Quantification of left ventricular functional parameter values using 3D spiral bSSFP and through-time Non-Cartesian GRAPPA
BACKGROUND: The standard clinical acquisition for left ventricular functional parameter analysis with cardiovascular magnetic resonance (CMR) uses a multi-breathhold multi-slice segmented balanced SSFP sequence. Performing multiple long breathholds in quick succession for ventricular coverage in the...
Autores principales: | , , , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
BioMed Central
2014
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4160541/ https://www.ncbi.nlm.nih.gov/pubmed/25231607 http://dx.doi.org/10.1186/s12968-014-0065-1 |
_version_ | 1782334397615177728 |
---|---|
author | Barkauskas, Kestutis J Rajiah, Prabhakar Ashwath, Ravi Hamilton, Jesse I Chen, Yong Ma, Dan Wright, Katherine L Gulani, Vikas Griswold, Mark A Seiberlich, Nicole |
author_facet | Barkauskas, Kestutis J Rajiah, Prabhakar Ashwath, Ravi Hamilton, Jesse I Chen, Yong Ma, Dan Wright, Katherine L Gulani, Vikas Griswold, Mark A Seiberlich, Nicole |
author_sort | Barkauskas, Kestutis J |
collection | PubMed |
description | BACKGROUND: The standard clinical acquisition for left ventricular functional parameter analysis with cardiovascular magnetic resonance (CMR) uses a multi-breathhold multi-slice segmented balanced SSFP sequence. Performing multiple long breathholds in quick succession for ventricular coverage in the short-axis orientation can lead to fatigue and is challenging in patients with severe cardiac or respiratory disorders. This study combines the encoding efficiency of a six-fold undersampled 3D stack of spirals balanced SSFP sequence with 3D through-time spiral GRAPPA parallel imaging reconstruction. This 3D spiral method requires only one breathhold to collect the dynamic data. METHODS: Ten healthy volunteers were recruited for imaging at 3 T. The 3D spiral technique was compared against 2D imaging in terms of systolic left ventricular functional parameter values (Bland-Altman plots), total scan time (Welch’s t-test) and qualitative image rating scores (Wilcoxon signed-rank test). RESULTS: Systolic left ventricular functional values were not significantly different (i.e. 3D-2D) between the methods. The 95% confidence interval for ejection fraction was −0.1 ± 1.6% (mean ± 1.96*SD). The total scan time for the 3D spiral technique was 48 s, which included one breathhold with an average duration of 14 s for the dynamic scan, plus 34 s to collect the calibration data under free-breathing conditions. The 2D method required an average of 5min40s for the same coverage of the left ventricle. The difference between 3D and 2D image rating scores was significantly different from zero (Wilcoxon signed-rank test, p < 0.05); however, the scores were at least 3 (i.e. average) or higher for 3D spiral imaging. CONCLUSION: The 3D through-time spiral GRAPPA method demonstrated equivalent systolic left ventricular functional parameter values, required significantly less total scan time and yielded acceptable image quality with respect to the 2D segmented multi-breathhold standard in this study. Moreover, the 3D spiral technique used just one breathhold for dynamic imaging, which is anticipated to reduce patient fatigue as part of the complete cardiac examination in future studies that include patients. ELECTRONIC SUPPLEMENTARY MATERIAL: The online version of this article (doi:10.1186/s12968-014-0065-1) contains supplementary material, which is available to authorized users. |
format | Online Article Text |
id | pubmed-4160541 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2014 |
publisher | BioMed Central |
record_format | MEDLINE/PubMed |
spelling | pubmed-41605412014-09-12 Quantification of left ventricular functional parameter values using 3D spiral bSSFP and through-time Non-Cartesian GRAPPA Barkauskas, Kestutis J Rajiah, Prabhakar Ashwath, Ravi Hamilton, Jesse I Chen, Yong Ma, Dan Wright, Katherine L Gulani, Vikas Griswold, Mark A Seiberlich, Nicole J Cardiovasc Magn Reson Technical Notes BACKGROUND: The standard clinical acquisition for left ventricular functional parameter analysis with cardiovascular magnetic resonance (CMR) uses a multi-breathhold multi-slice segmented balanced SSFP sequence. Performing multiple long breathholds in quick succession for ventricular coverage in the short-axis orientation can lead to fatigue and is challenging in patients with severe cardiac or respiratory disorders. This study combines the encoding efficiency of a six-fold undersampled 3D stack of spirals balanced SSFP sequence with 3D through-time spiral GRAPPA parallel imaging reconstruction. This 3D spiral method requires only one breathhold to collect the dynamic data. METHODS: Ten healthy volunteers were recruited for imaging at 3 T. The 3D spiral technique was compared against 2D imaging in terms of systolic left ventricular functional parameter values (Bland-Altman plots), total scan time (Welch’s t-test) and qualitative image rating scores (Wilcoxon signed-rank test). RESULTS: Systolic left ventricular functional values were not significantly different (i.e. 3D-2D) between the methods. The 95% confidence interval for ejection fraction was −0.1 ± 1.6% (mean ± 1.96*SD). The total scan time for the 3D spiral technique was 48 s, which included one breathhold with an average duration of 14 s for the dynamic scan, plus 34 s to collect the calibration data under free-breathing conditions. The 2D method required an average of 5min40s for the same coverage of the left ventricle. The difference between 3D and 2D image rating scores was significantly different from zero (Wilcoxon signed-rank test, p < 0.05); however, the scores were at least 3 (i.e. average) or higher for 3D spiral imaging. CONCLUSION: The 3D through-time spiral GRAPPA method demonstrated equivalent systolic left ventricular functional parameter values, required significantly less total scan time and yielded acceptable image quality with respect to the 2D segmented multi-breathhold standard in this study. Moreover, the 3D spiral technique used just one breathhold for dynamic imaging, which is anticipated to reduce patient fatigue as part of the complete cardiac examination in future studies that include patients. ELECTRONIC SUPPLEMENTARY MATERIAL: The online version of this article (doi:10.1186/s12968-014-0065-1) contains supplementary material, which is available to authorized users. BioMed Central 2014-09-11 /pmc/articles/PMC4160541/ /pubmed/25231607 http://dx.doi.org/10.1186/s12968-014-0065-1 Text en © Barkauskas et al.; licensee BioMed Central Ltd. 2014 This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly credited. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated. |
spellingShingle | Technical Notes Barkauskas, Kestutis J Rajiah, Prabhakar Ashwath, Ravi Hamilton, Jesse I Chen, Yong Ma, Dan Wright, Katherine L Gulani, Vikas Griswold, Mark A Seiberlich, Nicole Quantification of left ventricular functional parameter values using 3D spiral bSSFP and through-time Non-Cartesian GRAPPA |
title | Quantification of left ventricular functional parameter values using 3D spiral bSSFP and through-time Non-Cartesian GRAPPA |
title_full | Quantification of left ventricular functional parameter values using 3D spiral bSSFP and through-time Non-Cartesian GRAPPA |
title_fullStr | Quantification of left ventricular functional parameter values using 3D spiral bSSFP and through-time Non-Cartesian GRAPPA |
title_full_unstemmed | Quantification of left ventricular functional parameter values using 3D spiral bSSFP and through-time Non-Cartesian GRAPPA |
title_short | Quantification of left ventricular functional parameter values using 3D spiral bSSFP and through-time Non-Cartesian GRAPPA |
title_sort | quantification of left ventricular functional parameter values using 3d spiral bssfp and through-time non-cartesian grappa |
topic | Technical Notes |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4160541/ https://www.ncbi.nlm.nih.gov/pubmed/25231607 http://dx.doi.org/10.1186/s12968-014-0065-1 |
work_keys_str_mv | AT barkauskaskestutisj quantificationofleftventricularfunctionalparametervaluesusing3dspiralbssfpandthroughtimenoncartesiangrappa AT rajiahprabhakar quantificationofleftventricularfunctionalparametervaluesusing3dspiralbssfpandthroughtimenoncartesiangrappa AT ashwathravi quantificationofleftventricularfunctionalparametervaluesusing3dspiralbssfpandthroughtimenoncartesiangrappa AT hamiltonjessei quantificationofleftventricularfunctionalparametervaluesusing3dspiralbssfpandthroughtimenoncartesiangrappa AT chenyong quantificationofleftventricularfunctionalparametervaluesusing3dspiralbssfpandthroughtimenoncartesiangrappa AT madan quantificationofleftventricularfunctionalparametervaluesusing3dspiralbssfpandthroughtimenoncartesiangrappa AT wrightkatherinel quantificationofleftventricularfunctionalparametervaluesusing3dspiralbssfpandthroughtimenoncartesiangrappa AT gulanivikas quantificationofleftventricularfunctionalparametervaluesusing3dspiralbssfpandthroughtimenoncartesiangrappa AT griswoldmarka quantificationofleftventricularfunctionalparametervaluesusing3dspiralbssfpandthroughtimenoncartesiangrappa AT seiberlichnicole quantificationofleftventricularfunctionalparametervaluesusing3dspiralbssfpandthroughtimenoncartesiangrappa |