INTRODUCTION: Preanalytical variables account for most of laboratory errors. There is a wide range of factors that affect the reliability of laboratory report. Most convenient sample type for routine laboratory analysis is serum. BD Vacutainer(®) Rapid Serum Tube (RST) (Becton, Dickinson and Company...
Autores principales: | , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Croatian Society of Medical Biochemistry and Laboratory Medicine
2014
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4210257/ https://www.ncbi.nlm.nih.gov/pubmed/25351355 http://dx.doi.org/10.11613/BM.2014.039 |
_version_ | 1782341353430056960 |
---|---|
author | Kocijancic, Marija Cargonja, Jelena Delic-Knezevic, Alma |
author_facet | Kocijancic, Marija Cargonja, Jelena Delic-Knezevic, Alma |
author_sort | Kocijancic, Marija |
collection | PubMed |
description | INTRODUCTION: Preanalytical variables account for most of laboratory errors. There is a wide range of factors that affect the reliability of laboratory report. Most convenient sample type for routine laboratory analysis is serum. BD Vacutainer(®) Rapid Serum Tube (RST) (Becton, Dickinson and Company, Franklin Lakes, NJ, USA) blood collection tube provides rapid clotting time allowing fast serum separation. Our aim was to evaluate the comparability of routine chemistry parameters in BD Vacutainer(®) RST blood collection tube in reference with the BD Vacutainer(®) Serum Separating Tubes II Advance Tube (SST) (Becton, Dickinson and Company, Franklin Lakes, NJ, USA). MATERIALS AND METHODS: Blood specimens were collected from 90 participants for evaluation on its results, clotting time and stability study of six routine biochemistry parameters: glucose (Glu), aspartate aminotransferase (AST), alanine aminotransferase (ALT), calcium (Ca), lactate dehidrogenase (LD) and potassium (K) measured with Olympus AU2700 analyzer (Beckman Coulter, Tokyo, Japan). The significance of the differences between samples was assessed by paired t-test or Wilcoxon Matched-Pairs Rank test after checking for normality. RESULTS: Clotting process was significantly shorter in the RSTs compared to SSTs (2.49 min vs. 19.47 min, respectively; P < 0.001). There was a statistically significant difference between the RST and SST II tubes for glucose, calcium and LD (P < 0.001). Differences for glucose and LD were also clinically significant. Analyte stability studies showed that all analytes were stable for 24 h at 4 °C. CONCLUSIONS: Most results (except LD and glucose) from RST are comparable with those from SST. In addition, RST tube provides shorter clotting time. |
format | Online Article Text |
id | pubmed-4210257 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2014 |
publisher | Croatian Society of Medical Biochemistry and Laboratory Medicine |
record_format | MEDLINE/PubMed |
spelling | pubmed-42102572014-10-28 Evaluation of the BD Vacutainer(®) RST blood collection tube for routine chemistry analytes: clinical significance of differences and stability study Kocijancic, Marija Cargonja, Jelena Delic-Knezevic, Alma Biochem Med (Zagreb) Original Papers INTRODUCTION: Preanalytical variables account for most of laboratory errors. There is a wide range of factors that affect the reliability of laboratory report. Most convenient sample type for routine laboratory analysis is serum. BD Vacutainer(®) Rapid Serum Tube (RST) (Becton, Dickinson and Company, Franklin Lakes, NJ, USA) blood collection tube provides rapid clotting time allowing fast serum separation. Our aim was to evaluate the comparability of routine chemistry parameters in BD Vacutainer(®) RST blood collection tube in reference with the BD Vacutainer(®) Serum Separating Tubes II Advance Tube (SST) (Becton, Dickinson and Company, Franklin Lakes, NJ, USA). MATERIALS AND METHODS: Blood specimens were collected from 90 participants for evaluation on its results, clotting time and stability study of six routine biochemistry parameters: glucose (Glu), aspartate aminotransferase (AST), alanine aminotransferase (ALT), calcium (Ca), lactate dehidrogenase (LD) and potassium (K) measured with Olympus AU2700 analyzer (Beckman Coulter, Tokyo, Japan). The significance of the differences between samples was assessed by paired t-test or Wilcoxon Matched-Pairs Rank test after checking for normality. RESULTS: Clotting process was significantly shorter in the RSTs compared to SSTs (2.49 min vs. 19.47 min, respectively; P < 0.001). There was a statistically significant difference between the RST and SST II tubes for glucose, calcium and LD (P < 0.001). Differences for glucose and LD were also clinically significant. Analyte stability studies showed that all analytes were stable for 24 h at 4 °C. CONCLUSIONS: Most results (except LD and glucose) from RST are comparable with those from SST. In addition, RST tube provides shorter clotting time. Croatian Society of Medical Biochemistry and Laboratory Medicine 2014-10-15 /pmc/articles/PMC4210257/ /pubmed/25351355 http://dx.doi.org/10.11613/BM.2014.039 Text en © Copyright by Croatian Society of Medical Biochemistry and Laboratory Medicine. This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc-nd/3.0/) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Original Papers Kocijancic, Marija Cargonja, Jelena Delic-Knezevic, Alma Evaluation of the BD Vacutainer(®) RST blood collection tube for routine chemistry analytes: clinical significance of differences and stability study |
title | Evaluation of the BD Vacutainer(®) RST blood collection tube for routine chemistry analytes: clinical significance of differences and stability study |
title_full | Evaluation of the BD Vacutainer(®) RST blood collection tube for routine chemistry analytes: clinical significance of differences and stability study |
title_fullStr | Evaluation of the BD Vacutainer(®) RST blood collection tube for routine chemistry analytes: clinical significance of differences and stability study |
title_full_unstemmed | Evaluation of the BD Vacutainer(®) RST blood collection tube for routine chemistry analytes: clinical significance of differences and stability study |
title_short | Evaluation of the BD Vacutainer(®) RST blood collection tube for routine chemistry analytes: clinical significance of differences and stability study |
title_sort | evaluation of the bd vacutainer(®) rst blood collection tube for routine chemistry analytes: clinical significance of differences and stability study |
topic | Original Papers |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4210257/ https://www.ncbi.nlm.nih.gov/pubmed/25351355 http://dx.doi.org/10.11613/BM.2014.039 |
work_keys_str_mv | AT kocijancicmarija evaluationofthebdvacutainerrstbloodcollectiontubeforroutinechemistryanalytesclinicalsignificanceofdifferencesandstabilitystudy AT cargonjajelena evaluationofthebdvacutainerrstbloodcollectiontubeforroutinechemistryanalytesclinicalsignificanceofdifferencesandstabilitystudy AT delicknezevicalma evaluationofthebdvacutainerrstbloodcollectiontubeforroutinechemistryanalytesclinicalsignificanceofdifferencesandstabilitystudy |