingtheadviacentaurvitamindtotalassay AT sibleypaul influenceofvitamindbindingproteinonaccuracyof25hydroxyvitamindmeasurementusingtheadviacentaurvitamindtotalassay |