Cargando…

Construct Failure in an Atypical Femoral Fracture treated with Intramedullary Nailing: A Case Report

The aim of this paper is to document a rare case of construct failure in a 68-year old Filipina who sustained an atypical femoral fracture (AFF) in her left subtrochanteric area. The patient previously had a 40-month history of alendronate 70mg + vitamin D 5600u therapy for osteoporosis and underwen...

Descripción completa

Detalles Bibliográficos
Autores principales: Bonifacio, L, Syson, P
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Malaysian Orthopaedic Association 2014
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4093543/
https://www.ncbi.nlm.nih.gov/pubmed/25279084
http://dx.doi.org/10.5704/MOJ.1403.008
Descripción
Sumario:The aim of this paper is to document a rare case of construct failure in a 68-year old Filipina who sustained an atypical femoral fracture (AFF) in her left subtrochanteric area. The patient previously had a 40-month history of alendronate 70mg + vitamin D 5600u therapy for osteoporosis and underwent closed intramedullary nailing for the AFF. Six months postoperatively, she began to experience progressive pain in her operated thigh. Radiographs revealed a broken nail at the proximal screw hole and non-union of the AFF. The patient was treated with exposure of the fracture site, removal of the broken device, exchange intramedullary nailing, and iliac bone grafting. She had radiographic and clinical union and was full weight bearing after three months.