Influence of increment of gantry angle and number of arcs on esophageal volumetric modulated arc therapy planning in Monaco planning system: A planning study
The objective of this study was to analyze the influence of the increment of gantry angle and the number of arcs on esophageal volumetric modulated arc therapy plan. All plans were done in Monaco planning system for Elekta Synergy linear accelerator with 80 multileaf collimator (MLC). Volumetric mod...
Autores principales: | Nithya, L., Raj, N. Arunai Nambi, Rathinamuthu, Sasikumar, Sharma, Kanika, Pandey, Manish Bhushan |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Medknow Publications & Media Pvt Ltd
2014
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4258731/ https://www.ncbi.nlm.nih.gov/pubmed/25525311 http://dx.doi.org/10.4103/0971-6203.144488 |
Ejemplares similares
-
Comparative analysis of volumetric-modulated arc therapy and intensity-modulated radiotherapy for base of tongue cancer
por: Nithya, L., et al.
Publicado: (2014) -
Validation of the relative insensitivity of volumetric-modulated arc therapy (VMAT) plan quality to gantry space resolution
por: Murtaza, Ghulam, et al.
Publicado: (2017) -
Analyzing the characteristics of 6 MV photon beam at low monitor unit settings
por: Nithya, L., et al.
Publicado: (2016) -
The effect of gantry spacing resolution on plan quality in a single modulated arc optimization
por: Mihaylov, Ivaylo B., et al.
Publicado: (2011) -
RADI-02. Hippocampal-sparing whole brain volumetric modulated arc therapy (VMAT) planning in Monaco: a “How-to” not pull your hair out
por: Goss, Matthew, et al.
Publicado: (2021)