Changes in endotracheal tube cuff pressure in mechanically ventilated adult patients
During mechanical ventilation, endotracheal tube cuff pressure should be maintained within proper range. We investigated the effect of frequent adjustment on cuff pressure in 27 mechanically ventilated patients. Cuff pressure was recorded every 2 h and was adjusted to 24 cmH(2)O each time. We found...
Autores principales: | , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
BioMed Central
2014
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4267542/ https://www.ncbi.nlm.nih.gov/pubmed/25520824 http://dx.doi.org/10.1186/2052-0492-2-7 |
Sumario: | During mechanical ventilation, endotracheal tube cuff pressure should be maintained within proper range. We investigated the effect of frequent adjustment on cuff pressure in 27 mechanically ventilated patients. Cuff pressure was recorded every 2 h and was adjusted to 24 cmH(2)O each time. We found that cuff pressure was decreased by 4.9 ± 2.9 cmH(2)O from the target value. Cuff pressure decreased to less than 20 cmH(2)O in 45% of measurement occasions 2 h after adjusting it to 24 cmH(2)O. |
---|