A Recommendation for Revised Dose Calibrator Measurement Procedures for (89)Zr and (124)I
Because of their chemical properties and multiday half lives, iodine-124 and zirconium-89 are being used in a growing number of PET imaging studies. Some aspects of their quantitation, however, still need attention. For (89)Zr the PET images should, in principle, be as quantitatively accurate as sim...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Public Library of Science
2014
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4159273/ https://www.ncbi.nlm.nih.gov/pubmed/25202987 http://dx.doi.org/10.1371/journal.pone.0106868 |
_version_ | 1782334186391076864 |
---|---|
author | Beattie, Bradley J. Pentlow, Keith S. O'Donoghue, Joseph Humm, John L. |
author_facet | Beattie, Bradley J. Pentlow, Keith S. O'Donoghue, Joseph Humm, John L. |
author_sort | Beattie, Bradley J. |
collection | PubMed |
description | Because of their chemical properties and multiday half lives, iodine-124 and zirconium-89 are being used in a growing number of PET imaging studies. Some aspects of their quantitation, however, still need attention. For (89)Zr the PET images should, in principle, be as quantitatively accurate as similarly reconstructed (18)F measurements. We found, however, that images of a 20 cm well calibration phantom containing (89)Zr underestimated the activity by approximately 10% relative to a dose calibrator measurement (Capintec CRC-15R) using a published calibration setting number of 465. PET images of (124)I, in contrast, are complicated by the contribution of decays in cascade that add spurious coincident events to the PET data. When these cascade coincidences are properly accounted for, quantitatively accurate images should be possible. We found, however, that even with this correction we still encountered what appeared to be a large variability in the accuracy of the PET images when compared to dose calibrator measurements made using the calibration setting number, 570, recommended by Capintec. We derive new calibration setting numbers for (89)Zr and (124)I based on their 511 keV photon peaks as measured on an HPGe detector. The peaks were calibrated relative to an (18)F standard, the activity level of which was precisely measured in a dose calibrator under well-defined measurement conditions. When measuring (89)Zr on a Capintec CRC-15R we propose the use of calibration setting number 517. And for (124)I, we recommend the use of a copper filter surrounding the sample and the use of calibration setting number 494. The new dose calibrator measurement procedures we propose will result in more consistent and accurate radioactivity measurements of (89)Zr and (124)I. These and other positron emitting radionuclides can be accurately calibrated relative to (18)F based on measurements of their 511 keV peaks and knowledge of their relative positron abundances. |
format | Online Article Text |
id | pubmed-4159273 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2014 |
publisher | Public Library of Science |
record_format | MEDLINE/PubMed |
spelling | pubmed-41592732014-09-12 A Recommendation for Revised Dose Calibrator Measurement Procedures for (89)Zr and (124)I Beattie, Bradley J. Pentlow, Keith S. O'Donoghue, Joseph Humm, John L. PLoS One Research Article Because of their chemical properties and multiday half lives, iodine-124 and zirconium-89 are being used in a growing number of PET imaging studies. Some aspects of their quantitation, however, still need attention. For (89)Zr the PET images should, in principle, be as quantitatively accurate as similarly reconstructed (18)F measurements. We found, however, that images of a 20 cm well calibration phantom containing (89)Zr underestimated the activity by approximately 10% relative to a dose calibrator measurement (Capintec CRC-15R) using a published calibration setting number of 465. PET images of (124)I, in contrast, are complicated by the contribution of decays in cascade that add spurious coincident events to the PET data. When these cascade coincidences are properly accounted for, quantitatively accurate images should be possible. We found, however, that even with this correction we still encountered what appeared to be a large variability in the accuracy of the PET images when compared to dose calibrator measurements made using the calibration setting number, 570, recommended by Capintec. We derive new calibration setting numbers for (89)Zr and (124)I based on their 511 keV photon peaks as measured on an HPGe detector. The peaks were calibrated relative to an (18)F standard, the activity level of which was precisely measured in a dose calibrator under well-defined measurement conditions. When measuring (89)Zr on a Capintec CRC-15R we propose the use of calibration setting number 517. And for (124)I, we recommend the use of a copper filter surrounding the sample and the use of calibration setting number 494. The new dose calibrator measurement procedures we propose will result in more consistent and accurate radioactivity measurements of (89)Zr and (124)I. These and other positron emitting radionuclides can be accurately calibrated relative to (18)F based on measurements of their 511 keV peaks and knowledge of their relative positron abundances. Public Library of Science 2014-09-09 /pmc/articles/PMC4159273/ /pubmed/25202987 http://dx.doi.org/10.1371/journal.pone.0106868 Text en © 2014 Beattie et al http://creativecommons.org/licenses/by/4.0/ This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are properly credited. |
spellingShingle | Research Article Beattie, Bradley J. Pentlow, Keith S. O'Donoghue, Joseph Humm, John L. A Recommendation for Revised Dose Calibrator Measurement Procedures for (89)Zr and (124)I |
title | A Recommendation for Revised Dose Calibrator Measurement Procedures for (89)Zr and (124)I |
title_full | A Recommendation for Revised Dose Calibrator Measurement Procedures for (89)Zr and (124)I |
title_fullStr | A Recommendation for Revised Dose Calibrator Measurement Procedures for (89)Zr and (124)I |
title_full_unstemmed | A Recommendation for Revised Dose Calibrator Measurement Procedures for (89)Zr and (124)I |
title_short | A Recommendation for Revised Dose Calibrator Measurement Procedures for (89)Zr and (124)I |
title_sort | recommendation for revised dose calibrator measurement procedures for (89)zr and (124)i |
topic | Research Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4159273/ https://www.ncbi.nlm.nih.gov/pubmed/25202987 http://dx.doi.org/10.1371/journal.pone.0106868 |
work_keys_str_mv | AT beattiebradleyj arecommendationforreviseddosecalibratormeasurementproceduresfor89zrand124i AT pentlowkeiths arecommendationforreviseddosecalibratormeasurementproceduresfor89zrand124i AT odonoghuejoseph arecommendationforreviseddosecalibratormeasurementproceduresfor89zrand124i AT hummjohnl arecommendationforreviseddosecalibratormeasurementproceduresfor89zrand124i AT beattiebradleyj recommendationforreviseddosecalibratormeasurementproceduresfor89zrand124i AT pentlowkeiths recommendationforreviseddosecalibratormeasurementproceduresfor89zrand124i AT odonoghuejoseph recommendationforreviseddosecalibratormeasurementproceduresfor89zrand124i AT hummjohnl recommendationforreviseddosecalibratormeasurementproceduresfor89zrand124i |