Definition of Metrics to Evaluate Cochlear Array Insertion Forces Performed with Forceps, Insertion Tool, or Motorized Tool in Temporal Bone Specimens

Introduction. In order to achieve a minimal trauma to the inner ear structures during array insertion, it would be suitable to control insertion forces. The aim of this work was to compare the insertion forces of an array insertion into anatomical specimens with three different insertion techniques:...

Descripción completa

Detalles Bibliográficos
Autores principales: Nguyen, Yann, Kazmitcheff, Guillaume, De Seta, Daniele, Miroir, Mathieu, Ferrary, Evelyne, Sterkers, Olivier
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Hindawi Publishing Corporation 2014
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4122132/
https://www.ncbi.nlm.nih.gov/pubmed/25126565
http://dx.doi.org/10.1155/2014/532570
_version_ 1782329309940154368
author Nguyen, Yann
Kazmitcheff, Guillaume
De Seta, Daniele
Miroir, Mathieu
Ferrary, Evelyne
Sterkers, Olivier
author_facet Nguyen, Yann
Kazmitcheff, Guillaume
De Seta, Daniele
Miroir, Mathieu
Ferrary, Evelyne
Sterkers, Olivier
author_sort Nguyen, Yann
collection PubMed
description Introduction. In order to achieve a minimal trauma to the inner ear structures during array insertion, it would be suitable to control insertion forces. The aim of this work was to compare the insertion forces of an array insertion into anatomical specimens with three different insertion techniques: with forceps, with a commercial tool, and with a motorized tool. Materials and Methods. Temporal bones have been mounted on a 6-axis force sensor to record insertion forces. Each temporal bone has been inserted, with a lateral wall electrode array, in random order, with each of the 3 techniques. Results. Forceps manual and commercial tool insertions generated multiple jerks during whole length insertion related to fits and starts. On the contrary, insertion force with the motorized tool only rose at the end of the insertion. Overall force momentum was 1.16 ± 0.505 N (mean ± SD, n = 10), 1.337 ± 0.408 N (n = 8), and 1.573 ± 0.764 N (n = 8) for manual insertion with forceps and commercial and motorized tools, respectively. Conclusion. Considering force momentum, no difference between the three techniques was observed. Nevertheless, a more predictable force profile could be observed with the motorized tool with a smoother rise of insertion forces.
format Online
Article
Text
id pubmed-4122132
institution National Center for Biotechnology Information
language English
publishDate 2014
publisher Hindawi Publishing Corporation
record_format MEDLINE/PubMed
spelling pubmed-41221322014-08-14 Definition of Metrics to Evaluate Cochlear Array Insertion Forces Performed with Forceps, Insertion Tool, or Motorized Tool in Temporal Bone Specimens Nguyen, Yann Kazmitcheff, Guillaume De Seta, Daniele Miroir, Mathieu Ferrary, Evelyne Sterkers, Olivier Biomed Res Int Research Article Introduction. In order to achieve a minimal trauma to the inner ear structures during array insertion, it would be suitable to control insertion forces. The aim of this work was to compare the insertion forces of an array insertion into anatomical specimens with three different insertion techniques: with forceps, with a commercial tool, and with a motorized tool. Materials and Methods. Temporal bones have been mounted on a 6-axis force sensor to record insertion forces. Each temporal bone has been inserted, with a lateral wall electrode array, in random order, with each of the 3 techniques. Results. Forceps manual and commercial tool insertions generated multiple jerks during whole length insertion related to fits and starts. On the contrary, insertion force with the motorized tool only rose at the end of the insertion. Overall force momentum was 1.16 ± 0.505 N (mean ± SD, n = 10), 1.337 ± 0.408 N (n = 8), and 1.573 ± 0.764 N (n = 8) for manual insertion with forceps and commercial and motorized tools, respectively. Conclusion. Considering force momentum, no difference between the three techniques was observed. Nevertheless, a more predictable force profile could be observed with the motorized tool with a smoother rise of insertion forces. Hindawi Publishing Corporation 2014 2014-07-15 /pmc/articles/PMC4122132/ /pubmed/25126565 http://dx.doi.org/10.1155/2014/532570 Text en Copyright © 2014 Yann Nguyen et al. https://creativecommons.org/licenses/by/3.0/ This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Research Article
Nguyen, Yann
Kazmitcheff, Guillaume
De Seta, Daniele
Miroir, Mathieu
Ferrary, Evelyne
Sterkers, Olivier
Definition of Metrics to Evaluate Cochlear Array Insertion Forces Performed with Forceps, Insertion Tool, or Motorized Tool in Temporal Bone Specimens
title Definition of Metrics to Evaluate Cochlear Array Insertion Forces Performed with Forceps, Insertion Tool, or Motorized Tool in Temporal Bone Specimens
title_full Definition of Metrics to Evaluate Cochlear Array Insertion Forces Performed with Forceps, Insertion Tool, or Motorized Tool in Temporal Bone Specimens
title_fullStr Definition of Metrics to Evaluate Cochlear Array Insertion Forces Performed with Forceps, Insertion Tool, or Motorized Tool in Temporal Bone Specimens
title_full_unstemmed Definition of Metrics to Evaluate Cochlear Array Insertion Forces Performed with Forceps, Insertion Tool, or Motorized Tool in Temporal Bone Specimens
title_short Definition of Metrics to Evaluate Cochlear Array Insertion Forces Performed with Forceps, Insertion Tool, or Motorized Tool in Temporal Bone Specimens
title_sort definition of metrics to evaluate cochlear array insertion forces performed with forceps, insertion tool, or motorized tool in temporal bone specimens
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4122132/
https://www.ncbi.nlm.nih.gov/pubmed/25126565
http://dx.doi.org/10.1155/2014/532570
work_keys_str_mv AT nguyenyann definitionofmetricstoevaluatecochleararrayinsertionforcesperformedwithforcepsinsertiontoolormotorizedtoolintemporalbonespecimens
AT kazmitcheffguillaume definitionofmetricstoevaluatecochleararrayinsertionforcesperformedwithforcepsinsertiontoolormotorizedtoolintemporalbonespecimens
AT desetadaniele definitionofmetricstoevaluatecochleararrayinsertionforcesperformedwithforcepsinsertiontoolormotorizedtoolintemporalbonespecimens
AT miroirmathieu definitionofmetricstoevaluatecochleararrayinsertionforcesperformedwithforcepsinsertiontoolormotorizedtoolintemporalbonespecimens
AT ferraryevelyne definitionofmetricstoevaluatecochleararrayinsertionforcesperformedwithforcepsinsertiontoolormotorizedtoolintemporalbonespecimens
AT sterkersolivier definitionofmetricstoevaluatecochleararrayinsertionforcesperformedwithforcepsinsertiontoolormotorizedtoolintemporalbonespecimens