A Osteogenesis Distraction Device Enabling Control of Vertical Direction for Syndromic Craniosynostosis

BACKGROUND: We have developed a hybrid facial osteogenesis distraction system that combines the advantages of external and internal distraction devices to enable control of both the distraction distance and vector. However, when the advanced maxilla has excessive clockwise rotation and shifts more d...

Descripción completa

Detalles Bibliográficos
Autores principales: Kobayashi, Shinji, Fukawa, Toshihiko, Hirakawa, Takashi, Maegawa, Jiro
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Wolters Kluwer Health 2014
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4173834/
https://www.ncbi.nlm.nih.gov/pubmed/25289307
http://dx.doi.org/10.1097/GOX.0000000000000060
Descripción
Sumario:BACKGROUND: We have developed a hybrid facial osteogenesis distraction system that combines the advantages of external and internal distraction devices to enable control of both the distraction distance and vector. However, when the advanced maxilla has excessive clockwise rotation and shifts more downward vertically than planned, it might be impossible to pull it up to correct it. We invented devices attached to external distraction systems that can control the vertical vector of distraction to resolve this problem. The purpose of this article is to describe the result of utilizing the distraction system for syndromic craniosynostosis. METHODS: In addition to a previously reported hybrid facial distraction system, the devices for controlling the vertical direction of the advanced maxilla were attached to the external distraction device. The vertical direction of the advanced maxilla can be controlled by adjustment of the spindle units. This system was used for 2 patients with Crouzon and Apert syndrome. RESULTS: The system enabled control of the vertical distance, with no complications during the procedures. As a result, the maxilla could be advanced into the planned position including overcorrection without excessive clockwise rotation of distraction. CONCLUSION: Our system can alter the cases and bring them into the planned position, by controlling the vertical vector of distraction. We believe that this system might be effective in infants with syndromic craniosynostosis as it involves 2 osteotomies and horizontal and vertical direction of elongation can be controlled.