Cargando…

A Simple Technique to Deflate a Tissue Expander

SUMMARY: Removal of an intact, inflated tissue expander can result in uncontrolled spillage of expander contents on the drapes, surgeons, staff, or patient. The technique described has been used for the past 4 years to eliminate these spillage concerns. It is simple and inexpensive using common disp...

Descripción completa

Detalles Bibliográficos
Autores principales: Topol, Bruce M., Dalton, Edward
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Wolters Kluwer Health 2014
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4174205/
https://www.ncbi.nlm.nih.gov/pubmed/25289324
http://dx.doi.org/10.1097/GOX.0000000000000076
Descripción
Sumario:SUMMARY: Removal of an intact, inflated tissue expander can result in uncontrolled spillage of expander contents on the drapes, surgeons, staff, or patient. The technique described has been used for the past 4 years to eliminate these spillage concerns. It is simple and inexpensive using common disposable supplies found in any operating room. An irrigating syringe is modified by cutting away its 2 flanges. The bulb is removed and suction is applied to the syringe tip. The expander is punctured with a scalpel while the puncture hole is covered by the syringe opening. The applied suction aspirates the contents thereby deflating the expander. Uncontrolled content spillage and spray are eliminated.