Assessment of Two Portable Real-Time Particle Monitors Used in Nanomaterial Workplace Exposure Evaluations

BACKGROUND: Nanoparticle emission assessment technique was developed to semi-quantitatively evaluate nanomaterial exposures and employs a combination of filter based samples and portable real-time particle monitors, including a condensation particle counter (CPC) and an optical particle counter (OPC...

Descripción completa

Detalles Bibliográficos
Autores principales: Liu, Yuewei, Beaucham, Catherine C., Pearce, Terri A., Zhuang, Ziqing
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Public Library of Science 2014
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4141826/
https://www.ncbi.nlm.nih.gov/pubmed/25148239
http://dx.doi.org/10.1371/journal.pone.0105769
_version_ 1782331700466941952
author Liu, Yuewei
Beaucham, Catherine C.
Pearce, Terri A.
Zhuang, Ziqing
author_facet Liu, Yuewei
Beaucham, Catherine C.
Pearce, Terri A.
Zhuang, Ziqing
author_sort Liu, Yuewei
collection PubMed
description BACKGROUND: Nanoparticle emission assessment technique was developed to semi-quantitatively evaluate nanomaterial exposures and employs a combination of filter based samples and portable real-time particle monitors, including a condensation particle counter (CPC) and an optical particle counter (OPC), to detect nanomaterial releases. This laboratory study evaluated the results from CPC and OPC simultaneously measuring a polydisperse aerosol to assess their variability and accuracy. METHODS AND RESULTS: Two CPCs and two OPCs were used to evaluate a polydisperse sodium chloride aerosol within an enclosed chamber. The measurement results for number concentration versus time were compared between paired particle monitors of the same type, and to results from the Scanning Mobility Particle Spectrometer (SMPS) which was widely used to measure concentration of size-specific particles. According to analyses by using the Bland-Altman method, the CPCs displayed a constant mean percent difference of −3.8% (95% agreement limits: −9.1 to 1.6%; range of 95% agreement limit: 10.7%) with the chamber particle concentration below its dynamic upper limit (100,000 particles per cubic centimeter). The mean percent difference increased from −3.4% to −12.0% (range of 95% agreement limits: 7.1%) with increasing particle concentrations that were above the dynamic upper limit. The OPC results showed the percent difference within 15% for measurements in particles with size ranges of 300 to 500 and 500 to 1000 regardless of the particle concentration. Compared with SMPS measurements, the CPC gave a mean percent difference of 22.9% (95% agreement limits: 10.5% to 35.2%); whereas the measurements from OPC were not comparable. CONCLUSIONS: This study demonstrated that CPC and OPC are useful for measuring nanoparticle exposures but the results from an individual monitor should be interpreted based upon the instrument's technical parameters. Future research should challenge these monitors with particles of different sizes, shapes, or composition, to determine measurement comparability and accuracy across various workplace nanomaterials.
format Online
Article
Text
id pubmed-4141826
institution National Center for Biotechnology Information
language English
publishDate 2014
publisher Public Library of Science
record_format MEDLINE/PubMed
spelling pubmed-41418262014-08-25 Assessment of Two Portable Real-Time Particle Monitors Used in Nanomaterial Workplace Exposure Evaluations Liu, Yuewei Beaucham, Catherine C. Pearce, Terri A. Zhuang, Ziqing PLoS One Research Article BACKGROUND: Nanoparticle emission assessment technique was developed to semi-quantitatively evaluate nanomaterial exposures and employs a combination of filter based samples and portable real-time particle monitors, including a condensation particle counter (CPC) and an optical particle counter (OPC), to detect nanomaterial releases. This laboratory study evaluated the results from CPC and OPC simultaneously measuring a polydisperse aerosol to assess their variability and accuracy. METHODS AND RESULTS: Two CPCs and two OPCs were used to evaluate a polydisperse sodium chloride aerosol within an enclosed chamber. The measurement results for number concentration versus time were compared between paired particle monitors of the same type, and to results from the Scanning Mobility Particle Spectrometer (SMPS) which was widely used to measure concentration of size-specific particles. According to analyses by using the Bland-Altman method, the CPCs displayed a constant mean percent difference of −3.8% (95% agreement limits: −9.1 to 1.6%; range of 95% agreement limit: 10.7%) with the chamber particle concentration below its dynamic upper limit (100,000 particles per cubic centimeter). The mean percent difference increased from −3.4% to −12.0% (range of 95% agreement limits: 7.1%) with increasing particle concentrations that were above the dynamic upper limit. The OPC results showed the percent difference within 15% for measurements in particles with size ranges of 300 to 500 and 500 to 1000 regardless of the particle concentration. Compared with SMPS measurements, the CPC gave a mean percent difference of 22.9% (95% agreement limits: 10.5% to 35.2%); whereas the measurements from OPC were not comparable. CONCLUSIONS: This study demonstrated that CPC and OPC are useful for measuring nanoparticle exposures but the results from an individual monitor should be interpreted based upon the instrument's technical parameters. Future research should challenge these monitors with particles of different sizes, shapes, or composition, to determine measurement comparability and accuracy across various workplace nanomaterials. Public Library of Science 2014-08-22 /pmc/articles/PMC4141826/ /pubmed/25148239 http://dx.doi.org/10.1371/journal.pone.0105769 Text en © 2014 Liu et al http://creativecommons.org/licenses/by/4.0/ This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are properly credited.
spellingShingle Research Article
Liu, Yuewei
Beaucham, Catherine C.
Pearce, Terri A.
Zhuang, Ziqing
Assessment of Two Portable Real-Time Particle Monitors Used in Nanomaterial Workplace Exposure Evaluations
title Assessment of Two Portable Real-Time Particle Monitors Used in Nanomaterial Workplace Exposure Evaluations
title_full Assessment of Two Portable Real-Time Particle Monitors Used in Nanomaterial Workplace Exposure Evaluations
title_fullStr Assessment of Two Portable Real-Time Particle Monitors Used in Nanomaterial Workplace Exposure Evaluations
title_full_unstemmed Assessment of Two Portable Real-Time Particle Monitors Used in Nanomaterial Workplace Exposure Evaluations
title_short Assessment of Two Portable Real-Time Particle Monitors Used in Nanomaterial Workplace Exposure Evaluations
title_sort assessment of two portable real-time particle monitors used in nanomaterial workplace exposure evaluations
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4141826/
https://www.ncbi.nlm.nih.gov/pubmed/25148239
http://dx.doi.org/10.1371/journal.pone.0105769
work_keys_str_mv AT liuyuewei assessmentoftwoportablerealtimeparticlemonitorsusedinnanomaterialworkplaceexposureevaluations
AT beauchamcatherinec assessmentoftwoportablerealtimeparticlemonitorsusedinnanomaterialworkplaceexposureevaluations
AT pearceterria assessmentoftwoportablerealtimeparticlemonitorsusedinnanomaterialworkplaceexposureevaluations
AT zhuangziqing assessmentoftwoportablerealtimeparticlemonitorsusedinnanomaterialworkplaceexposureevaluations