Rectifying calibration error of Goldmann applanation tonometer is easy!
PURPOSE: Goldmann applanation tonometer (GAT) is the current Gold standard tonometer. However, its calibration error is common and can go unnoticed in clinics. Its company repair has limitations. The purpose of this report is to describe a self-taught technique of rectifying calibration error of GAT...
Autores principales: | , , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Medknow Publications & Media Pvt Ltd
2014
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4290199/ https://www.ncbi.nlm.nih.gov/pubmed/25494251 http://dx.doi.org/10.4103/0301-4738.146761 |
_version_ | 1782352220794126336 |
---|---|
author | Choudhari, Nikhil S Moorthy, P Krishna Tungikar, Vinod B Kumar, Mohan George, Ronnie Rao, Harsha L Senthil, Sirisha Vijaya, Lingam Garudadri, Chandra Sekhar |
author_facet | Choudhari, Nikhil S Moorthy, P Krishna Tungikar, Vinod B Kumar, Mohan George, Ronnie Rao, Harsha L Senthil, Sirisha Vijaya, Lingam Garudadri, Chandra Sekhar |
author_sort | Choudhari, Nikhil S |
collection | PubMed |
description | PURPOSE: Goldmann applanation tonometer (GAT) is the current Gold standard tonometer. However, its calibration error is common and can go unnoticed in clinics. Its company repair has limitations. The purpose of this report is to describe a self-taught technique of rectifying calibration error of GAT. MATERIALS AND METHODS: Twenty-nine slit-lamp-mounted Haag-Streit Goldmann tonometers (Model AT 900 C/M; Haag-Streit, Switzerland) were included in this cross-sectional interventional pilot study. The technique of rectification of calibration error of the tonometer involved cleaning and lubrication of the instrument followed by alignment of weights when lubrication alone didn’t suffice. We followed the South East Asia Glaucoma Interest Group's definition of calibration error tolerance (acceptable GAT calibration error within ±2, ±3 and ±4 mm Hg at the 0, 20 and 60-mm Hg testing levels, respectively). RESULTS: Twelve out of 29 (41.3%) GATs were out of calibration. The range of positive and negative calibration error at the clinically most important 20-mm Hg testing level was 0.5 to 20 mm Hg and -0.5 to -18 mm Hg, respectively. Cleaning and lubrication alone sufficed to rectify calibration error of 11 (91.6%) faulty instruments. Only one (8.3%) faulty GAT required alignment of the counter-weight. CONCLUSIONS: Rectification of calibration error of GAT is possible in-house. Cleaning and lubrication of GAT can be carried out even by eye care professionals and may suffice to rectify calibration error in the majority of faulty instruments. Such an exercise may drastically reduce the downtime of the Gold standard tonometer. |
format | Online Article Text |
id | pubmed-4290199 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2014 |
publisher | Medknow Publications & Media Pvt Ltd |
record_format | MEDLINE/PubMed |
spelling | pubmed-42901992015-01-15 Rectifying calibration error of Goldmann applanation tonometer is easy! Choudhari, Nikhil S Moorthy, P Krishna Tungikar, Vinod B Kumar, Mohan George, Ronnie Rao, Harsha L Senthil, Sirisha Vijaya, Lingam Garudadri, Chandra Sekhar Indian J Ophthalmol Original Article PURPOSE: Goldmann applanation tonometer (GAT) is the current Gold standard tonometer. However, its calibration error is common and can go unnoticed in clinics. Its company repair has limitations. The purpose of this report is to describe a self-taught technique of rectifying calibration error of GAT. MATERIALS AND METHODS: Twenty-nine slit-lamp-mounted Haag-Streit Goldmann tonometers (Model AT 900 C/M; Haag-Streit, Switzerland) were included in this cross-sectional interventional pilot study. The technique of rectification of calibration error of the tonometer involved cleaning and lubrication of the instrument followed by alignment of weights when lubrication alone didn’t suffice. We followed the South East Asia Glaucoma Interest Group's definition of calibration error tolerance (acceptable GAT calibration error within ±2, ±3 and ±4 mm Hg at the 0, 20 and 60-mm Hg testing levels, respectively). RESULTS: Twelve out of 29 (41.3%) GATs were out of calibration. The range of positive and negative calibration error at the clinically most important 20-mm Hg testing level was 0.5 to 20 mm Hg and -0.5 to -18 mm Hg, respectively. Cleaning and lubrication alone sufficed to rectify calibration error of 11 (91.6%) faulty instruments. Only one (8.3%) faulty GAT required alignment of the counter-weight. CONCLUSIONS: Rectification of calibration error of GAT is possible in-house. Cleaning and lubrication of GAT can be carried out even by eye care professionals and may suffice to rectify calibration error in the majority of faulty instruments. Such an exercise may drastically reduce the downtime of the Gold standard tonometer. Medknow Publications & Media Pvt Ltd 2014-11 /pmc/articles/PMC4290199/ /pubmed/25494251 http://dx.doi.org/10.4103/0301-4738.146761 Text en Copyright: © Indian Journal of Ophthalmology http://creativecommons.org/licenses/by-nc-sa/3.0 This is an open-access article distributed under the terms of the Creative Commons Attribution-Noncommercial-Share Alike 3.0 Unported, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Original Article Choudhari, Nikhil S Moorthy, P Krishna Tungikar, Vinod B Kumar, Mohan George, Ronnie Rao, Harsha L Senthil, Sirisha Vijaya, Lingam Garudadri, Chandra Sekhar Rectifying calibration error of Goldmann applanation tonometer is easy! |
title | Rectifying calibration error of Goldmann applanation tonometer is easy! |
title_full | Rectifying calibration error of Goldmann applanation tonometer is easy! |
title_fullStr | Rectifying calibration error of Goldmann applanation tonometer is easy! |
title_full_unstemmed | Rectifying calibration error of Goldmann applanation tonometer is easy! |
title_short | Rectifying calibration error of Goldmann applanation tonometer is easy! |
title_sort | rectifying calibration error of goldmann applanation tonometer is easy! |
topic | Original Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4290199/ https://www.ncbi.nlm.nih.gov/pubmed/25494251 http://dx.doi.org/10.4103/0301-4738.146761 |
work_keys_str_mv | AT choudharinikhils rectifyingcalibrationerrorofgoldmannapplanationtonometeriseasy AT moorthypkrishna rectifyingcalibrationerrorofgoldmannapplanationtonometeriseasy AT tungikarvinodb rectifyingcalibrationerrorofgoldmannapplanationtonometeriseasy AT kumarmohan rectifyingcalibrationerrorofgoldmannapplanationtonometeriseasy AT georgeronnie rectifyingcalibrationerrorofgoldmannapplanationtonometeriseasy AT raoharshal rectifyingcalibrationerrorofgoldmannapplanationtonometeriseasy AT senthilsirisha rectifyingcalibrationerrorofgoldmannapplanationtonometeriseasy AT vijayalingam rectifyingcalibrationerrorofgoldmannapplanationtonometeriseasy AT garudadrichandrasekhar rectifyingcalibrationerrorofgoldmannapplanationtonometeriseasy |