Are the review criteria for automated complete blood counts of the International Society of Laboratory Hematology suitable for all hematology laboratories?

OBJECTIVE: to verify whether the review criteria for automated blood counts suggested by the International Consensus Group for Hematology Review of the International Society for Laboratory Hematology are suitable for the Hematology Laboratory of Hospital de Clinicas, Universidade Federal do Paraná....

Descripción completa

Detalles Bibliográficos
Autores principales: Comar, Samuel Ricardo, Malvezzi, Mariester, Pasquini, Ricardo
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Sociedade Brasileira de Hematologia e Hemoterapia 2014
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4109747/
https://www.ncbi.nlm.nih.gov/pubmed/25031063
http://dx.doi.org/10.1016/j.bjhh.2014.03.011
_version_ 1782327907176153088
author Comar, Samuel Ricardo
Malvezzi, Mariester
Pasquini, Ricardo
author_facet Comar, Samuel Ricardo
Malvezzi, Mariester
Pasquini, Ricardo
author_sort Comar, Samuel Ricardo
collection PubMed
description OBJECTIVE: to verify whether the review criteria for automated blood counts suggested by the International Consensus Group for Hematology Review of the International Society for Laboratory Hematology are suitable for the Hematology Laboratory of Hospital de Clinicas, Universidade Federal do Paraná. METHODS: initially, the review criteria of the International Society for Laboratory Hematology were adapted due to limitations in the Institution's electronic hospital records and interfacing systems. The adapted review criteria were tested using 1977 samples. After this first assessment, an additional 180 inpatient samples were analyzed to evaluate the screening criteria of the review criteria in conjunction with positive smear findings established by the institution. The performance of the review criteria was verified by determining false positive, false negative, true positive and true negative rates, sensitivity, specificity, positive predictive value, negative predictive value, microscopic review rate and efficiency. RESULTS: initial analysis showed false negatives = 6.73%, false positives = 23.27%, microscopic review rate = 46.03% and efficiency = 70.0%. An evaluation of the screening criteria adapted from the review criteria together with the positive smear findings of the institution showed false negatives = 15.5%, false positives = 10.5%, microscopic review rate = 37.3% and efficiency = 73.8%. In both situations the safety limit (false negative <5%) recommended by the review criteria was exceeded. CONCLUSIONS: the review criteria adapted from the International Society for Laboratory Hematology are neither suitable nor safe for use in the hematology laboratory of the Hospital de Clinicas. This implies a need to develop and validate institution-specific review criteria in order to decrease false negative results to an acceptable and safe rate for patients.
format Online
Article
Text
id pubmed-4109747
institution National Center for Biotechnology Information
language English
publishDate 2014
publisher Sociedade Brasileira de Hematologia e Hemoterapia
record_format MEDLINE/PubMed
spelling pubmed-41097472014-07-25 Are the review criteria for automated complete blood counts of the International Society of Laboratory Hematology suitable for all hematology laboratories? Comar, Samuel Ricardo Malvezzi, Mariester Pasquini, Ricardo Rev Bras Hematol Hemoter Update Article OBJECTIVE: to verify whether the review criteria for automated blood counts suggested by the International Consensus Group for Hematology Review of the International Society for Laboratory Hematology are suitable for the Hematology Laboratory of Hospital de Clinicas, Universidade Federal do Paraná. METHODS: initially, the review criteria of the International Society for Laboratory Hematology were adapted due to limitations in the Institution's electronic hospital records and interfacing systems. The adapted review criteria were tested using 1977 samples. After this first assessment, an additional 180 inpatient samples were analyzed to evaluate the screening criteria of the review criteria in conjunction with positive smear findings established by the institution. The performance of the review criteria was verified by determining false positive, false negative, true positive and true negative rates, sensitivity, specificity, positive predictive value, negative predictive value, microscopic review rate and efficiency. RESULTS: initial analysis showed false negatives = 6.73%, false positives = 23.27%, microscopic review rate = 46.03% and efficiency = 70.0%. An evaluation of the screening criteria adapted from the review criteria together with the positive smear findings of the institution showed false negatives = 15.5%, false positives = 10.5%, microscopic review rate = 37.3% and efficiency = 73.8%. In both situations the safety limit (false negative <5%) recommended by the review criteria was exceeded. CONCLUSIONS: the review criteria adapted from the International Society for Laboratory Hematology are neither suitable nor safe for use in the hematology laboratory of the Hospital de Clinicas. This implies a need to develop and validate institution-specific review criteria in order to decrease false negative results to an acceptable and safe rate for patients. Sociedade Brasileira de Hematologia e Hemoterapia 2014 2014-04-03 /pmc/articles/PMC4109747/ /pubmed/25031063 http://dx.doi.org/10.1016/j.bjhh.2014.03.011 Text en © 2014 Associação Brasileira de Hematologia, Hemoterapia e Terapia Celular. Published by Elsevier Editora Ltda. All rights reserved. http://creativecommons.org/licenses/by-nc-nd/3.0/ This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/3.0/).
spellingShingle Update Article
Comar, Samuel Ricardo
Malvezzi, Mariester
Pasquini, Ricardo
Are the review criteria for automated complete blood counts of the International Society of Laboratory Hematology suitable for all hematology laboratories?
title Are the review criteria for automated complete blood counts of the International Society of Laboratory Hematology suitable for all hematology laboratories?
title_full Are the review criteria for automated complete blood counts of the International Society of Laboratory Hematology suitable for all hematology laboratories?
title_fullStr Are the review criteria for automated complete blood counts of the International Society of Laboratory Hematology suitable for all hematology laboratories?
title_full_unstemmed Are the review criteria for automated complete blood counts of the International Society of Laboratory Hematology suitable for all hematology laboratories?
title_short Are the review criteria for automated complete blood counts of the International Society of Laboratory Hematology suitable for all hematology laboratories?
title_sort are the review criteria for automated complete blood counts of the international society of laboratory hematology suitable for all hematology laboratories?
topic Update Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4109747/
https://www.ncbi.nlm.nih.gov/pubmed/25031063
http://dx.doi.org/10.1016/j.bjhh.2014.03.011
work_keys_str_mv AT comarsamuelricardo arethereviewcriteriaforautomatedcompletebloodcountsoftheinternationalsocietyoflaboratoryhematologysuitableforallhematologylaboratories
AT malvezzimariester arethereviewcriteriaforautomatedcompletebloodcountsoftheinternationalsocietyoflaboratoryhematologysuitableforallhematologylaboratories
AT pasquiniricardo arethereviewcriteriaforautomatedcompletebloodcountsoftheinternationalsocietyoflaboratoryhematologysuitableforallhematologylaboratories