Insights into the Problem of Alarm Fatigue with Physiologic Monitor Devices: A Comprehensive Observational Study of Consecutive Intensive Care Unit Patients
PURPOSE: Physiologic monitors are plagued with alarms that create a cacophony of sounds and visual alerts causing “alarm fatigue” which creates an unsafe patient environment because a life-threatening event may be missed in this milieu of sensory overload. Using a state-of-the-art technology acquisi...
Autores principales: | , , , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Public Library of Science
2014
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4206416/ https://www.ncbi.nlm.nih.gov/pubmed/25338067 http://dx.doi.org/10.1371/journal.pone.0110274 |
_version_ | 1782340820086554624 |
---|---|
author | Drew, Barbara J. Harris, Patricia Zègre-Hemsey, Jessica K. Mammone, Tina Schindler, Daniel Salas-Boni, Rebeca Bai, Yong Tinoco, Adelita Ding, Quan Hu, Xiao |
author_facet | Drew, Barbara J. Harris, Patricia Zègre-Hemsey, Jessica K. Mammone, Tina Schindler, Daniel Salas-Boni, Rebeca Bai, Yong Tinoco, Adelita Ding, Quan Hu, Xiao |
author_sort | Drew, Barbara J. |
collection | PubMed |
description | PURPOSE: Physiologic monitors are plagued with alarms that create a cacophony of sounds and visual alerts causing “alarm fatigue” which creates an unsafe patient environment because a life-threatening event may be missed in this milieu of sensory overload. Using a state-of-the-art technology acquisition infrastructure, all monitor data including 7 ECG leads, all pressure, SpO2, and respiration waveforms as well as user settings and alarms were stored on 461 adults treated in intensive care units. Using a well-defined alarm annotation protocol, nurse scientists with 95% inter-rater reliability annotated 12,671 arrhythmia alarms. RESULTS: A total of 2,558,760 unique alarms occurred in the 31-day study period: arrhythmia, 1,154,201; parameter, 612,927; technical, 791,632. There were 381,560 audible alarms for an audible alarm burden of 187/bed/day. 88.8% of the 12,671 annotated arrhythmia alarms were false positives. Conditions causing excessive alarms included inappropriate alarm settings, persistent atrial fibrillation, and non-actionable events such as PVC's and brief spikes in ST segments. Low amplitude QRS complexes in some, but not all available ECG leads caused undercounting and false arrhythmia alarms. Wide QRS complexes due to bundle branch block or ventricular pacemaker rhythm caused false alarms. 93% of the 168 true ventricular tachycardia alarms were not sustained long enough to warrant treatment. DISCUSSION: The excessive number of physiologic monitor alarms is a complex interplay of inappropriate user settings, patient conditions, and algorithm deficiencies. Device solutions should focus on use of all available ECG leads to identify non-artifact leads and leads with adequate QRS amplitude. Devices should provide prompts to aide in more appropriate tailoring of alarm settings to individual patients. Atrial fibrillation alarms should be limited to new onset and termination of the arrhythmia and delays for ST-segment and other parameter alarms should be configurable. Because computer devices are more reliable than humans, an opportunity exists to improve physiologic monitoring and reduce alarm fatigue. |
format | Online Article Text |
id | pubmed-4206416 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2014 |
publisher | Public Library of Science |
record_format | MEDLINE/PubMed |
spelling | pubmed-42064162014-10-27 Insights into the Problem of Alarm Fatigue with Physiologic Monitor Devices: A Comprehensive Observational Study of Consecutive Intensive Care Unit Patients Drew, Barbara J. Harris, Patricia Zègre-Hemsey, Jessica K. Mammone, Tina Schindler, Daniel Salas-Boni, Rebeca Bai, Yong Tinoco, Adelita Ding, Quan Hu, Xiao PLoS One Research Article PURPOSE: Physiologic monitors are plagued with alarms that create a cacophony of sounds and visual alerts causing “alarm fatigue” which creates an unsafe patient environment because a life-threatening event may be missed in this milieu of sensory overload. Using a state-of-the-art technology acquisition infrastructure, all monitor data including 7 ECG leads, all pressure, SpO2, and respiration waveforms as well as user settings and alarms were stored on 461 adults treated in intensive care units. Using a well-defined alarm annotation protocol, nurse scientists with 95% inter-rater reliability annotated 12,671 arrhythmia alarms. RESULTS: A total of 2,558,760 unique alarms occurred in the 31-day study period: arrhythmia, 1,154,201; parameter, 612,927; technical, 791,632. There were 381,560 audible alarms for an audible alarm burden of 187/bed/day. 88.8% of the 12,671 annotated arrhythmia alarms were false positives. Conditions causing excessive alarms included inappropriate alarm settings, persistent atrial fibrillation, and non-actionable events such as PVC's and brief spikes in ST segments. Low amplitude QRS complexes in some, but not all available ECG leads caused undercounting and false arrhythmia alarms. Wide QRS complexes due to bundle branch block or ventricular pacemaker rhythm caused false alarms. 93% of the 168 true ventricular tachycardia alarms were not sustained long enough to warrant treatment. DISCUSSION: The excessive number of physiologic monitor alarms is a complex interplay of inappropriate user settings, patient conditions, and algorithm deficiencies. Device solutions should focus on use of all available ECG leads to identify non-artifact leads and leads with adequate QRS amplitude. Devices should provide prompts to aide in more appropriate tailoring of alarm settings to individual patients. Atrial fibrillation alarms should be limited to new onset and termination of the arrhythmia and delays for ST-segment and other parameter alarms should be configurable. Because computer devices are more reliable than humans, an opportunity exists to improve physiologic monitoring and reduce alarm fatigue. Public Library of Science 2014-10-22 /pmc/articles/PMC4206416/ /pubmed/25338067 http://dx.doi.org/10.1371/journal.pone.0110274 Text en © 2014 Drew et al http://creativecommons.org/licenses/by/4.0/ This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are properly credited. |
spellingShingle | Research Article Drew, Barbara J. Harris, Patricia Zègre-Hemsey, Jessica K. Mammone, Tina Schindler, Daniel Salas-Boni, Rebeca Bai, Yong Tinoco, Adelita Ding, Quan Hu, Xiao Insights into the Problem of Alarm Fatigue with Physiologic Monitor Devices: A Comprehensive Observational Study of Consecutive Intensive Care Unit Patients |
title | Insights into the Problem of Alarm Fatigue with Physiologic Monitor Devices: A Comprehensive Observational Study of Consecutive Intensive Care Unit Patients |
title_full | Insights into the Problem of Alarm Fatigue with Physiologic Monitor Devices: A Comprehensive Observational Study of Consecutive Intensive Care Unit Patients |
title_fullStr | Insights into the Problem of Alarm Fatigue with Physiologic Monitor Devices: A Comprehensive Observational Study of Consecutive Intensive Care Unit Patients |
title_full_unstemmed | Insights into the Problem of Alarm Fatigue with Physiologic Monitor Devices: A Comprehensive Observational Study of Consecutive Intensive Care Unit Patients |
title_short | Insights into the Problem of Alarm Fatigue with Physiologic Monitor Devices: A Comprehensive Observational Study of Consecutive Intensive Care Unit Patients |
title_sort | insights into the problem of alarm fatigue with physiologic monitor devices: a comprehensive observational study of consecutive intensive care unit patients |
topic | Research Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4206416/ https://www.ncbi.nlm.nih.gov/pubmed/25338067 http://dx.doi.org/10.1371/journal.pone.0110274 |
work_keys_str_mv | AT drewbarbaraj insightsintotheproblemofalarmfatiguewithphysiologicmonitordevicesacomprehensiveobservationalstudyofconsecutiveintensivecareunitpatients AT harrispatricia insightsintotheproblemofalarmfatiguewithphysiologicmonitordevicesacomprehensiveobservationalstudyofconsecutiveintensivecareunitpatients AT zegrehemseyjessicak insightsintotheproblemofalarmfatiguewithphysiologicmonitordevicesacomprehensiveobservationalstudyofconsecutiveintensivecareunitpatients AT mammonetina insightsintotheproblemofalarmfatiguewithphysiologicmonitordevicesacomprehensiveobservationalstudyofconsecutiveintensivecareunitpatients AT schindlerdaniel insightsintotheproblemofalarmfatiguewithphysiologicmonitordevicesacomprehensiveobservationalstudyofconsecutiveintensivecareunitpatients AT salasbonirebeca insightsintotheproblemofalarmfatiguewithphysiologicmonitordevicesacomprehensiveobservationalstudyofconsecutiveintensivecareunitpatients AT baiyong insightsintotheproblemofalarmfatiguewithphysiologicmonitordevicesacomprehensiveobservationalstudyofconsecutiveintensivecareunitpatients AT tinocoadelita insightsintotheproblemofalarmfatiguewithphysiologicmonitordevicesacomprehensiveobservationalstudyofconsecutiveintensivecareunitpatients AT dingquan insightsintotheproblemofalarmfatiguewithphysiologicmonitordevicesacomprehensiveobservationalstudyofconsecutiveintensivecareunitpatients AT huxiao insightsintotheproblemofalarmfatiguewithphysiologicmonitordevicesacomprehensiveobservationalstudyofconsecutiveintensivecareunitpatients |