Cargando…

On Interlayer Stability and High-Cycle Simulator Performance of Diamond-Like Carbon Layers for Articulating Joint Replacements

Diamond like carbon (DLC) coatings have been proven to be an excellent choice for wear reduction in many technical applications. However, for successful adaption to the orthopaedic field, layer performance, stability and adhesion in physiologically relevant setups are crucial and not consistently in...

Descripción completa

Detalles Bibliográficos
Autores principales: Thorwarth, Kerstin, Thorwarth, Götz, Figi, Renato, Weisse, Bernhard, Stiefel, Michael, Hauert, Roland
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2014
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4100166/
https://www.ncbi.nlm.nih.gov/pubmed/24921709
http://dx.doi.org/10.3390/ijms150610527
_version_ 1782326626240954368
author Thorwarth, Kerstin
Thorwarth, Götz
Figi, Renato
Weisse, Bernhard
Stiefel, Michael
Hauert, Roland
author_facet Thorwarth, Kerstin
Thorwarth, Götz
Figi, Renato
Weisse, Bernhard
Stiefel, Michael
Hauert, Roland
author_sort Thorwarth, Kerstin
collection PubMed
description Diamond like carbon (DLC) coatings have been proven to be an excellent choice for wear reduction in many technical applications. However, for successful adaption to the orthopaedic field, layer performance, stability and adhesion in physiologically relevant setups are crucial and not consistently investigated. In vitro wear testing as well as adequate corrosion tests of interfaces and interlayers are of great importance to verify the long term stability of DLC coated load bearing implants in the human body. DLC coatings were deposited on articulating lumbar spinal disks made of CoCr28Mo6 biomedical implant alloy using a plasma-activated chemical vapor deposition (PACVD) process. As an adhesion promoting interlayer, tantalum films were deposited by magnetron sputtering. Wear tests of coated and uncoated implants were performed in physiological solution up to a maximum of 101 million articulation cycles with an amplitude of ±2° and −3/+6° in successive intervals at a preload of 1200 N. The implants were characterized by gravimetry, inductively coupled plasma optical emission spectrometry (ICP-OES) and cross section scanning electron microscopy (SEM) analysis. It is shown that DLC coated surfaces with uncontaminated tantalum interlayers perform very well and no corrosive or mechanical failure could be observed. This also holds true in tests featuring overload and third-body wear by cortical bone chips present in the bearing pairs. Regarding the interlayer tolerance towards interlayer contamination (oxygen), limits for initiation of potential failure modes were established. It was found that mechanical failure is the most critical aspect and this mode is hypothetically linked to the α-β tantalum phase switch induced by increasing oxygen levels as observed by X-ray diffraction (XRD). It is concluded that DLC coatings are a feasible candidate for near zero wear articulations on implants, potentially even surpassing the performance of ceramic vs. ceramic.
format Online
Article
Text
id pubmed-4100166
institution National Center for Biotechnology Information
language English
publishDate 2014
publisher MDPI
record_format MEDLINE/PubMed
spelling pubmed-41001662014-07-16 On Interlayer Stability and High-Cycle Simulator Performance of Diamond-Like Carbon Layers for Articulating Joint Replacements Thorwarth, Kerstin Thorwarth, Götz Figi, Renato Weisse, Bernhard Stiefel, Michael Hauert, Roland Int J Mol Sci Article Diamond like carbon (DLC) coatings have been proven to be an excellent choice for wear reduction in many technical applications. However, for successful adaption to the orthopaedic field, layer performance, stability and adhesion in physiologically relevant setups are crucial and not consistently investigated. In vitro wear testing as well as adequate corrosion tests of interfaces and interlayers are of great importance to verify the long term stability of DLC coated load bearing implants in the human body. DLC coatings were deposited on articulating lumbar spinal disks made of CoCr28Mo6 biomedical implant alloy using a plasma-activated chemical vapor deposition (PACVD) process. As an adhesion promoting interlayer, tantalum films were deposited by magnetron sputtering. Wear tests of coated and uncoated implants were performed in physiological solution up to a maximum of 101 million articulation cycles with an amplitude of ±2° and −3/+6° in successive intervals at a preload of 1200 N. The implants were characterized by gravimetry, inductively coupled plasma optical emission spectrometry (ICP-OES) and cross section scanning electron microscopy (SEM) analysis. It is shown that DLC coated surfaces with uncontaminated tantalum interlayers perform very well and no corrosive or mechanical failure could be observed. This also holds true in tests featuring overload and third-body wear by cortical bone chips present in the bearing pairs. Regarding the interlayer tolerance towards interlayer contamination (oxygen), limits for initiation of potential failure modes were established. It was found that mechanical failure is the most critical aspect and this mode is hypothetically linked to the α-β tantalum phase switch induced by increasing oxygen levels as observed by X-ray diffraction (XRD). It is concluded that DLC coatings are a feasible candidate for near zero wear articulations on implants, potentially even surpassing the performance of ceramic vs. ceramic. MDPI 2014-06-11 /pmc/articles/PMC4100166/ /pubmed/24921709 http://dx.doi.org/10.3390/ijms150610527 Text en © 2014 by the authors; licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution license (http://creativecommons.org/licenses/by/3.0/).
spellingShingle Article
Thorwarth, Kerstin
Thorwarth, Götz
Figi, Renato
Weisse, Bernhard
Stiefel, Michael
Hauert, Roland
On Interlayer Stability and High-Cycle Simulator Performance of Diamond-Like Carbon Layers for Articulating Joint Replacements
title On Interlayer Stability and High-Cycle Simulator Performance of Diamond-Like Carbon Layers for Articulating Joint Replacements
title_full On Interlayer Stability and High-Cycle Simulator Performance of Diamond-Like Carbon Layers for Articulating Joint Replacements
title_fullStr On Interlayer Stability and High-Cycle Simulator Performance of Diamond-Like Carbon Layers for Articulating Joint Replacements
title_full_unstemmed On Interlayer Stability and High-Cycle Simulator Performance of Diamond-Like Carbon Layers for Articulating Joint Replacements
title_short On Interlayer Stability and High-Cycle Simulator Performance of Diamond-Like Carbon Layers for Articulating Joint Replacements
title_sort on interlayer stability and high-cycle simulator performance of diamond-like carbon layers for articulating joint replacements
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4100166/
https://www.ncbi.nlm.nih.gov/pubmed/24921709
http://dx.doi.org/10.3390/ijms150610527
work_keys_str_mv AT thorwarthkerstin oninterlayerstabilityandhighcyclesimulatorperformanceofdiamondlikecarbonlayersforarticulatingjointreplacements
AT thorwarthgotz oninterlayerstabilityandhighcyclesimulatorperformanceofdiamondlikecarbonlayersforarticulatingjointreplacements
AT figirenato oninterlayerstabilityandhighcyclesimulatorperformanceofdiamondlikecarbonlayersforarticulatingjointreplacements
AT weissebernhard oninterlayerstabilityandhighcyclesimulatorperformanceofdiamondlikecarbonlayersforarticulatingjointreplacements
AT stiefelmichael oninterlayerstabilityandhighcyclesimulatorperformanceofdiamondlikecarbonlayersforarticulatingjointreplacements
AT hauertroland oninterlayerstabilityandhighcyclesimulatorperformanceofdiamondlikecarbonlayersforarticulatingjointreplacements