Cargando…

Inter-Vendor Reproducibility of Pseudo-Continuous Arterial Spin Labeling at 3 Tesla

PURPOSE: Prior to the implementation of arterial spin labeling (ASL) in clinical multi-center studies, it is important to establish its status quo inter-vendor reproducibility. This study evaluates and compares the intra- and inter-vendor reproducibility of pseudo-continuous ASL (pCASL) as clinicall...

Descripción completa

Detalles Bibliográficos
Autores principales: Mutsaerts, Henri J. M. M., Steketee, Rebecca M. E., Heijtel, Dennis F. R., Kuijer, Joost P. A., van Osch, Matthias J. P., Majoie, Charles B. L. M., Smits, Marion, Nederveen, Aart J.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Public Library of Science 2014
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4121318/
https://www.ncbi.nlm.nih.gov/pubmed/25090654
http://dx.doi.org/10.1371/journal.pone.0104108
Descripción
Sumario:PURPOSE: Prior to the implementation of arterial spin labeling (ASL) in clinical multi-center studies, it is important to establish its status quo inter-vendor reproducibility. This study evaluates and compares the intra- and inter-vendor reproducibility of pseudo-continuous ASL (pCASL) as clinically implemented by GE and Philips. MATERIAL AND METHODS: 22 healthy volunteers were scanned twice on both a 3T GE and a 3T Philips scanner. The main difference in implementation between the vendors was the readout module: spiral 3D fast spin echo vs. 2D gradient-echo echo-planar imaging respectively. Mean and variation of cerebral blood flow (CBF) were compared for the total gray matter (GM) and white matter (WM), and on a voxel-level. RESULTS: Whereas the mean GM CBF of both vendors was almost equal (p = 1.0), the mean WM CBF was significantly different (p<0.01). The inter-vendor GM variation did not differ from the intra-vendor GM variation (p = 0.3 and p = 0.5 for GE and Philips respectively). Spatial inter-vendor CBF and variation differences were observed in several GM regions and in the WM. CONCLUSION: These results show that total GM CBF-values can be exchanged between vendors. For the inter-vendor comparison of GM regions or WM, these results encourage further standardization of ASL implementation among vendors.