Opportunities to improve storage and transportation of blood specimens for CD4 testing in a rural district in Zimbabwe using BD vacutainer CD4 stabilization tubes: a stability and diagnostic accuracy study
BACKGROUND: CD4+ T-cell testing of blood specimens collected in standard EDTA Vacutainer tubes and transported at ambient temperature, must be completed within 48 hours with the BD FACSCount™ flow cytometer, restricting specimen collection in remote clinics with no on-site testing and limited specim...
Autores principales: | , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
BioMed Central
2014
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4209078/ https://www.ncbi.nlm.nih.gov/pubmed/25330812 http://dx.doi.org/10.1186/s12879-014-0553-9 |
_version_ | 1782341219220717568 |
---|---|
author | Fajardo, Emmanuel Metcalf, Carol Mbofana, Elton van Vyve, Charlotte Munyaradzi, Dhodho Simons, Sandra Kuhudzayi, Misheck Bygrave, Helen |
author_facet | Fajardo, Emmanuel Metcalf, Carol Mbofana, Elton van Vyve, Charlotte Munyaradzi, Dhodho Simons, Sandra Kuhudzayi, Misheck Bygrave, Helen |
author_sort | Fajardo, Emmanuel |
collection | PubMed |
description | BACKGROUND: CD4+ T-cell testing of blood specimens collected in standard EDTA Vacutainer tubes and transported at ambient temperature, must be completed within 48 hours with the BD FACSCount™ flow cytometer, restricting specimen collection in remote clinics with no on-site testing and limited specimen transport services. We conducted a study in Buhera District, Zimbabwe, to assess the stability and accuracy of CD4+ T-cell results of samples collected in Stabilization Tubes (ST) and stored at ambient temperature for varying time periods. METHODS: Paired EDTA and ST samples were collected from 51 HIV-positive patients aged 18 years and older. CD4+ T-cell testing was done on arrival in the laboratory (Day 0). ST samples were retested on Days 3, 5, and 7. Nineteen ST samples were stored for an additional week and retested on Day 14. RESULTS: There was a strong correlation between absolute CD4+ T-cell counts measured in the EDTA Day 0 reference sample and Day 7 ST sample (Spearman's rho: 0.9778; mean difference: -4.9 cells/μL and limits of agreement (LOA): 98.5 and 88.7 cells/μL); and the reference sample and Day 14 ST sample (Spearman's rho: 0.9632; mean difference 5.1 cells/μL and LOA: -99.6 and 109.8 cells/μL. Using a 350 cells/μL threshold, the sensitivity, specificity, positive predictive value (PPV) and negative predictive value (NPV) were all 100% on Day 7, and 83.3%, 100%, 100% and 92.9% on Day 14. Using a 500 cells/μL threshold, the sensitivity, specificity, PPV and NVP were 100%, 88.5%, 88.5% and 100% on Day 7 and 88.9%, 80.0%, 80.0% and 88.9% on Day 14. CONCLUSIONS: CD4 ST can be used and stored up to 7 days as a reliable alternative to standard EDTA tubes in settings where CD4+ T-cell testing within 48 hours is not feasible. Despite the small sample size, results suggest that ST may be stored up to 14 days at room temperature for CD4 testing, without compromising accuracy. However, further studies with larger sample sizes are needed to confirm this preliminary finding. ELECTRONIC SUPPLEMENTARY MATERIAL: The online version of this article (doi:10.1186/s12879-014-0553-9) contains supplementary material, which is available to authorized users. |
format | Online Article Text |
id | pubmed-4209078 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2014 |
publisher | BioMed Central |
record_format | MEDLINE/PubMed |
spelling | pubmed-42090782014-10-28 Opportunities to improve storage and transportation of blood specimens for CD4 testing in a rural district in Zimbabwe using BD vacutainer CD4 stabilization tubes: a stability and diagnostic accuracy study Fajardo, Emmanuel Metcalf, Carol Mbofana, Elton van Vyve, Charlotte Munyaradzi, Dhodho Simons, Sandra Kuhudzayi, Misheck Bygrave, Helen BMC Infect Dis Research Article BACKGROUND: CD4+ T-cell testing of blood specimens collected in standard EDTA Vacutainer tubes and transported at ambient temperature, must be completed within 48 hours with the BD FACSCount™ flow cytometer, restricting specimen collection in remote clinics with no on-site testing and limited specimen transport services. We conducted a study in Buhera District, Zimbabwe, to assess the stability and accuracy of CD4+ T-cell results of samples collected in Stabilization Tubes (ST) and stored at ambient temperature for varying time periods. METHODS: Paired EDTA and ST samples were collected from 51 HIV-positive patients aged 18 years and older. CD4+ T-cell testing was done on arrival in the laboratory (Day 0). ST samples were retested on Days 3, 5, and 7. Nineteen ST samples were stored for an additional week and retested on Day 14. RESULTS: There was a strong correlation between absolute CD4+ T-cell counts measured in the EDTA Day 0 reference sample and Day 7 ST sample (Spearman's rho: 0.9778; mean difference: -4.9 cells/μL and limits of agreement (LOA): 98.5 and 88.7 cells/μL); and the reference sample and Day 14 ST sample (Spearman's rho: 0.9632; mean difference 5.1 cells/μL and LOA: -99.6 and 109.8 cells/μL. Using a 350 cells/μL threshold, the sensitivity, specificity, positive predictive value (PPV) and negative predictive value (NPV) were all 100% on Day 7, and 83.3%, 100%, 100% and 92.9% on Day 14. Using a 500 cells/μL threshold, the sensitivity, specificity, PPV and NVP were 100%, 88.5%, 88.5% and 100% on Day 7 and 88.9%, 80.0%, 80.0% and 88.9% on Day 14. CONCLUSIONS: CD4 ST can be used and stored up to 7 days as a reliable alternative to standard EDTA tubes in settings where CD4+ T-cell testing within 48 hours is not feasible. Despite the small sample size, results suggest that ST may be stored up to 14 days at room temperature for CD4 testing, without compromising accuracy. However, further studies with larger sample sizes are needed to confirm this preliminary finding. ELECTRONIC SUPPLEMENTARY MATERIAL: The online version of this article (doi:10.1186/s12879-014-0553-9) contains supplementary material, which is available to authorized users. BioMed Central 2014-10-22 /pmc/articles/PMC4209078/ /pubmed/25330812 http://dx.doi.org/10.1186/s12879-014-0553-9 Text en © Fajardo et al.; licensee BioMed Central Ltd. 2014 This article is published under license to BioMed Central Ltd. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly credited. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated. |
spellingShingle | Research Article Fajardo, Emmanuel Metcalf, Carol Mbofana, Elton van Vyve, Charlotte Munyaradzi, Dhodho Simons, Sandra Kuhudzayi, Misheck Bygrave, Helen Opportunities to improve storage and transportation of blood specimens for CD4 testing in a rural district in Zimbabwe using BD vacutainer CD4 stabilization tubes: a stability and diagnostic accuracy study |
title | Opportunities to improve storage and transportation of blood specimens for CD4 testing in a rural district in Zimbabwe using BD vacutainer CD4 stabilization tubes: a stability and diagnostic accuracy study |
title_full | Opportunities to improve storage and transportation of blood specimens for CD4 testing in a rural district in Zimbabwe using BD vacutainer CD4 stabilization tubes: a stability and diagnostic accuracy study |
title_fullStr | Opportunities to improve storage and transportation of blood specimens for CD4 testing in a rural district in Zimbabwe using BD vacutainer CD4 stabilization tubes: a stability and diagnostic accuracy study |
title_full_unstemmed | Opportunities to improve storage and transportation of blood specimens for CD4 testing in a rural district in Zimbabwe using BD vacutainer CD4 stabilization tubes: a stability and diagnostic accuracy study |
title_short | Opportunities to improve storage and transportation of blood specimens for CD4 testing in a rural district in Zimbabwe using BD vacutainer CD4 stabilization tubes: a stability and diagnostic accuracy study |
title_sort | opportunities to improve storage and transportation of blood specimens for cd4 testing in a rural district in zimbabwe using bd vacutainer cd4 stabilization tubes: a stability and diagnostic accuracy study |
topic | Research Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4209078/ https://www.ncbi.nlm.nih.gov/pubmed/25330812 http://dx.doi.org/10.1186/s12879-014-0553-9 |
work_keys_str_mv | AT fajardoemmanuel opportunitiestoimprovestorageandtransportationofbloodspecimensforcd4testinginaruraldistrictinzimbabweusingbdvacutainercd4stabilizationtubesastabilityanddiagnosticaccuracystudy AT metcalfcarol opportunitiestoimprovestorageandtransportationofbloodspecimensforcd4testinginaruraldistrictinzimbabweusingbdvacutainercd4stabilizationtubesastabilityanddiagnosticaccuracystudy AT mbofanaelton opportunitiestoimprovestorageandtransportationofbloodspecimensforcd4testinginaruraldistrictinzimbabweusingbdvacutainercd4stabilizationtubesastabilityanddiagnosticaccuracystudy AT vanvyvecharlotte opportunitiestoimprovestorageandtransportationofbloodspecimensforcd4testinginaruraldistrictinzimbabweusingbdvacutainercd4stabilizationtubesastabilityanddiagnosticaccuracystudy AT munyaradzidhodho opportunitiestoimprovestorageandtransportationofbloodspecimensforcd4testinginaruraldistrictinzimbabweusingbdvacutainercd4stabilizationtubesastabilityanddiagnosticaccuracystudy AT simonssandra opportunitiestoimprovestorageandtransportationofbloodspecimensforcd4testinginaruraldistrictinzimbabweusingbdvacutainercd4stabilizationtubesastabilityanddiagnosticaccuracystudy AT kuhudzayimisheck opportunitiestoimprovestorageandtransportationofbloodspecimensforcd4testinginaruraldistrictinzimbabweusingbdvacutainercd4stabilizationtubesastabilityanddiagnosticaccuracystudy AT bygravehelen opportunitiestoimprovestorageandtransportationofbloodspecimensforcd4testinginaruraldistrictinzimbabweusingbdvacutainercd4stabilizationtubesastabilityanddiagnosticaccuracystudy |