PET reconstruction artifact can be minimized by using sinogram correction and filtered back-projection technique

Filtered Back-Projection (FBP) has become an outdated image reconstruction technique in new-generation positron emission tomography (PET)/computed tomography (CT) scanners. Iterative reconstruction used in all new-generation PET scanners is a much improved reconstruction technique. Though a well-cal...

Descripción completa

Detalles Bibliográficos
Autores principales: Jha, Ashish Kumar, Purandare, Nilendu C, Shah, Sneha, Agrawal, Archi, Puranik, Ameya D, Rangarajan, Venkatesh
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Medknow Publications & Media Pvt Ltd 2014
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4094957/
https://www.ncbi.nlm.nih.gov/pubmed/25024515
http://dx.doi.org/10.4103/0971-3026.134379
Descripción
Sumario:Filtered Back-Projection (FBP) has become an outdated image reconstruction technique in new-generation positron emission tomography (PET)/computed tomography (CT) scanners. Iterative reconstruction used in all new-generation PET scanners is a much improved reconstruction technique. Though a well-calibrated PET system can only be used for clinical imaging in few situations like ours, when compromised PET scanner with one PET module bypassed was used for PET acquisition, FBP with sinogram correction proved to be a better reconstruction technique to minimize streak artifact present in the image reconstructed by the iterative technique.