Soft, Brown Rupture: Clinical Signs and Symptoms Associated with Ruptured PIP Breast Implants
BACKGROUND: Preoperative signs and symptoms of patients with Poly Implant Prothese (PIP) implants could be predictive of device failure. Based on clinical observation and intraoperative findings 4 hypotheses were raised: (1) Preoperative clinical signs including acquired asymmetry, breast enlargemen...
Autores principales: | , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Wolters Kluwer Health
2014
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4255892/ https://www.ncbi.nlm.nih.gov/pubmed/25506532 http://dx.doi.org/10.1097/GOX.0000000000000212 |
_version_ | 1782347506637602816 |
---|---|
author | Godwin, Yvette Duncan, Robert T. Feig, Christine Reintals, Michelle Hill, Sarah |
author_facet | Godwin, Yvette Duncan, Robert T. Feig, Christine Reintals, Michelle Hill, Sarah |
author_sort | Godwin, Yvette |
collection | PubMed |
description | BACKGROUND: Preoperative signs and symptoms of patients with Poly Implant Prothese (PIP) implants could be predictive of device failure. Based on clinical observation and intraoperative findings 4 hypotheses were raised: (1) Preoperative clinical signs including acquired asymmetry, breast enlargement, fullness of the lower pole, decreased mound projection, and change in breast consistency could be indicative of implant rupture. (2) Device failure correlates with a low preoperative Baker grade of capsule. (3) Brown-stained implants are more prone to implant failure. (4) The brown gel could be indicative of iodine ingression through a substandard elastomer shell. METHODS: Preoperative clinical signs were compared with intraoperative findings for 27 patients undergoing PIP implant explantation. RESULTS: Acquired asymmetry (P = 0.0003), breast enlargement (P = 0.0002), fuller lower pole (P < 0.0001), and loss of lateral projection (P < 0.0001) were all significantly predictive of device failure. Capsule Baker grade was lower preoperatively for ruptured implants. The lack of palpable and visible preoperative capsular contracture could be secondary to the elastic nature of the capsular tissue found. Brown implants failed significantly more often than white implants. Analysis of brown gel revealed the presence of iodine, suggesting povidone iodine ingression at implantation. CONCLUSIONS: Preoperative signs can be predictive of PIP implant failure. Brown-stained implants are more prone to rupture. The presence of iodine in the gel suggests unacceptable permeability of the shell early in the implant’s life span. A noninvasive screening test to detect brown implants in situ could help identify implants at risk of failure in those who elect to keep their implants. |
format | Online Article Text |
id | pubmed-4255892 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2014 |
publisher | Wolters Kluwer Health |
record_format | MEDLINE/PubMed |
spelling | pubmed-42558922014-12-12 Soft, Brown Rupture: Clinical Signs and Symptoms Associated with Ruptured PIP Breast Implants Godwin, Yvette Duncan, Robert T. Feig, Christine Reintals, Michelle Hill, Sarah Plast Reconstr Surg Glob Open Original Articles BACKGROUND: Preoperative signs and symptoms of patients with Poly Implant Prothese (PIP) implants could be predictive of device failure. Based on clinical observation and intraoperative findings 4 hypotheses were raised: (1) Preoperative clinical signs including acquired asymmetry, breast enlargement, fullness of the lower pole, decreased mound projection, and change in breast consistency could be indicative of implant rupture. (2) Device failure correlates with a low preoperative Baker grade of capsule. (3) Brown-stained implants are more prone to implant failure. (4) The brown gel could be indicative of iodine ingression through a substandard elastomer shell. METHODS: Preoperative clinical signs were compared with intraoperative findings for 27 patients undergoing PIP implant explantation. RESULTS: Acquired asymmetry (P = 0.0003), breast enlargement (P = 0.0002), fuller lower pole (P < 0.0001), and loss of lateral projection (P < 0.0001) were all significantly predictive of device failure. Capsule Baker grade was lower preoperatively for ruptured implants. The lack of palpable and visible preoperative capsular contracture could be secondary to the elastic nature of the capsular tissue found. Brown implants failed significantly more often than white implants. Analysis of brown gel revealed the presence of iodine, suggesting povidone iodine ingression at implantation. CONCLUSIONS: Preoperative signs can be predictive of PIP implant failure. Brown-stained implants are more prone to rupture. The presence of iodine in the gel suggests unacceptable permeability of the shell early in the implant’s life span. A noninvasive screening test to detect brown implants in situ could help identify implants at risk of failure in those who elect to keep their implants. Wolters Kluwer Health 2014-12-05 /pmc/articles/PMC4255892/ /pubmed/25506532 http://dx.doi.org/10.1097/GOX.0000000000000212 Text en Copyright © 2014 The Authors. Published by Lippincott Williams & Wilkins on behalf of The American Society of Plastic Surgeons. PRS Global Open is a publication of the American Society of Plastic Surgeons. This is an open-access article distributed under the terms of the Creative Commons Attribution-NonCommercial-NoDerivatives 3.0 License, where it is permissible to download and share the work provided it is properly cited. The work cannot be changed in any way or used commercially. |
spellingShingle | Original Articles Godwin, Yvette Duncan, Robert T. Feig, Christine Reintals, Michelle Hill, Sarah Soft, Brown Rupture: Clinical Signs and Symptoms Associated with Ruptured PIP Breast Implants |
title | Soft, Brown Rupture: Clinical Signs and Symptoms Associated with Ruptured PIP Breast Implants |
title_full | Soft, Brown Rupture: Clinical Signs and Symptoms Associated with Ruptured PIP Breast Implants |
title_fullStr | Soft, Brown Rupture: Clinical Signs and Symptoms Associated with Ruptured PIP Breast Implants |
title_full_unstemmed | Soft, Brown Rupture: Clinical Signs and Symptoms Associated with Ruptured PIP Breast Implants |
title_short | Soft, Brown Rupture: Clinical Signs and Symptoms Associated with Ruptured PIP Breast Implants |
title_sort | soft, brown rupture: clinical signs and symptoms associated with ruptured pip breast implants |
topic | Original Articles |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4255892/ https://www.ncbi.nlm.nih.gov/pubmed/25506532 http://dx.doi.org/10.1097/GOX.0000000000000212 |
work_keys_str_mv | AT godwinyvette softbrownruptureclinicalsignsandsymptomsassociatedwithrupturedpipbreastimplants AT duncanrobertt softbrownruptureclinicalsignsandsymptomsassociatedwithrupturedpipbreastimplants AT feigchristine softbrownruptureclinicalsignsandsymptomsassociatedwithrupturedpipbreastimplants AT reintalsmichelle softbrownruptureclinicalsignsandsymptomsassociatedwithrupturedpipbreastimplants AT hillsarah softbrownruptureclinicalsignsandsymptomsassociatedwithrupturedpipbreastimplants |