Influence of Vitamin D Binding Protein on Accuracy of 25-Hydroxyvitamin D Measurement Using the ADVIA Centaur Vitamin D Total Assay

Vitamin D status in different populations relies on accurate measurement of total serum 25-hydroxyvitamin D [25(OH)D] concentrations [i.e., 25(OH)D(3) and 25(OH)D(2)]. This study evaluated agreement between the ADVIA Centaur Vitamin D Total assay for 25(OH)D testing (traceable to the NIST-Ghent refe...

Descripción completa

Detalles Bibliográficos
Autores principales: Freeman, James, Wilson, Kimberly, Spears, Ryan, Shalhoub, Victoria, Sibley, Paul
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Hindawi Publishing Corporation 2014
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4089903/
https://www.ncbi.nlm.nih.gov/pubmed/25045351
http://dx.doi.org/10.1155/2014/691679
_version_ 1782325189633114112
author Freeman, James
Wilson, Kimberly
Spears, Ryan
Shalhoub, Victoria
Sibley, Paul
author_facet Freeman, James
Wilson, Kimberly
Spears, Ryan
Shalhoub, Victoria
Sibley, Paul
author_sort Freeman, James
collection PubMed
description Vitamin D status in different populations relies on accurate measurement of total serum 25-hydroxyvitamin D [25(OH)D] concentrations [i.e., 25(OH)D(3) and 25(OH)D(2)]. This study evaluated agreement between the ADVIA Centaur Vitamin D Total assay for 25(OH)D testing (traceable to the NIST-Ghent reference method procedure) and a liquid chromatography tandem mass spectrometry (LC-MS/MS) method for various populations with different levels of vitamin D binding protein (DBP). Total serum 25(OH)D concentrations were measured for 36 pregnant women, 40 hemodialysis patients, and 30 samples (DBP-spiked or not) from healthy subjects. ELISA measured DBP levels. The mean serum DBP concentrations were higher for pregnancy (415 μg/mL) and lower for hemodialysis subjects (198 μg/mL) than for healthy subjects and were highest for spiked serum (545 μg/mL). The average bias between the ADVIA Centaur assay and the LC-MS/MS method was −1.4% (healthy), −6.1% (pregnancy), and 4.4% (hemodialysis). The slightly greater bias for samples from some pregnancy and hemodialysis subjects with serum DBP levels outside of the normal healthy range fell within a clinically acceptable range—reflected by analysis of their low-range (≤136 μg/mL), medium-range (137–559 μg/mL), and high-range (≥560 μg/mL) DBP groups. Thus, the ADVIA Centaur Vitamin D Total assay demonstrates acceptable performance compared with an LC-MS/MS method for populations containing different amounts of DBP.
format Online
Article
Text
id pubmed-4089903
institution National Center for Biotechnology Information
language English
publishDate 2014
publisher Hindawi Publishing Corporation
record_format MEDLINE/PubMed
spelling pubmed-40899032014-07-20 Influence of Vitamin D Binding Protein on Accuracy of 25-Hydroxyvitamin D Measurement Using the ADVIA Centaur Vitamin D Total Assay Freeman, James Wilson, Kimberly Spears, Ryan Shalhoub, Victoria Sibley, Paul Int J Endocrinol Research Article Vitamin D status in different populations relies on accurate measurement of total serum 25-hydroxyvitamin D [25(OH)D] concentrations [i.e., 25(OH)D(3) and 25(OH)D(2)]. This study evaluated agreement between the ADVIA Centaur Vitamin D Total assay for 25(OH)D testing (traceable to the NIST-Ghent reference method procedure) and a liquid chromatography tandem mass spectrometry (LC-MS/MS) method for various populations with different levels of vitamin D binding protein (DBP). Total serum 25(OH)D concentrations were measured for 36 pregnant women, 40 hemodialysis patients, and 30 samples (DBP-spiked or not) from healthy subjects. ELISA measured DBP levels. The mean serum DBP concentrations were higher for pregnancy (415 μg/mL) and lower for hemodialysis subjects (198 μg/mL) than for healthy subjects and were highest for spiked serum (545 μg/mL). The average bias between the ADVIA Centaur assay and the LC-MS/MS method was −1.4% (healthy), −6.1% (pregnancy), and 4.4% (hemodialysis). The slightly greater bias for samples from some pregnancy and hemodialysis subjects with serum DBP levels outside of the normal healthy range fell within a clinically acceptable range—reflected by analysis of their low-range (≤136 μg/mL), medium-range (137–559 μg/mL), and high-range (≥560 μg/mL) DBP groups. Thus, the ADVIA Centaur Vitamin D Total assay demonstrates acceptable performance compared with an LC-MS/MS method for populations containing different amounts of DBP. Hindawi Publishing Corporation 2014 2014-06-19 /pmc/articles/PMC4089903/ /pubmed/25045351 http://dx.doi.org/10.1155/2014/691679 Text en Copyright © 2014 James Freeman et al. https://creativecommons.org/licenses/by/3.0/ This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Research Article
Freeman, James
Wilson, Kimberly
Spears, Ryan
Shalhoub, Victoria
Sibley, Paul
Influence of Vitamin D Binding Protein on Accuracy of 25-Hydroxyvitamin D Measurement Using the ADVIA Centaur Vitamin D Total Assay
title Influence of Vitamin D Binding Protein on Accuracy of 25-Hydroxyvitamin D Measurement Using the ADVIA Centaur Vitamin D Total Assay
title_full Influence of Vitamin D Binding Protein on Accuracy of 25-Hydroxyvitamin D Measurement Using the ADVIA Centaur Vitamin D Total Assay
title_fullStr Influence of Vitamin D Binding Protein on Accuracy of 25-Hydroxyvitamin D Measurement Using the ADVIA Centaur Vitamin D Total Assay
title_full_unstemmed Influence of Vitamin D Binding Protein on Accuracy of 25-Hydroxyvitamin D Measurement Using the ADVIA Centaur Vitamin D Total Assay
title_short Influence of Vitamin D Binding Protein on Accuracy of 25-Hydroxyvitamin D Measurement Using the ADVIA Centaur Vitamin D Total Assay
title_sort influence of vitamin d binding protein on accuracy of 25-hydroxyvitamin d measurement using the advia centaur vitamin d total assay
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4089903/
https://www.ncbi.nlm.nih.gov/pubmed/25045351
http://dx.doi.org/10.1155/2014/691679
work_keys_str_mv AT freemanjames influenceofvitamindbindingproteinonaccuracyof25hydroxyvitamindmeasurementusingtheadviacentaurvitamindtotalassay
AT wilsonkimberly influenceofvitamindbindingproteinonaccuracyof25hydroxyvitamindmeasurementusingtheadviacentaurvitamindtotalassay
AT spearsryan influenceofvitamindbindingproteinonaccuracyof25hydroxyvitamindmeasurementusingtheadviacentaurvitamindtotalassay
AT shalhoubvictoria influenceofvitamindbindingproteinonaccuracyof25hydroxyvitamindmeasurementusingtheadviacentaurvitamindtotalassay
AT sibleypaul influenceofvitamindbindingproteinonaccuracyof25hydroxyvitamindmeasurementusingtheadviacentaurvitamindtotalassay