Accuracy of noninvasive hemoglobin and invasive point-of-care hemoglobin testing compared with a laboratory analyzer

INTRODUCTION: Hemoglobin concentration is assessed to detect anemia and its associated morbidities. Hemoglobin is usually determined from venous or capillary blood samples run on a laboratory analyzer. However, this method requires a needle stick and results can be delayed. It also exposes caregiver...

Descripción completa

Detalles Bibliográficos
Autores principales: Shah, N, Osea, E A, Martinez, G J
Formato: Online Artículo Texto
Lenguaje:English
Publicado: BlackWell Publishing Ltd 2014
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4232003/
https://www.ncbi.nlm.nih.gov/pubmed/23809685
http://dx.doi.org/10.1111/ijlh.12118
_version_ 1782344512546275328
author Shah, N
Osea, E A
Martinez, G J
author_facet Shah, N
Osea, E A
Martinez, G J
author_sort Shah, N
collection PubMed
description INTRODUCTION: Hemoglobin concentration is assessed to detect anemia and its associated morbidities. Hemoglobin is usually determined from venous or capillary blood samples run on a laboratory analyzer. However, this method requires a needle stick and results can be delayed. It also exposes caregivers to risks associated with needle sticks and blood exposure. Noninvasive hemoglobin determination would be of benefit to patients and caregivers because it would allow for quick and painless point-of-care assessment. METHODS: Hemoglobin determination from a noninvasive spot check hemoglobin device (Pronto-7 with SpHb, Masimo) and an invasive point-of-care device (HemoCue) was compared with venous blood samples run on a laboratory hematology analyzer. RESULTS: A total of 440 outpatients and healthy volunteers were included (mean age 36 years, 62% female). Compared with the hematology analyzer, the bias ± standard deviation of was −0.1 ± 1.1 g/dL for SpHb and −0.1 ± 1.6 g/dL for HemoCue. CONCLUSION: Noninvasive hemoglobin testing with SpHb provided similar accuracy as invasive point-of-care hemoglobin testing and may enable more efficient and effective patient care.
format Online
Article
Text
id pubmed-4232003
institution National Center for Biotechnology Information
language English
publishDate 2014
publisher BlackWell Publishing Ltd
record_format MEDLINE/PubMed
spelling pubmed-42320032014-12-15 Accuracy of noninvasive hemoglobin and invasive point-of-care hemoglobin testing compared with a laboratory analyzer Shah, N Osea, E A Martinez, G J Int J Lab Hematol Original Articles INTRODUCTION: Hemoglobin concentration is assessed to detect anemia and its associated morbidities. Hemoglobin is usually determined from venous or capillary blood samples run on a laboratory analyzer. However, this method requires a needle stick and results can be delayed. It also exposes caregivers to risks associated with needle sticks and blood exposure. Noninvasive hemoglobin determination would be of benefit to patients and caregivers because it would allow for quick and painless point-of-care assessment. METHODS: Hemoglobin determination from a noninvasive spot check hemoglobin device (Pronto-7 with SpHb, Masimo) and an invasive point-of-care device (HemoCue) was compared with venous blood samples run on a laboratory hematology analyzer. RESULTS: A total of 440 outpatients and healthy volunteers were included (mean age 36 years, 62% female). Compared with the hematology analyzer, the bias ± standard deviation of was −0.1 ± 1.1 g/dL for SpHb and −0.1 ± 1.6 g/dL for HemoCue. CONCLUSION: Noninvasive hemoglobin testing with SpHb provided similar accuracy as invasive point-of-care hemoglobin testing and may enable more efficient and effective patient care. BlackWell Publishing Ltd 2014-02 2013-06-27 /pmc/articles/PMC4232003/ /pubmed/23809685 http://dx.doi.org/10.1111/ijlh.12118 Text en © 2013 The Authors. International Journal of Laboratory Hematology published by John Wiley & Sons Ltd. http://creativecommons.org/licenses/by-nc/3.0/ This is an open access article under the terms of the Creative Commons Attribution-NonCommercial License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited and is not used for commercial purposes.
spellingShingle Original Articles
Shah, N
Osea, E A
Martinez, G J
Accuracy of noninvasive hemoglobin and invasive point-of-care hemoglobin testing compared with a laboratory analyzer
title Accuracy of noninvasive hemoglobin and invasive point-of-care hemoglobin testing compared with a laboratory analyzer
title_full Accuracy of noninvasive hemoglobin and invasive point-of-care hemoglobin testing compared with a laboratory analyzer
title_fullStr Accuracy of noninvasive hemoglobin and invasive point-of-care hemoglobin testing compared with a laboratory analyzer
title_full_unstemmed Accuracy of noninvasive hemoglobin and invasive point-of-care hemoglobin testing compared with a laboratory analyzer
title_short Accuracy of noninvasive hemoglobin and invasive point-of-care hemoglobin testing compared with a laboratory analyzer
title_sort accuracy of noninvasive hemoglobin and invasive point-of-care hemoglobin testing compared with a laboratory analyzer
topic Original Articles
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4232003/
https://www.ncbi.nlm.nih.gov/pubmed/23809685
http://dx.doi.org/10.1111/ijlh.12118
work_keys_str_mv AT shahn accuracyofnoninvasivehemoglobinandinvasivepointofcarehemoglobintestingcomparedwithalaboratoryanalyzer
AT oseaea accuracyofnoninvasivehemoglobinandinvasivepointofcarehemoglobintestingcomparedwithalaboratoryanalyzer
AT martinezgj accuracyofnoninvasivehemoglobinandinvasivepointofcarehemoglobintestingcomparedwithalaboratoryanalyzer