Finger volume pulse waveforms facilitate reliable assessment of heart rate variability, but not blood pressure variability or baroreflex function

BACKGROUND: We sought to determine whether heart rate variability (HRV), blood pressure (BP) variability, and baroreceptor-heart rate reflex sensitivity can be reliably assessed using finger volume pulse waveforms obtained from the commercially available EndoPAT device. METHODS: Non-invasive BP (Fin...

Descripción completa

Detalles Bibliográficos
Autores principales: Linder, Jonathan R, Stauss, Harald M, Gindes, Holly, Pierce, Gary L, Von Bergen, Nicholas H, Haynes, William G, Fiedorowicz, Jess G
Formato: Online Artículo Texto
Lenguaje:English
Publicado: BioMed Central 2014
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4269858/
https://www.ncbi.nlm.nih.gov/pubmed/25487432
http://dx.doi.org/10.1186/1471-2261-14-180
_version_ 1782349402216595456
author Linder, Jonathan R
Stauss, Harald M
Gindes, Holly
Pierce, Gary L
Von Bergen, Nicholas H
Haynes, William G
Fiedorowicz, Jess G
author_facet Linder, Jonathan R
Stauss, Harald M
Gindes, Holly
Pierce, Gary L
Von Bergen, Nicholas H
Haynes, William G
Fiedorowicz, Jess G
author_sort Linder, Jonathan R
collection PubMed
description BACKGROUND: We sought to determine whether heart rate variability (HRV), blood pressure (BP) variability, and baroreceptor-heart rate reflex sensitivity can be reliably assessed using finger volume pulse waveforms obtained from the commercially available EndoPAT device. METHODS: Non-invasive BP (Finometer Pro as a non-invasive standard) and finger volume (EndoPAT) waveforms were recorded in 65 adults (37 ± 14 years; 60% female) and systolic BP and heart rate (HR) time series were derived after calibrating the EndoPAT signal based on systolic and diastolic BP values obtained by a sphygomomanometer. Transfer function analyses were performed to test for coherence between systolic BP and HR time series derived from the Finometer and EndoPAT devices. Time-domain HRV parameters, frequency domain HR and systolic BP variability parameters, and baroreflex sensitivity (sequence technique) were computed from Finometer- and EndoPAT-derived time series and intraclass correlation coefficients (ICC) were calculated. RESULTS: Squared coherence between systolic BP time series derived from the Finometer and EndoPAT devices was low, suggesting poor correlation. In contrast, squared coherence between HR time series derived from the two devices was excellent [High Frequency (HF) = 0.80, Low Frequency (LF) = 0.81], with gain values close to 1.0. ICC values for time- and frequency-domain HRV parameters were excellent (>0.9 except for relative HF HRV, which was 0.77), while ICC values for frequency-domain BP variability parameters and baroreceptor-HR reflex sensitivity were low. CONCLUSIONS: Finger volume pulse waveforms can be used to reliably assess both time-domain and frequency-domain HR variability. However, frequency domain BP variability parameters cannot be reliably assessed from finger volume pulse waveforms using the simple calibration technique used in this study.
format Online
Article
Text
id pubmed-4269858
institution National Center for Biotechnology Information
language English
publishDate 2014
publisher BioMed Central
record_format MEDLINE/PubMed
spelling pubmed-42698582014-12-18 Finger volume pulse waveforms facilitate reliable assessment of heart rate variability, but not blood pressure variability or baroreflex function Linder, Jonathan R Stauss, Harald M Gindes, Holly Pierce, Gary L Von Bergen, Nicholas H Haynes, William G Fiedorowicz, Jess G BMC Cardiovasc Disord Research Article BACKGROUND: We sought to determine whether heart rate variability (HRV), blood pressure (BP) variability, and baroreceptor-heart rate reflex sensitivity can be reliably assessed using finger volume pulse waveforms obtained from the commercially available EndoPAT device. METHODS: Non-invasive BP (Finometer Pro as a non-invasive standard) and finger volume (EndoPAT) waveforms were recorded in 65 adults (37 ± 14 years; 60% female) and systolic BP and heart rate (HR) time series were derived after calibrating the EndoPAT signal based on systolic and diastolic BP values obtained by a sphygomomanometer. Transfer function analyses were performed to test for coherence between systolic BP and HR time series derived from the Finometer and EndoPAT devices. Time-domain HRV parameters, frequency domain HR and systolic BP variability parameters, and baroreflex sensitivity (sequence technique) were computed from Finometer- and EndoPAT-derived time series and intraclass correlation coefficients (ICC) were calculated. RESULTS: Squared coherence between systolic BP time series derived from the Finometer and EndoPAT devices was low, suggesting poor correlation. In contrast, squared coherence between HR time series derived from the two devices was excellent [High Frequency (HF) = 0.80, Low Frequency (LF) = 0.81], with gain values close to 1.0. ICC values for time- and frequency-domain HRV parameters were excellent (>0.9 except for relative HF HRV, which was 0.77), while ICC values for frequency-domain BP variability parameters and baroreceptor-HR reflex sensitivity were low. CONCLUSIONS: Finger volume pulse waveforms can be used to reliably assess both time-domain and frequency-domain HR variability. However, frequency domain BP variability parameters cannot be reliably assessed from finger volume pulse waveforms using the simple calibration technique used in this study. BioMed Central 2014-12-09 /pmc/articles/PMC4269858/ /pubmed/25487432 http://dx.doi.org/10.1186/1471-2261-14-180 Text en © Linder et al.; licensee BioMed Central Ltd. 2014 This article is published under license to BioMed Central Ltd. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly credited. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.
spellingShingle Research Article
Linder, Jonathan R
Stauss, Harald M
Gindes, Holly
Pierce, Gary L
Von Bergen, Nicholas H
Haynes, William G
Fiedorowicz, Jess G
Finger volume pulse waveforms facilitate reliable assessment of heart rate variability, but not blood pressure variability or baroreflex function
title Finger volume pulse waveforms facilitate reliable assessment of heart rate variability, but not blood pressure variability or baroreflex function
title_full Finger volume pulse waveforms facilitate reliable assessment of heart rate variability, but not blood pressure variability or baroreflex function
title_fullStr Finger volume pulse waveforms facilitate reliable assessment of heart rate variability, but not blood pressure variability or baroreflex function
title_full_unstemmed Finger volume pulse waveforms facilitate reliable assessment of heart rate variability, but not blood pressure variability or baroreflex function
title_short Finger volume pulse waveforms facilitate reliable assessment of heart rate variability, but not blood pressure variability or baroreflex function
title_sort finger volume pulse waveforms facilitate reliable assessment of heart rate variability, but not blood pressure variability or baroreflex function
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4269858/
https://www.ncbi.nlm.nih.gov/pubmed/25487432
http://dx.doi.org/10.1186/1471-2261-14-180
work_keys_str_mv AT linderjonathanr fingervolumepulsewaveformsfacilitatereliableassessmentofheartratevariabilitybutnotbloodpressurevariabilityorbaroreflexfunction
AT staussharaldm fingervolumepulsewaveformsfacilitatereliableassessmentofheartratevariabilitybutnotbloodpressurevariabilityorbaroreflexfunction
AT gindesholly fingervolumepulsewaveformsfacilitatereliableassessmentofheartratevariabilitybutnotbloodpressurevariabilityorbaroreflexfunction
AT piercegaryl fingervolumepulsewaveformsfacilitatereliableassessmentofheartratevariabilitybutnotbloodpressurevariabilityorbaroreflexfunction
AT vonbergennicholash fingervolumepulsewaveformsfacilitatereliableassessmentofheartratevariabilitybutnotbloodpressurevariabilityorbaroreflexfunction
AT hayneswilliamg fingervolumepulsewaveformsfacilitatereliableassessmentofheartratevariabilitybutnotbloodpressurevariabilityorbaroreflexfunction
AT fiedorowiczjessg fingervolumepulsewaveformsfacilitatereliableassessmentofheartratevariabilitybutnotbloodpressurevariabilityorbaroreflexfunction