Prefilled syringes and usability of ophthalmic viscosurgical devices
PURPOSE: To examine the influence of the configuration of prefilled syringes on the usability of ophthalmic viscosurgical devices (OVDs). SETTING: Department of Ophthalmology, Jikei University School of Medicine, Tokyo, Japan. DESIGN: Experimental study. METHODS: The maximum force needed to extrude...
Autores principales: | , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Dove Medical Press
2014
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4160326/ https://www.ncbi.nlm.nih.gov/pubmed/25228786 http://dx.doi.org/10.2147/OPTH.S67689 |
_version_ | 1782334385670848512 |
---|---|
author | Shiba, Takuya Tsuneoka, Hiroshi |
author_facet | Shiba, Takuya Tsuneoka, Hiroshi |
author_sort | Shiba, Takuya |
collection | PubMed |
description | PURPOSE: To examine the influence of the configuration of prefilled syringes on the usability of ophthalmic viscosurgical devices (OVDs). SETTING: Department of Ophthalmology, Jikei University School of Medicine, Tokyo, Japan. DESIGN: Experimental study. METHODS: The maximum force needed to extrude the whole OVD (extrusion force) was compared among viscous cohesive OVDs (OPEGAN Hi(®) 0.85 mL and Healon(®) 0.85 mL) and very low viscosity dispersive OVDs (OPEGAN(®) 0.6 mL and OPEGAN(®) 1.1 mL). Additionally, to exclude the influence of any differences between syringes of viscous cohesive OVDs on the extrusion force, empty syringes of the same configuration were refilled with the same products. In addition, the syringe inner surface and that of the piston attached to the tip of the plunger were measured. RESULTS: The extrusion force of Healon 0.85 mL (3.28±0.19 kgf) was significantly (P<0.01) higher than that of OPEGAN Hi 0.85 mL (2.54±0.23 kgf). The inner syringe chamber for Healon 0.85 mL was vial shaped and had a built-in needle in the Luer tip, which was clearly different from OPEGAN Hi 0.85 mL. There were no significant differences in the extrusion force between refilled syringes. The extrusion force of OPEGAN 1.1 mL (3.44±0.12 kgf) was significantly (P<0.01) higher than that of OPEGAN 0.6 mL (1.63±0.06 kgf). The syringe of OPEGAN 1.1 mL was obviously bigger than that of OPEGAN 0.6 mL. CONCLUSION: It was confirmed that the configuration of the syringes is another determinant of the extrusion force and a factor related to the difference in usability among products that meet the same cohesive class. |
format | Online Article Text |
id | pubmed-4160326 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2014 |
publisher | Dove Medical Press |
record_format | MEDLINE/PubMed |
spelling | pubmed-41603262014-09-16 Prefilled syringes and usability of ophthalmic viscosurgical devices Shiba, Takuya Tsuneoka, Hiroshi Clin Ophthalmol Original Research PURPOSE: To examine the influence of the configuration of prefilled syringes on the usability of ophthalmic viscosurgical devices (OVDs). SETTING: Department of Ophthalmology, Jikei University School of Medicine, Tokyo, Japan. DESIGN: Experimental study. METHODS: The maximum force needed to extrude the whole OVD (extrusion force) was compared among viscous cohesive OVDs (OPEGAN Hi(®) 0.85 mL and Healon(®) 0.85 mL) and very low viscosity dispersive OVDs (OPEGAN(®) 0.6 mL and OPEGAN(®) 1.1 mL). Additionally, to exclude the influence of any differences between syringes of viscous cohesive OVDs on the extrusion force, empty syringes of the same configuration were refilled with the same products. In addition, the syringe inner surface and that of the piston attached to the tip of the plunger were measured. RESULTS: The extrusion force of Healon 0.85 mL (3.28±0.19 kgf) was significantly (P<0.01) higher than that of OPEGAN Hi 0.85 mL (2.54±0.23 kgf). The inner syringe chamber for Healon 0.85 mL was vial shaped and had a built-in needle in the Luer tip, which was clearly different from OPEGAN Hi 0.85 mL. There were no significant differences in the extrusion force between refilled syringes. The extrusion force of OPEGAN 1.1 mL (3.44±0.12 kgf) was significantly (P<0.01) higher than that of OPEGAN 0.6 mL (1.63±0.06 kgf). The syringe of OPEGAN 1.1 mL was obviously bigger than that of OPEGAN 0.6 mL. CONCLUSION: It was confirmed that the configuration of the syringes is another determinant of the extrusion force and a factor related to the difference in usability among products that meet the same cohesive class. Dove Medical Press 2014-09-04 /pmc/articles/PMC4160326/ /pubmed/25228786 http://dx.doi.org/10.2147/OPTH.S67689 Text en © 2014 Shiba and Tsuneoka. This work is published by Dove Medical Press Limited, and licensed under Creative Commons Attribution – Non Commercial (unported, v3.0) License The full terms of the License are available at http://creativecommons.org/licenses/by-nc/3.0/. Non-commercial uses of the work are permitted without any further permission from Dove Medical Press Limited, provided the work is properly attributed. |
spellingShingle | Original Research Shiba, Takuya Tsuneoka, Hiroshi Prefilled syringes and usability of ophthalmic viscosurgical devices |
title | Prefilled syringes and usability of ophthalmic viscosurgical devices |
title_full | Prefilled syringes and usability of ophthalmic viscosurgical devices |
title_fullStr | Prefilled syringes and usability of ophthalmic viscosurgical devices |
title_full_unstemmed | Prefilled syringes and usability of ophthalmic viscosurgical devices |
title_short | Prefilled syringes and usability of ophthalmic viscosurgical devices |
title_sort | prefilled syringes and usability of ophthalmic viscosurgical devices |
topic | Original Research |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4160326/ https://www.ncbi.nlm.nih.gov/pubmed/25228786 http://dx.doi.org/10.2147/OPTH.S67689 |
work_keys_str_mv | AT shibatakuya prefilledsyringesandusabilityofophthalmicviscosurgicaldevices AT tsuneokahiroshi prefilledsyringesandusabilityofophthalmicviscosurgicaldevices |