Rectifying calibration error of Goldmann applanation tonometer is easy!

PURPOSE: Goldmann applanation tonometer (GAT) is the current Gold standard tonometer. However, its calibration error is common and can go unnoticed in clinics. Its company repair has limitations. The purpose of this report is to describe a self-taught technique of rectifying calibration error of GAT...

Descripción completa

Detalles Bibliográficos
Autores principales: Choudhari, Nikhil S, Moorthy, P Krishna, Tungikar, Vinod B, Kumar, Mohan, George, Ronnie, Rao, Harsha L, Senthil, Sirisha, Vijaya, Lingam, Garudadri, Chandra Sekhar
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Medknow Publications & Media Pvt Ltd 2014
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4290199/
https://www.ncbi.nlm.nih.gov/pubmed/25494251
http://dx.doi.org/10.4103/0301-4738.146761
_version_ 1782352220794126336
author Choudhari, Nikhil S
Moorthy, P Krishna
Tungikar, Vinod B
Kumar, Mohan
George, Ronnie
Rao, Harsha L
Senthil, Sirisha
Vijaya, Lingam
Garudadri, Chandra Sekhar
author_facet Choudhari, Nikhil S
Moorthy, P Krishna
Tungikar, Vinod B
Kumar, Mohan
George, Ronnie
Rao, Harsha L
Senthil, Sirisha
Vijaya, Lingam
Garudadri, Chandra Sekhar
author_sort Choudhari, Nikhil S
collection PubMed
description PURPOSE: Goldmann applanation tonometer (GAT) is the current Gold standard tonometer. However, its calibration error is common and can go unnoticed in clinics. Its company repair has limitations. The purpose of this report is to describe a self-taught technique of rectifying calibration error of GAT. MATERIALS AND METHODS: Twenty-nine slit-lamp-mounted Haag-Streit Goldmann tonometers (Model AT 900 C/M; Haag-Streit, Switzerland) were included in this cross-sectional interventional pilot study. The technique of rectification of calibration error of the tonometer involved cleaning and lubrication of the instrument followed by alignment of weights when lubrication alone didn’t suffice. We followed the South East Asia Glaucoma Interest Group's definition of calibration error tolerance (acceptable GAT calibration error within ±2, ±3 and ±4 mm Hg at the 0, 20 and 60-mm Hg testing levels, respectively). RESULTS: Twelve out of 29 (41.3%) GATs were out of calibration. The range of positive and negative calibration error at the clinically most important 20-mm Hg testing level was 0.5 to 20 mm Hg and -0.5 to -18 mm Hg, respectively. Cleaning and lubrication alone sufficed to rectify calibration error of 11 (91.6%) faulty instruments. Only one (8.3%) faulty GAT required alignment of the counter-weight. CONCLUSIONS: Rectification of calibration error of GAT is possible in-house. Cleaning and lubrication of GAT can be carried out even by eye care professionals and may suffice to rectify calibration error in the majority of faulty instruments. Such an exercise may drastically reduce the downtime of the Gold standard tonometer.
format Online
Article
Text
id pubmed-4290199
institution National Center for Biotechnology Information
language English
publishDate 2014
publisher Medknow Publications & Media Pvt Ltd
record_format MEDLINE/PubMed
spelling pubmed-42901992015-01-15 Rectifying calibration error of Goldmann applanation tonometer is easy! Choudhari, Nikhil S Moorthy, P Krishna Tungikar, Vinod B Kumar, Mohan George, Ronnie Rao, Harsha L Senthil, Sirisha Vijaya, Lingam Garudadri, Chandra Sekhar Indian J Ophthalmol Original Article PURPOSE: Goldmann applanation tonometer (GAT) is the current Gold standard tonometer. However, its calibration error is common and can go unnoticed in clinics. Its company repair has limitations. The purpose of this report is to describe a self-taught technique of rectifying calibration error of GAT. MATERIALS AND METHODS: Twenty-nine slit-lamp-mounted Haag-Streit Goldmann tonometers (Model AT 900 C/M; Haag-Streit, Switzerland) were included in this cross-sectional interventional pilot study. The technique of rectification of calibration error of the tonometer involved cleaning and lubrication of the instrument followed by alignment of weights when lubrication alone didn’t suffice. We followed the South East Asia Glaucoma Interest Group's definition of calibration error tolerance (acceptable GAT calibration error within ±2, ±3 and ±4 mm Hg at the 0, 20 and 60-mm Hg testing levels, respectively). RESULTS: Twelve out of 29 (41.3%) GATs were out of calibration. The range of positive and negative calibration error at the clinically most important 20-mm Hg testing level was 0.5 to 20 mm Hg and -0.5 to -18 mm Hg, respectively. Cleaning and lubrication alone sufficed to rectify calibration error of 11 (91.6%) faulty instruments. Only one (8.3%) faulty GAT required alignment of the counter-weight. CONCLUSIONS: Rectification of calibration error of GAT is possible in-house. Cleaning and lubrication of GAT can be carried out even by eye care professionals and may suffice to rectify calibration error in the majority of faulty instruments. Such an exercise may drastically reduce the downtime of the Gold standard tonometer. Medknow Publications & Media Pvt Ltd 2014-11 /pmc/articles/PMC4290199/ /pubmed/25494251 http://dx.doi.org/10.4103/0301-4738.146761 Text en Copyright: © Indian Journal of Ophthalmology http://creativecommons.org/licenses/by-nc-sa/3.0 This is an open-access article distributed under the terms of the Creative Commons Attribution-Noncommercial-Share Alike 3.0 Unported, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Original Article
Choudhari, Nikhil S
Moorthy, P Krishna
Tungikar, Vinod B
Kumar, Mohan
George, Ronnie
Rao, Harsha L
Senthil, Sirisha
Vijaya, Lingam
Garudadri, Chandra Sekhar
Rectifying calibration error of Goldmann applanation tonometer is easy!
title Rectifying calibration error of Goldmann applanation tonometer is easy!
title_full Rectifying calibration error of Goldmann applanation tonometer is easy!
title_fullStr Rectifying calibration error of Goldmann applanation tonometer is easy!
title_full_unstemmed Rectifying calibration error of Goldmann applanation tonometer is easy!
title_short Rectifying calibration error of Goldmann applanation tonometer is easy!
title_sort rectifying calibration error of goldmann applanation tonometer is easy!
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4290199/
https://www.ncbi.nlm.nih.gov/pubmed/25494251
http://dx.doi.org/10.4103/0301-4738.146761
work_keys_str_mv AT choudharinikhils rectifyingcalibrationerrorofgoldmannapplanationtonometeriseasy
AT moorthypkrishna rectifyingcalibrationerrorofgoldmannapplanationtonometeriseasy
AT tungikarvinodb rectifyingcalibrationerrorofgoldmannapplanationtonometeriseasy
AT kumarmohan rectifyingcalibrationerrorofgoldmannapplanationtonometeriseasy
AT georgeronnie rectifyingcalibrationerrorofgoldmannapplanationtonometeriseasy
AT raoharshal rectifyingcalibrationerrorofgoldmannapplanationtonometeriseasy
AT senthilsirisha rectifyingcalibrationerrorofgoldmannapplanationtonometeriseasy
AT vijayalingam rectifyingcalibrationerrorofgoldmannapplanationtonometeriseasy
AT garudadrichandrasekhar rectifyingcalibrationerrorofgoldmannapplanationtonometeriseasy