Cargando…
Cerebral oxygenation after birth – a comparison of INVOS(®) and FORE‐SIGHT™ near‐infrared spectroscopy oximeters
AIM: To compare absolute values of regional cerebral tissue oxygenation (cStO(2)) during haemodynamic transition after birth and repeatability during steady state for two commercial near‐infrared spectroscopy (NIRS) devices. METHODS: In a prospective observational study, the INVOS 5100C and FORE‐SIG...
Autores principales: | , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Wiley-Blackwell
2014
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4112844/ https://www.ncbi.nlm.nih.gov/pubmed/24456266 http://dx.doi.org/10.1111/apa.12567 |
Sumario: | AIM: To compare absolute values of regional cerebral tissue oxygenation (cStO(2)) during haemodynamic transition after birth and repeatability during steady state for two commercial near‐infrared spectroscopy (NIRS) devices. METHODS: In a prospective observational study, the INVOS 5100C and FORE‐SIGHT were compared on 12 term newborns delivered by elective caesarean section. During the 10 min following umbilical cord clamping, cStO(2) was measured simultaneously with the neonatal sensors from each device. Repeated measurements were taken the following day. RESULTS: Three and 8 min after clamping, the mean cStO(2) value increased from 53.4% (CI 36.8–69.9%) to 86.0% (CI 80.2–91.7%) for INVOS and from 61.6% (CI 55.4–67.8%) to 82.2% (CI 77.7–86.7%) for FORE‐SIGHT. The Bland–Altman plot revealed decreasing difference (INVOS minus FORE‐SIGHT) (D) in absolute values (A) with increasing cStO2 (D = 0.5A – 38.19 p = <0.001). The mean steady‐state value on day two was 78.4% (CI 74.6–82.2%) and 86.2% (CI 85.0–87.4%) for INVOS and FORE‐SIGHT, respectively. The within‐subject standard deviation during steady‐state repeated measurements was 4.8% ± 0.86 for INVOS and 2.8% ± 0.5 for FORE‐SIGHT. CONCLUSION: The INVOS and FORE‐SIGHT cStO(2) estimates showed oxygenation‐level‐dependent difference during birth transition. The better repeatability of FORE‐SIGHT could be due to the lower response to change in saturation. |
---|