“First experience with JenaValve™: a single-centre cohort”
AIMS: Since the introduction of transcatheter aortic valve implantation (TAVI), newer generation and novel devices such as the retrievable JenaValve™ have been developed. We evaluated the procedural and 6-month results of our first experience with implantation of the JenaValve™. METHODS AND RESULTS:...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Bohn Stafleu van Loghum
2014
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4268214/ https://www.ncbi.nlm.nih.gov/pubmed/25326104 http://dx.doi.org/10.1007/s12471-014-0619-8 |
Sumario: | AIMS: Since the introduction of transcatheter aortic valve implantation (TAVI), newer generation and novel devices such as the retrievable JenaValve™ have been developed. We evaluated the procedural and 6-month results of our first experience with implantation of the JenaValve™. METHODS AND RESULTS: From June 2012 to December 2013, 24 consecutive patients (mean age 80 ± 7 years, 42 % male) underwent an elective transapical TAVI with the JenaValve™. Device success was 88 %. The mortality rate was 4 % at 30 days and 31 % at 6 months. TAVI reduced the mean transvalvular gradient (44.2 ± 11.1 mmHg vs. 12.3 ± 4.3 mmHg, p < 0.001) and increased the mean aortic valve area (0.8 3 ± 0.23 to 1.70 ± 0.44 cm(2)). A mild paravalvular leakage (PVL) occurred in 4 patients (18 %) and a moderate PVL in 1 patient (4 %). Mean New York Heart Association Functional Class improved from 2.9 ± 0.5 to 2.0 ± 0.8 at 30 days. CONCLUSION: TAVI using the JenaValve™ prosthesis seems adequate and safe in this first experience cohort. |
---|