Reliability of Corneal Dynamic Scheimpflug Analyser Measurements in Virgin and Post-PRK Eyes

PURPOSE: To determine the measurement reliability of CorVis ST, a dynamic Scheimpflug analyser, in virgin and post-photorefractive keratectomy (PRK) eyes and compare the results between these two groups. METHODS: Forty virgin eyes and 42 post-PRK eyes underwent CorVis ST measurements performed by tw...

Descripción completa

Detalles Bibliográficos
Autores principales: Chen, Xiangjun, Stojanovic, Aleksandar, Hua, Yanjun, Eidet, Jon Roger, Hu, Di, Wang, Jingting, Utheim, Tor Paaske
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Public Library of Science 2014
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4193795/
https://www.ncbi.nlm.nih.gov/pubmed/25302580
http://dx.doi.org/10.1371/journal.pone.0109577
_version_ 1782339034899546112
author Chen, Xiangjun
Stojanovic, Aleksandar
Hua, Yanjun
Eidet, Jon Roger
Hu, Di
Wang, Jingting
Utheim, Tor Paaske
author_facet Chen, Xiangjun
Stojanovic, Aleksandar
Hua, Yanjun
Eidet, Jon Roger
Hu, Di
Wang, Jingting
Utheim, Tor Paaske
author_sort Chen, Xiangjun
collection PubMed
description PURPOSE: To determine the measurement reliability of CorVis ST, a dynamic Scheimpflug analyser, in virgin and post-photorefractive keratectomy (PRK) eyes and compare the results between these two groups. METHODS: Forty virgin eyes and 42 post-PRK eyes underwent CorVis ST measurements performed by two technicians. Repeatability was evaluated by comparing three consecutive measurements by technician A. Reproducibility was determined by comparing the first measurement by technician A with one performed by technician B. Intraobserver and interobserver intraclass correlation coefficients (ICCs) were calculated. Univariate analysis of covariance (ANCOVA) was used to compare measured parameters between virgin and post-PRK eyes. RESULTS: The intraocular pressure (IOP), central corneal thickness (CCT) and 1(st) applanation time demonstrated good intraobserver repeatability and interobserver reproducibility (ICC≧0.90) in virgin and post-PRK eyes. The deformation amplitude showed a good or close to good repeatability and reproducibility in both groups (ICC≧0.88). The CCT correlated positively with 1(st) applanation time (r = 0.437 and 0.483, respectively, p<0.05) and negatively with deformation amplitude (r = −0.384 and −0.375, respectively, p<0.05) in both groups. Compared to post-PRK eyes, virgin eyes showed longer 1(st) applanation time (7.29±0.21 vs. 6.96±0.17 ms, p<0.05) and lower deformation amplitude (1.06±0.07 vs. 1.17±0.08 mm, p<0.05). CONCLUSIONS: CorVis ST demonstrated reliable measurements for CCT, IOP, and 1(st) applanation time, as well as relatively reliable measurement for deformation amplitude in both virgin and post-PRK eyes. There were differences in 1(st) applanation time and deformation amplitude between virgin and post-PRK eyes, which may reflect corneal biomechanical changes occurring after the surgery in the latter.
format Online
Article
Text
id pubmed-4193795
institution National Center for Biotechnology Information
language English
publishDate 2014
publisher Public Library of Science
record_format MEDLINE/PubMed
spelling pubmed-41937952014-10-14 Reliability of Corneal Dynamic Scheimpflug Analyser Measurements in Virgin and Post-PRK Eyes Chen, Xiangjun Stojanovic, Aleksandar Hua, Yanjun Eidet, Jon Roger Hu, Di Wang, Jingting Utheim, Tor Paaske PLoS One Research Article PURPOSE: To determine the measurement reliability of CorVis ST, a dynamic Scheimpflug analyser, in virgin and post-photorefractive keratectomy (PRK) eyes and compare the results between these two groups. METHODS: Forty virgin eyes and 42 post-PRK eyes underwent CorVis ST measurements performed by two technicians. Repeatability was evaluated by comparing three consecutive measurements by technician A. Reproducibility was determined by comparing the first measurement by technician A with one performed by technician B. Intraobserver and interobserver intraclass correlation coefficients (ICCs) were calculated. Univariate analysis of covariance (ANCOVA) was used to compare measured parameters between virgin and post-PRK eyes. RESULTS: The intraocular pressure (IOP), central corneal thickness (CCT) and 1(st) applanation time demonstrated good intraobserver repeatability and interobserver reproducibility (ICC≧0.90) in virgin and post-PRK eyes. The deformation amplitude showed a good or close to good repeatability and reproducibility in both groups (ICC≧0.88). The CCT correlated positively with 1(st) applanation time (r = 0.437 and 0.483, respectively, p<0.05) and negatively with deformation amplitude (r = −0.384 and −0.375, respectively, p<0.05) in both groups. Compared to post-PRK eyes, virgin eyes showed longer 1(st) applanation time (7.29±0.21 vs. 6.96±0.17 ms, p<0.05) and lower deformation amplitude (1.06±0.07 vs. 1.17±0.08 mm, p<0.05). CONCLUSIONS: CorVis ST demonstrated reliable measurements for CCT, IOP, and 1(st) applanation time, as well as relatively reliable measurement for deformation amplitude in both virgin and post-PRK eyes. There were differences in 1(st) applanation time and deformation amplitude between virgin and post-PRK eyes, which may reflect corneal biomechanical changes occurring after the surgery in the latter. Public Library of Science 2014-10-10 /pmc/articles/PMC4193795/ /pubmed/25302580 http://dx.doi.org/10.1371/journal.pone.0109577 Text en © 2014 Chen et al http://creativecommons.org/licenses/by/4.0/ This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are properly credited.
spellingShingle Research Article
Chen, Xiangjun
Stojanovic, Aleksandar
Hua, Yanjun
Eidet, Jon Roger
Hu, Di
Wang, Jingting
Utheim, Tor Paaske
Reliability of Corneal Dynamic Scheimpflug Analyser Measurements in Virgin and Post-PRK Eyes
title Reliability of Corneal Dynamic Scheimpflug Analyser Measurements in Virgin and Post-PRK Eyes
title_full Reliability of Corneal Dynamic Scheimpflug Analyser Measurements in Virgin and Post-PRK Eyes
title_fullStr Reliability of Corneal Dynamic Scheimpflug Analyser Measurements in Virgin and Post-PRK Eyes
title_full_unstemmed Reliability of Corneal Dynamic Scheimpflug Analyser Measurements in Virgin and Post-PRK Eyes
title_short Reliability of Corneal Dynamic Scheimpflug Analyser Measurements in Virgin and Post-PRK Eyes
title_sort reliability of corneal dynamic scheimpflug analyser measurements in virgin and post-prk eyes
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4193795/
https://www.ncbi.nlm.nih.gov/pubmed/25302580
http://dx.doi.org/10.1371/journal.pone.0109577
work_keys_str_mv AT chenxiangjun reliabilityofcornealdynamicscheimpfluganalysermeasurementsinvirginandpostprkeyes
AT stojanovicaleksandar reliabilityofcornealdynamicscheimpfluganalysermeasurementsinvirginandpostprkeyes
AT huayanjun reliabilityofcornealdynamicscheimpfluganalysermeasurementsinvirginandpostprkeyes
AT eidetjonroger reliabilityofcornealdynamicscheimpfluganalysermeasurementsinvirginandpostprkeyes
AT hudi reliabilityofcornealdynamicscheimpfluganalysermeasurementsinvirginandpostprkeyes
AT wangjingting reliabilityofcornealdynamicscheimpfluganalysermeasurementsinvirginandpostprkeyes
AT utheimtorpaaske reliabilityofcornealdynamicscheimpfluganalysermeasurementsinvirginandpostprkeyes