Evaluation of the BD Vacutainer(®) RST blood collection tube for routine chemistry analytes: clinical significance of differences and stability study

INTRODUCTION: Preanalytical variables account for most of laboratory errors. There is a wide range of factors that affect the reliability of laboratory report. Most convenient sample type for routine laboratory analysis is serum. BD Vacutainer(®) Rapid Serum Tube (RST) (Becton, Dickinson and Company...

Descripción completa

Detalles Bibliográficos
Autores principales: Kocijancic, Marija, Cargonja, Jelena, Delic-Knezevic, Alma
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Croatian Society of Medical Biochemistry and Laboratory Medicine 2014
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4210257/
https://www.ncbi.nlm.nih.gov/pubmed/25351355
http://dx.doi.org/10.11613/BM.2014.039
_version_ 1782341353430056960
author Kocijancic, Marija
Cargonja, Jelena
Delic-Knezevic, Alma
author_facet Kocijancic, Marija
Cargonja, Jelena
Delic-Knezevic, Alma
author_sort Kocijancic, Marija
collection PubMed
description INTRODUCTION: Preanalytical variables account for most of laboratory errors. There is a wide range of factors that affect the reliability of laboratory report. Most convenient sample type for routine laboratory analysis is serum. BD Vacutainer(®) Rapid Serum Tube (RST) (Becton, Dickinson and Company, Franklin Lakes, NJ, USA) blood collection tube provides rapid clotting time allowing fast serum separation. Our aim was to evaluate the comparability of routine chemistry parameters in BD Vacutainer(®) RST blood collection tube in reference with the BD Vacutainer(®) Serum Separating Tubes II Advance Tube (SST) (Becton, Dickinson and Company, Franklin Lakes, NJ, USA). MATERIALS AND METHODS: Blood specimens were collected from 90 participants for evaluation on its results, clotting time and stability study of six routine biochemistry parameters: glucose (Glu), aspartate aminotransferase (AST), alanine aminotransferase (ALT), calcium (Ca), lactate dehidrogenase (LD) and potassium (K) measured with Olympus AU2700 analyzer (Beckman Coulter, Tokyo, Japan). The significance of the differences between samples was assessed by paired t-test or Wilcoxon Matched-Pairs Rank test after checking for normality. RESULTS: Clotting process was significantly shorter in the RSTs compared to SSTs (2.49 min vs. 19.47 min, respectively; P < 0.001). There was a statistically significant difference between the RST and SST II tubes for glucose, calcium and LD (P < 0.001). Differences for glucose and LD were also clinically significant. Analyte stability studies showed that all analytes were stable for 24 h at 4 °C. CONCLUSIONS: Most results (except LD and glucose) from RST are comparable with those from SST. In addition, RST tube provides shorter clotting time.
format Online
Article
Text
id pubmed-4210257
institution National Center for Biotechnology Information
language English
publishDate 2014
publisher Croatian Society of Medical Biochemistry and Laboratory Medicine
record_format MEDLINE/PubMed
spelling pubmed-42102572014-10-28 Evaluation of the BD Vacutainer(®) RST blood collection tube for routine chemistry analytes: clinical significance of differences and stability study Kocijancic, Marija Cargonja, Jelena Delic-Knezevic, Alma Biochem Med (Zagreb) Original Papers INTRODUCTION: Preanalytical variables account for most of laboratory errors. There is a wide range of factors that affect the reliability of laboratory report. Most convenient sample type for routine laboratory analysis is serum. BD Vacutainer(®) Rapid Serum Tube (RST) (Becton, Dickinson and Company, Franklin Lakes, NJ, USA) blood collection tube provides rapid clotting time allowing fast serum separation. Our aim was to evaluate the comparability of routine chemistry parameters in BD Vacutainer(®) RST blood collection tube in reference with the BD Vacutainer(®) Serum Separating Tubes II Advance Tube (SST) (Becton, Dickinson and Company, Franklin Lakes, NJ, USA). MATERIALS AND METHODS: Blood specimens were collected from 90 participants for evaluation on its results, clotting time and stability study of six routine biochemistry parameters: glucose (Glu), aspartate aminotransferase (AST), alanine aminotransferase (ALT), calcium (Ca), lactate dehidrogenase (LD) and potassium (K) measured with Olympus AU2700 analyzer (Beckman Coulter, Tokyo, Japan). The significance of the differences between samples was assessed by paired t-test or Wilcoxon Matched-Pairs Rank test after checking for normality. RESULTS: Clotting process was significantly shorter in the RSTs compared to SSTs (2.49 min vs. 19.47 min, respectively; P < 0.001). There was a statistically significant difference between the RST and SST II tubes for glucose, calcium and LD (P < 0.001). Differences for glucose and LD were also clinically significant. Analyte stability studies showed that all analytes were stable for 24 h at 4 °C. CONCLUSIONS: Most results (except LD and glucose) from RST are comparable with those from SST. In addition, RST tube provides shorter clotting time. Croatian Society of Medical Biochemistry and Laboratory Medicine 2014-10-15 /pmc/articles/PMC4210257/ /pubmed/25351355 http://dx.doi.org/10.11613/BM.2014.039 Text en © Copyright by Croatian Society of Medical Biochemistry and Laboratory Medicine. This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc-nd/3.0/) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Original Papers
Kocijancic, Marija
Cargonja, Jelena
Delic-Knezevic, Alma
Evaluation of the BD Vacutainer(®) RST blood collection tube for routine chemistry analytes: clinical significance of differences and stability study
title Evaluation of the BD Vacutainer(®) RST blood collection tube for routine chemistry analytes: clinical significance of differences and stability study
title_full Evaluation of the BD Vacutainer(®) RST blood collection tube for routine chemistry analytes: clinical significance of differences and stability study
title_fullStr Evaluation of the BD Vacutainer(®) RST blood collection tube for routine chemistry analytes: clinical significance of differences and stability study
title_full_unstemmed Evaluation of the BD Vacutainer(®) RST blood collection tube for routine chemistry analytes: clinical significance of differences and stability study
title_short Evaluation of the BD Vacutainer(®) RST blood collection tube for routine chemistry analytes: clinical significance of differences and stability study
title_sort evaluation of the bd vacutainer(®) rst blood collection tube for routine chemistry analytes: clinical significance of differences and stability study
topic Original Papers
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4210257/
https://www.ncbi.nlm.nih.gov/pubmed/25351355
http://dx.doi.org/10.11613/BM.2014.039
work_keys_str_mv AT kocijancicmarija evaluationofthebdvacutainerrstbloodcollectiontubeforroutinechemistryanalytesclinicalsignificanceofdifferencesandstabilitystudy
AT cargonjajelena evaluationofthebdvacutainerrstbloodcollectiontubeforroutinechemistryanalytesclinicalsignificanceofdifferencesandstabilitystudy
AT delicknezevicalma evaluationofthebdvacutainerrstbloodcollectiontubeforroutinechemistryanalytesclinicalsignificanceofdifferencesandstabilitystudy