Cargando…

Combined use of covered stent and flow diversion to seal iatrogenic carotid injury with vessel preservation during transsphenoidal endoscopic resection of clival tumor

BACKGROUND: Transsphenoidal tumor resection can lead to internal carotid artery (ICA) injury. Vascular disruption is often treated with emergent vessel deconstruction, incurring complications in a subset of patients with poor collateral circulation and resulting in minor and major ischemic strokes....

Descripción completa

Detalles Bibliográficos
Autores principales: Shakir, Hakeem J., Garson, Alex D., Sorkin, Grant C., Mokin, Maxim, Eller, Jorge L., Dumont, Travis M., Popat, Saurin R., Leonardo, Jody, Siddiqui, Adnan H.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Medknow Publications & Media Pvt Ltd 2014
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4093738/
https://www.ncbi.nlm.nih.gov/pubmed/25024881
http://dx.doi.org/10.4103/2152-7806.133638
Descripción
Sumario:BACKGROUND: Transsphenoidal tumor resection can lead to internal carotid artery (ICA) injury. Vascular disruption is often treated with emergent vessel deconstruction, incurring complications in a subset of patients with poor collateral circulation and resulting in minor and major ischemic strokes. METHODS: We attempted a novel approach combining a covered stent graft (Jostent) and two flow diverter stents [Pipeline embolization devices (PEDs)] to treat active extravasation from a disrupted right ICA that was the result of a transsphenoidal surgery complication. This disruption occurred during clival tumor surgery and required immediate sphenoidal sinus packing. Emergent angiography revealed continued petrous carotid artery extravasation, warranting emergent vessel repair or deconstruction for treatment. To preserve the vessel, we utilized a covered Jostent. Due to tortuosity and lack of optimal wall apposition, there was reduced, yet persistent extravasation from an endoleak after Jostent deployment that failed to resolve despite multiple angioplasties. Therefore, we used PEDs to divert the flow. RESULTS: Flow diversion relieved the extravasation. The patient remained neurologically intact post-procedure. CONCLUSIONS: This case demonstrates successful combined use of a covered stent and flow diverters to treat acute vascular injury resulting from transsphenoidal surgery. However, concerns remain, including the requirement of dual antiplatelet agents increasing postoperative bleeding risks, stent-related thromboembolic events, and delayed in-stent restenosis rates.