Development of Chemiluminescent Lateral Flow Assay for the Detection of Nucleic Acids

Rapid, sensitive detection methods are of utmost importance for the identification of pathogens related to health and safety. Herein we report the development of a nucleic acid sequence-based lateral flow assay which achieves a low limit of detection using chemiluminescence. On-membrane enzymatic si...

Descripción completa

Detalles Bibliográficos
Autores principales: Wang, Yuhong, Fill, Catherine, Nugen, Sam R.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2012
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4263543/
https://www.ncbi.nlm.nih.gov/pubmed/25585630
http://dx.doi.org/10.3390/bios2010032
Descripción
Sumario:Rapid, sensitive detection methods are of utmost importance for the identification of pathogens related to health and safety. Herein we report the development of a nucleic acid sequence-based lateral flow assay which achieves a low limit of detection using chemiluminescence. On-membrane enzymatic signal amplification is used to reduce the limit of detection to the sub-femtomol level. To demonstrate this assay, we detected synthetic nucleic acid sequences representative of Trypanosoma mRNA, the causative agent for African sleeping sickness, with relevance in human and animal health in sub-Saharan Africa. The intensity of the chemiluminescent signal was evaluated by using a charge-coupled device as well as a microtiter plate reader. We demonstrated that our lateral flow chemiluminescent assay has a very low limit of detection and is easy to use. The limit of detection was determined to be 0.5 fmols of nucleic acid target.