Accuracy of corneal trephination depth using the Moria single-use adjustable depth vacuum trephine system
BACKGROUND: The Moria single-use adjustable depth trephine is a device that allows a goal trephination depth to be set prior to the surgical procedure. METHODS: Eleven fresh human cadaveric eyes were trephined using 8.0 mm Moria single-use adjustable vacuum trephines. Prior to trephination, the aver...
Autores principales: | , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Dove Medical Press
2014
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4251574/ https://www.ncbi.nlm.nih.gov/pubmed/25473266 http://dx.doi.org/10.2147/OPTH.S73591 |
_version_ | 1782347072220954624 |
---|---|
author | Fenzl, Carlton R Gess, Adam J Moshirfar, Majid |
author_facet | Fenzl, Carlton R Gess, Adam J Moshirfar, Majid |
author_sort | Fenzl, Carlton R |
collection | PubMed |
description | BACKGROUND: The Moria single-use adjustable depth trephine is a device that allows a goal trephination depth to be set prior to the surgical procedure. METHODS: Eleven fresh human cadaveric eyes were trephined using 8.0 mm Moria single-use adjustable vacuum trephines. Prior to trephination, the average corneal pachymetry in the peripheral 7–10 mm range was obtained using anterior segment optical coherence tomography. The trephination depth was set to 80% of that value. Light microscopy was used to image anteroposterior cross-sections of each corneal specimen. Digital protractor software was used to evaluate the trephination angle, depth, and length. All adequately processed specimens were included in the analysis. In addition, trephination angle data from a previous publication by Moshirfar et al were used as a comparison with those of this study. RESULTS: Trephination analysis of depth compared with pachymetry revealed a mean of 83.7%±6.53% (95% confidence interval 79.8–87.6). Maximum and minimum trephined depths were 95.35% and 71.3%, respectively. Trephination depth compared with angular corneal thickness yielded a mean of 66.2%±4.79% (95% confidence interval 63.0–69.4). Maximum and minimum depths were 73.7% and 59.7%, respectively. Analysis of trephination angle yielded a mean of 130.2±3.57 degrees (95% confidence interval 127.8–132.61). Maximum and minimum angles were 135.5 degrees and 126 degrees, respectively. The standard deviation of the trephination angle of the Moria trephine was found to be significantly less than that of Hessburg-Barron and Hanna trephines calculated in the previous study. CONCLUSION: The Moria adjustable vacuum trephine is an accurate method of trephination when a specific depth is desired. Further investigation is needed to determine the relevance of this in relation to deep anterior lamellar keratoplasty. |
format | Online Article Text |
id | pubmed-4251574 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2014 |
publisher | Dove Medical Press |
record_format | MEDLINE/PubMed |
spelling | pubmed-42515742014-12-03 Accuracy of corneal trephination depth using the Moria single-use adjustable depth vacuum trephine system Fenzl, Carlton R Gess, Adam J Moshirfar, Majid Clin Ophthalmol Original Research BACKGROUND: The Moria single-use adjustable depth trephine is a device that allows a goal trephination depth to be set prior to the surgical procedure. METHODS: Eleven fresh human cadaveric eyes were trephined using 8.0 mm Moria single-use adjustable vacuum trephines. Prior to trephination, the average corneal pachymetry in the peripheral 7–10 mm range was obtained using anterior segment optical coherence tomography. The trephination depth was set to 80% of that value. Light microscopy was used to image anteroposterior cross-sections of each corneal specimen. Digital protractor software was used to evaluate the trephination angle, depth, and length. All adequately processed specimens were included in the analysis. In addition, trephination angle data from a previous publication by Moshirfar et al were used as a comparison with those of this study. RESULTS: Trephination analysis of depth compared with pachymetry revealed a mean of 83.7%±6.53% (95% confidence interval 79.8–87.6). Maximum and minimum trephined depths were 95.35% and 71.3%, respectively. Trephination depth compared with angular corneal thickness yielded a mean of 66.2%±4.79% (95% confidence interval 63.0–69.4). Maximum and minimum depths were 73.7% and 59.7%, respectively. Analysis of trephination angle yielded a mean of 130.2±3.57 degrees (95% confidence interval 127.8–132.61). Maximum and minimum angles were 135.5 degrees and 126 degrees, respectively. The standard deviation of the trephination angle of the Moria trephine was found to be significantly less than that of Hessburg-Barron and Hanna trephines calculated in the previous study. CONCLUSION: The Moria adjustable vacuum trephine is an accurate method of trephination when a specific depth is desired. Further investigation is needed to determine the relevance of this in relation to deep anterior lamellar keratoplasty. Dove Medical Press 2014-11-27 /pmc/articles/PMC4251574/ /pubmed/25473266 http://dx.doi.org/10.2147/OPTH.S73591 Text en © 2014 Fenzl et al. This work is published by Dove Medical Press Limited, and licensed under Creative Commons Attribution – Non Commercial (unported, v3.0) License The full terms of the License are available at http://creativecommons.org/licenses/by-nc/3.0/. Non-commercial uses of the work are permitted without any further permission from Dove Medical Press Limited, provided the work is properly attributed. |
spellingShingle | Original Research Fenzl, Carlton R Gess, Adam J Moshirfar, Majid Accuracy of corneal trephination depth using the Moria single-use adjustable depth vacuum trephine system |
title | Accuracy of corneal trephination depth using the Moria single-use adjustable depth vacuum trephine system |
title_full | Accuracy of corneal trephination depth using the Moria single-use adjustable depth vacuum trephine system |
title_fullStr | Accuracy of corneal trephination depth using the Moria single-use adjustable depth vacuum trephine system |
title_full_unstemmed | Accuracy of corneal trephination depth using the Moria single-use adjustable depth vacuum trephine system |
title_short | Accuracy of corneal trephination depth using the Moria single-use adjustable depth vacuum trephine system |
title_sort | accuracy of corneal trephination depth using the moria single-use adjustable depth vacuum trephine system |
topic | Original Research |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4251574/ https://www.ncbi.nlm.nih.gov/pubmed/25473266 http://dx.doi.org/10.2147/OPTH.S73591 |
work_keys_str_mv | AT fenzlcarltonr accuracyofcornealtrephinationdepthusingthemoriasingleuseadjustabledepthvacuumtrephinesystem AT gessadamj accuracyofcornealtrephinationdepthusingthemoriasingleuseadjustabledepthvacuumtrephinesystem AT moshirfarmajid accuracyofcornealtrephinationdepthusingthemoriasingleuseadjustabledepthvacuumtrephinesystem |