Cargando…

A study on the mechanical characteristics of the EBM-printed Ti-6Al-4V LCP plates in vitro

PURPOSE: The electron beam melting (EBM) Ti-6Al-4V material technology has been developed over a short time period. It was introduced through a research to develop Ti-6Al-4V implants for patients, but EBM printed locking compression plates have not been used for clinical implants. The main purpose o...

Descripción completa

Detalles Bibliográficos
Autores principales: Liu, Peng-cheng, Yang, Yun-ji, Liu, Run, Shu, He-xi, Gong, Jin-peng, Yang, Yong, Sun, Qi, Wu, Xing, Cai, Ming
Formato: Online Artículo Texto
Lenguaje:English
Publicado: BioMed Central 2014
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4221727/
https://www.ncbi.nlm.nih.gov/pubmed/25370215
http://dx.doi.org/10.1186/s13018-014-0106-3
_version_ 1782342924857507840
author Liu, Peng-cheng
Yang, Yun-ji
Liu, Run
Shu, He-xi
Gong, Jin-peng
Yang, Yong
Sun, Qi
Wu, Xing
Cai, Ming
author_facet Liu, Peng-cheng
Yang, Yun-ji
Liu, Run
Shu, He-xi
Gong, Jin-peng
Yang, Yong
Sun, Qi
Wu, Xing
Cai, Ming
author_sort Liu, Peng-cheng
collection PubMed
description PURPOSE: The electron beam melting (EBM) Ti-6Al-4V material technology has been developed over a short time period. It was introduced through a research to develop Ti-6Al-4V implants for patients, but EBM printed locking compression plates have not been used for clinical implants. The main purpose of this study is to find whether the EBM Ti-6Al-4V plate suit for clinical implants. METHODS: First, we scanned an AO-locking compression plate (LCP) and printed LCP samples using EBM. Next, we evaluated the EBM plate surface roughness through optical microscopy as well as the LCP and EBM plates’ mechanical characteristics using the ASTM standard, which is commonly used to test the mechanical properties of bone plates subject to bending. Each sample was examined using a single-cycle four-point bending test and hardness testing to acquire data on bending stiffness, bending strength, bending structural stiffness, and hardness. RESULTS: The results show significant differences in bending stiffness, bending strength, bending structural stiffness, and hardness between the samples using EBM and the original LCP plates. The EBM-printed samples’ surface roughness was 0.49 ± 0.02 μm. The mean hardness of the LCP sample was 266.67 HV10 ± 5.8, and the EBM-printed sample mean hardness was 341.1 HV10 ± 1.93. The EBM samples’ bending stiffness was 87.67%, which is greater than using the LCP plates’; and the bending strength was 190.7% greater, the bending structural stiffness was 73.2% greater, and the hardness was 27.9% greater. CONCLUSIONS: The results show that the EBM plates’ general mechanical strength was significantly greater than the LCP plates. An EBM plate is advantageous for clinical implants because it can be customized with great potential for improvement.
format Online
Article
Text
id pubmed-4221727
institution National Center for Biotechnology Information
language English
publishDate 2014
publisher BioMed Central
record_format MEDLINE/PubMed
spelling pubmed-42217272014-11-07 A study on the mechanical characteristics of the EBM-printed Ti-6Al-4V LCP plates in vitro Liu, Peng-cheng Yang, Yun-ji Liu, Run Shu, He-xi Gong, Jin-peng Yang, Yong Sun, Qi Wu, Xing Cai, Ming J Orthop Surg Res Research Article PURPOSE: The electron beam melting (EBM) Ti-6Al-4V material technology has been developed over a short time period. It was introduced through a research to develop Ti-6Al-4V implants for patients, but EBM printed locking compression plates have not been used for clinical implants. The main purpose of this study is to find whether the EBM Ti-6Al-4V plate suit for clinical implants. METHODS: First, we scanned an AO-locking compression plate (LCP) and printed LCP samples using EBM. Next, we evaluated the EBM plate surface roughness through optical microscopy as well as the LCP and EBM plates’ mechanical characteristics using the ASTM standard, which is commonly used to test the mechanical properties of bone plates subject to bending. Each sample was examined using a single-cycle four-point bending test and hardness testing to acquire data on bending stiffness, bending strength, bending structural stiffness, and hardness. RESULTS: The results show significant differences in bending stiffness, bending strength, bending structural stiffness, and hardness between the samples using EBM and the original LCP plates. The EBM-printed samples’ surface roughness was 0.49 ± 0.02 μm. The mean hardness of the LCP sample was 266.67 HV10 ± 5.8, and the EBM-printed sample mean hardness was 341.1 HV10 ± 1.93. The EBM samples’ bending stiffness was 87.67%, which is greater than using the LCP plates’; and the bending strength was 190.7% greater, the bending structural stiffness was 73.2% greater, and the hardness was 27.9% greater. CONCLUSIONS: The results show that the EBM plates’ general mechanical strength was significantly greater than the LCP plates. An EBM plate is advantageous for clinical implants because it can be customized with great potential for improvement. BioMed Central 2014-11-05 /pmc/articles/PMC4221727/ /pubmed/25370215 http://dx.doi.org/10.1186/s13018-014-0106-3 Text en © Liu et al.; licensee BioMed Central Ltd. 2014 This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly credited. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.
spellingShingle Research Article
Liu, Peng-cheng
Yang, Yun-ji
Liu, Run
Shu, He-xi
Gong, Jin-peng
Yang, Yong
Sun, Qi
Wu, Xing
Cai, Ming
A study on the mechanical characteristics of the EBM-printed Ti-6Al-4V LCP plates in vitro
title A study on the mechanical characteristics of the EBM-printed Ti-6Al-4V LCP plates in vitro
title_full A study on the mechanical characteristics of the EBM-printed Ti-6Al-4V LCP plates in vitro
title_fullStr A study on the mechanical characteristics of the EBM-printed Ti-6Al-4V LCP plates in vitro
title_full_unstemmed A study on the mechanical characteristics of the EBM-printed Ti-6Al-4V LCP plates in vitro
title_short A study on the mechanical characteristics of the EBM-printed Ti-6Al-4V LCP plates in vitro
title_sort study on the mechanical characteristics of the ebm-printed ti-6al-4v lcp plates in vitro
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4221727/
https://www.ncbi.nlm.nih.gov/pubmed/25370215
http://dx.doi.org/10.1186/s13018-014-0106-3
work_keys_str_mv AT liupengcheng astudyonthemechanicalcharacteristicsoftheebmprintedti6al4vlcpplatesinvitro
AT yangyunji astudyonthemechanicalcharacteristicsoftheebmprintedti6al4vlcpplatesinvitro
AT liurun astudyonthemechanicalcharacteristicsoftheebmprintedti6al4vlcpplatesinvitro
AT shuhexi astudyonthemechanicalcharacteristicsoftheebmprintedti6al4vlcpplatesinvitro
AT gongjinpeng astudyonthemechanicalcharacteristicsoftheebmprintedti6al4vlcpplatesinvitro
AT yangyong astudyonthemechanicalcharacteristicsoftheebmprintedti6al4vlcpplatesinvitro
AT sunqi astudyonthemechanicalcharacteristicsoftheebmprintedti6al4vlcpplatesinvitro
AT wuxing astudyonthemechanicalcharacteristicsoftheebmprintedti6al4vlcpplatesinvitro
AT caiming astudyonthemechanicalcharacteristicsoftheebmprintedti6al4vlcpplatesinvitro
AT liupengcheng studyonthemechanicalcharacteristicsoftheebmprintedti6al4vlcpplatesinvitro
AT yangyunji studyonthemechanicalcharacteristicsoftheebmprintedti6al4vlcpplatesinvitro
AT liurun studyonthemechanicalcharacteristicsoftheebmprintedti6al4vlcpplatesinvitro
AT shuhexi studyonthemechanicalcharacteristicsoftheebmprintedti6al4vlcpplatesinvitro
AT gongjinpeng studyonthemechanicalcharacteristicsoftheebmprintedti6al4vlcpplatesinvitro
AT yangyong studyonthemechanicalcharacteristicsoftheebmprintedti6al4vlcpplatesinvitro
AT sunqi studyonthemechanicalcharacteristicsoftheebmprintedti6al4vlcpplatesinvitro
AT wuxing studyonthemechanicalcharacteristicsoftheebmprintedti6al4vlcpplatesinvitro
AT caiming studyonthemechanicalcharacteristicsoftheebmprintedti6al4vlcpplatesinvitro