Comparison of Anterior Segment Measurements with Scheimpflug/Placido Photography-Based Topography System and IOLMaster Partial Coherence Interferometry in Patients with Cataracts

Purpose. To assess the consistency of anterior segment measurements obtained using a Sirius Scheimpflug/Placido photography-based topography system (CSO, Italy) and IOLMaster partial coherence interferometry (Carl Zeiss Meditec, Germany) in eyes with cataracts. Methods. A total of 90 eyes of 90 pati...

Descripción completa

Detalles Bibliográficos
Autores principales: Huang, Jinhai, Liao, Na, Savini, Giacomo, Bao, Fangjun, Yu, Ye, Lu, Weicong, Hu, Qingjie, Wang, Qinmei
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Hindawi Publishing Corporation 2014
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4226174/
https://www.ncbi.nlm.nih.gov/pubmed/25400939
http://dx.doi.org/10.1155/2014/540760
_version_ 1782343590576390144
author Huang, Jinhai
Liao, Na
Savini, Giacomo
Bao, Fangjun
Yu, Ye
Lu, Weicong
Hu, Qingjie
Wang, Qinmei
author_facet Huang, Jinhai
Liao, Na
Savini, Giacomo
Bao, Fangjun
Yu, Ye
Lu, Weicong
Hu, Qingjie
Wang, Qinmei
author_sort Huang, Jinhai
collection PubMed
description Purpose. To assess the consistency of anterior segment measurements obtained using a Sirius Scheimpflug/Placido photography-based topography system (CSO, Italy) and IOLMaster partial coherence interferometry (Carl Zeiss Meditec, Germany) in eyes with cataracts. Methods. A total of 90 eyes of 90 patients were included in this prospective study. The anterior chamber depth (ACD), keratometry (K), corneal astigmatism axis, and white to white (WTW) values were randomly measured three times with Sirius and IOLMaster. Concordance between them was assessed by calculating 95% limits of agreement (LoA). Results. The ACD and K taken with the Sirius were statistically significantly higher than that taken with the IOLMaster; however, the Sirius significantly underestimated the WTW values compared with the IOLMaster. Good agreement was found for Km and ACD measurements, with 95% LoA of −0.20 to 0.54 mm and −0.16 to 0.34 mm, respectively. Poor agreement was observed for astigmatism axis and WTW measurements, as the 95% LoA was −23.96 to 23.36° and −1.15 to 0.37 mm, respectively. Conclusion. With the exception of astigmatism axis and WTW, anterior segment measurements taken by Sirius and IOLMaster devices showed good agreement and may be used interchangeably in patients with cataracts.
format Online
Article
Text
id pubmed-4226174
institution National Center for Biotechnology Information
language English
publishDate 2014
publisher Hindawi Publishing Corporation
record_format MEDLINE/PubMed
spelling pubmed-42261742014-11-16 Comparison of Anterior Segment Measurements with Scheimpflug/Placido Photography-Based Topography System and IOLMaster Partial Coherence Interferometry in Patients with Cataracts Huang, Jinhai Liao, Na Savini, Giacomo Bao, Fangjun Yu, Ye Lu, Weicong Hu, Qingjie Wang, Qinmei J Ophthalmol Research Article Purpose. To assess the consistency of anterior segment measurements obtained using a Sirius Scheimpflug/Placido photography-based topography system (CSO, Italy) and IOLMaster partial coherence interferometry (Carl Zeiss Meditec, Germany) in eyes with cataracts. Methods. A total of 90 eyes of 90 patients were included in this prospective study. The anterior chamber depth (ACD), keratometry (K), corneal astigmatism axis, and white to white (WTW) values were randomly measured three times with Sirius and IOLMaster. Concordance between them was assessed by calculating 95% limits of agreement (LoA). Results. The ACD and K taken with the Sirius were statistically significantly higher than that taken with the IOLMaster; however, the Sirius significantly underestimated the WTW values compared with the IOLMaster. Good agreement was found for Km and ACD measurements, with 95% LoA of −0.20 to 0.54 mm and −0.16 to 0.34 mm, respectively. Poor agreement was observed for astigmatism axis and WTW measurements, as the 95% LoA was −23.96 to 23.36° and −1.15 to 0.37 mm, respectively. Conclusion. With the exception of astigmatism axis and WTW, anterior segment measurements taken by Sirius and IOLMaster devices showed good agreement and may be used interchangeably in patients with cataracts. Hindawi Publishing Corporation 2014 2014-10-23 /pmc/articles/PMC4226174/ /pubmed/25400939 http://dx.doi.org/10.1155/2014/540760 Text en Copyright © 2014 Jinhai Huang et al. https://creativecommons.org/licenses/by/3.0/ This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Research Article
Huang, Jinhai
Liao, Na
Savini, Giacomo
Bao, Fangjun
Yu, Ye
Lu, Weicong
Hu, Qingjie
Wang, Qinmei
Comparison of Anterior Segment Measurements with Scheimpflug/Placido Photography-Based Topography System and IOLMaster Partial Coherence Interferometry in Patients with Cataracts
title Comparison of Anterior Segment Measurements with Scheimpflug/Placido Photography-Based Topography System and IOLMaster Partial Coherence Interferometry in Patients with Cataracts
title_full Comparison of Anterior Segment Measurements with Scheimpflug/Placido Photography-Based Topography System and IOLMaster Partial Coherence Interferometry in Patients with Cataracts
title_fullStr Comparison of Anterior Segment Measurements with Scheimpflug/Placido Photography-Based Topography System and IOLMaster Partial Coherence Interferometry in Patients with Cataracts
title_full_unstemmed Comparison of Anterior Segment Measurements with Scheimpflug/Placido Photography-Based Topography System and IOLMaster Partial Coherence Interferometry in Patients with Cataracts
title_short Comparison of Anterior Segment Measurements with Scheimpflug/Placido Photography-Based Topography System and IOLMaster Partial Coherence Interferometry in Patients with Cataracts
title_sort comparison of anterior segment measurements with scheimpflug/placido photography-based topography system and iolmaster partial coherence interferometry in patients with cataracts
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4226174/
https://www.ncbi.nlm.nih.gov/pubmed/25400939
http://dx.doi.org/10.1155/2014/540760
work_keys_str_mv AT huangjinhai comparisonofanteriorsegmentmeasurementswithscheimpflugplacidophotographybasedtopographysystemandiolmasterpartialcoherenceinterferometryinpatientswithcataracts
AT liaona comparisonofanteriorsegmentmeasurementswithscheimpflugplacidophotographybasedtopographysystemandiolmasterpartialcoherenceinterferometryinpatientswithcataracts
AT savinigiacomo comparisonofanteriorsegmentmeasurementswithscheimpflugplacidophotographybasedtopographysystemandiolmasterpartialcoherenceinterferometryinpatientswithcataracts
AT baofangjun comparisonofanteriorsegmentmeasurementswithscheimpflugplacidophotographybasedtopographysystemandiolmasterpartialcoherenceinterferometryinpatientswithcataracts
AT yuye comparisonofanteriorsegmentmeasurementswithscheimpflugplacidophotographybasedtopographysystemandiolmasterpartialcoherenceinterferometryinpatientswithcataracts
AT luweicong comparisonofanteriorsegmentmeasurementswithscheimpflugplacidophotographybasedtopographysystemandiolmasterpartialcoherenceinterferometryinpatientswithcataracts
AT huqingjie comparisonofanteriorsegmentmeasurementswithscheimpflugplacidophotographybasedtopographysystemandiolmasterpartialcoherenceinterferometryinpatientswithcataracts
AT wangqinmei comparisonofanteriorsegmentmeasurementswithscheimpflugplacidophotographybasedtopographysystemandiolmasterpartialcoherenceinterferometryinpatientswithcataracts