Effect of cyclic load on vertical misfit of prefabricated and cast implant single abutment

OBJECTIVE: The purpose of this in vitro study was to evaluate misfit alterations at the implant/abutment interface of external and internal connection implant systems when subjected to cyclic loading. MATERIAL AND METHODS: Standard metal crowns were fabricated for 5 groups (n=10) of implant/abutment...

Descripción completa

Detalles Bibliográficos
Autores principales: DE JESUS TAVAREZ, Rudys Rodolfo, BONACHELA, Wellington Cardoso, XIBLE, Anuar Antônio
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Faculdade de Odontologia de Bauru da Universidade de São Paulo 2011
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4245858/
https://www.ncbi.nlm.nih.gov/pubmed/21437464
http://dx.doi.org/10.1590/S1678-77572011000100005
_version_ 1782346437290360832
author DE JESUS TAVAREZ, Rudys Rodolfo
BONACHELA, Wellington Cardoso
XIBLE, Anuar Antônio
author_facet DE JESUS TAVAREZ, Rudys Rodolfo
BONACHELA, Wellington Cardoso
XIBLE, Anuar Antônio
author_sort DE JESUS TAVAREZ, Rudys Rodolfo
collection PubMed
description OBJECTIVE: The purpose of this in vitro study was to evaluate misfit alterations at the implant/abutment interface of external and internal connection implant systems when subjected to cyclic loading. MATERIAL AND METHODS: Standard metal crowns were fabricated for 5 groups (n=10) of implant/abutment assemblies: Group 1, external hexagon implant and UCLA cast-on premachined abutment; Group 2, internal hexagon implant and premachined abutment; Group 3, internal octagon implant and prefabricated abutment; Group 4, external hexagon implant and UCLA cast-on premachined abutment; and Group 5, external hexagon implant and Ceraone abutment. For groups 1, 2, 3 and 5, the crowns were cemented on the abutments and in group 4 crowns were screwed directly on the implant. The specimens were subjected to 500,000 cycles at 19.1 Hz of frequency and non-axial load of 133 N in a MTS 810 machine. The vertical misfit (μm) at the implant/abutment interface was evaluated before (B) and after (A) application of the cyclic loading. Data were analyzed statistically by using two-away ANOVA and Tukey’s post-hoc test (p<0.05). RESULTS: Before loading values showed no difference among groups 2 (4.33±3.13), 3 (4.79±3.43) and 5 (3.86±4.60); between groups 1 (12.88±6.43) and 4 (9.67±3.08), and among groups 2, 3 and 4. However, groups 1 and 4 were significantly different from groups 2, 3 and 5. After loading values of groups 1 (17.28±8.77) and 4 (17.78±10.99) were significantly different from those of groups 2 (4.83±4.50), 3 (8.07±4.31) and 5 (3.81±4.84). There was a significant increase in misfit values of groups 1, 3 and 4 after cyclic loading, but not for groups 2 and 5. CONCLUSION: The cyclic loading and type of implant/abutment connection may develop a role on the vertical misfit at the implant/abutment interface.
format Online
Article
Text
id pubmed-4245858
institution National Center for Biotechnology Information
language English
publishDate 2011
publisher Faculdade de Odontologia de Bauru da Universidade de São Paulo
record_format MEDLINE/PubMed
spelling pubmed-42458582014-11-28 Effect of cyclic load on vertical misfit of prefabricated and cast implant single abutment DE JESUS TAVAREZ, Rudys Rodolfo BONACHELA, Wellington Cardoso XIBLE, Anuar Antônio J Appl Oral Sci Original Articles OBJECTIVE: The purpose of this in vitro study was to evaluate misfit alterations at the implant/abutment interface of external and internal connection implant systems when subjected to cyclic loading. MATERIAL AND METHODS: Standard metal crowns were fabricated for 5 groups (n=10) of implant/abutment assemblies: Group 1, external hexagon implant and UCLA cast-on premachined abutment; Group 2, internal hexagon implant and premachined abutment; Group 3, internal octagon implant and prefabricated abutment; Group 4, external hexagon implant and UCLA cast-on premachined abutment; and Group 5, external hexagon implant and Ceraone abutment. For groups 1, 2, 3 and 5, the crowns were cemented on the abutments and in group 4 crowns were screwed directly on the implant. The specimens were subjected to 500,000 cycles at 19.1 Hz of frequency and non-axial load of 133 N in a MTS 810 machine. The vertical misfit (μm) at the implant/abutment interface was evaluated before (B) and after (A) application of the cyclic loading. Data were analyzed statistically by using two-away ANOVA and Tukey’s post-hoc test (p<0.05). RESULTS: Before loading values showed no difference among groups 2 (4.33±3.13), 3 (4.79±3.43) and 5 (3.86±4.60); between groups 1 (12.88±6.43) and 4 (9.67±3.08), and among groups 2, 3 and 4. However, groups 1 and 4 were significantly different from groups 2, 3 and 5. After loading values of groups 1 (17.28±8.77) and 4 (17.78±10.99) were significantly different from those of groups 2 (4.83±4.50), 3 (8.07±4.31) and 5 (3.81±4.84). There was a significant increase in misfit values of groups 1, 3 and 4 after cyclic loading, but not for groups 2 and 5. CONCLUSION: The cyclic loading and type of implant/abutment connection may develop a role on the vertical misfit at the implant/abutment interface. Faculdade de Odontologia de Bauru da Universidade de São Paulo 2011 /pmc/articles/PMC4245858/ /pubmed/21437464 http://dx.doi.org/10.1590/S1678-77572011000100005 Text en http://creativecommons.org/licenses/by-nc/3.0/ This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Original Articles
DE JESUS TAVAREZ, Rudys Rodolfo
BONACHELA, Wellington Cardoso
XIBLE, Anuar Antônio
Effect of cyclic load on vertical misfit of prefabricated and cast implant single abutment
title Effect of cyclic load on vertical misfit of prefabricated and cast implant single abutment
title_full Effect of cyclic load on vertical misfit of prefabricated and cast implant single abutment
title_fullStr Effect of cyclic load on vertical misfit of prefabricated and cast implant single abutment
title_full_unstemmed Effect of cyclic load on vertical misfit of prefabricated and cast implant single abutment
title_short Effect of cyclic load on vertical misfit of prefabricated and cast implant single abutment
title_sort effect of cyclic load on vertical misfit of prefabricated and cast implant single abutment
topic Original Articles
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4245858/
https://www.ncbi.nlm.nih.gov/pubmed/21437464
http://dx.doi.org/10.1590/S1678-77572011000100005
work_keys_str_mv AT dejesustavarezrudysrodolfo effectofcyclicloadonverticalmisfitofprefabricatedandcastimplantsingleabutment
AT bonachelawellingtoncardoso effectofcyclicloadonverticalmisfitofprefabricatedandcastimplantsingleabutment
AT xibleanuarantonio effectofcyclicloadonverticalmisfitofprefabricatedandcastimplantsingleabutment