Cargando…
Diagnostic accuracy of single baseline measurement of Elecsys Troponin T high-sensitive assay for diagnosis of acute myocardial infarction in emergency department: systematic review and meta-analysis
Objective To obtain summary estimates of the accuracy of a single baseline measurement of the Elecsys Troponin T high-sensitive assay (Roche Diagnostics) for the diagnosis of acute myocardial infarction in patients presenting to the emergency department. Design Systematic review and meta-analysis of...
Autores principales: | , , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
BMJ Publishing Group Ltd.
2015
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4300724/ https://www.ncbi.nlm.nih.gov/pubmed/25646632 http://dx.doi.org/10.1136/bmj.h15 |
_version_ | 1782353554785173504 |
---|---|
author | Zhelev, Zhivko Hyde, Christopher Youngman, Emily Rogers, Morwenna Fleming, Simon Slade, Toby Coelho, Helen Jones-Hughes, Tracey Nikolaou, Vasilis |
author_facet | Zhelev, Zhivko Hyde, Christopher Youngman, Emily Rogers, Morwenna Fleming, Simon Slade, Toby Coelho, Helen Jones-Hughes, Tracey Nikolaou, Vasilis |
author_sort | Zhelev, Zhivko |
collection | PubMed |
description | Objective To obtain summary estimates of the accuracy of a single baseline measurement of the Elecsys Troponin T high-sensitive assay (Roche Diagnostics) for the diagnosis of acute myocardial infarction in patients presenting to the emergency department. Design Systematic review and meta-analysis of diagnostic test accuracy studies. Data sources Medline, Embase, and other relevant electronic databases were searched for papers published between January 2006 and December 2013. Study selection Studies were included if they evaluated the diagnostic accuracy of a single baseline measurement of Elecsys Troponin T high-sensitive assay for the diagnosis of acute myocardial infarction in patients presenting to the emergency department with suspected acute coronary syndrome. Study appraisal and data synthesis The first author screened all titles and abstracts identified through the searches and selected all potentially relevant papers. The screening of the full texts, the data extraction, and the methodological quality assessment, using the adapted QUADAS-2 tool, were conducted independently by two reviewers with disagreements being resolved through discussion or arbitration. If appropriate, meta-analysis was conducted using the hierarchical bivariate model. Results Twenty three studies reported the performance of the evaluated assay at presentation. The results for 14 ng/L and 3-5 ng/L cut-off values were pooled separately. At 14 ng/L (20 papers), the summary sensitivity was 89.5% (95% confidence interval 86.3% to 92.1%) and the summary specificity was 77.1% (68.7% to 83.7%). At 3-5 ng/L (six papers), the summary sensitivity was 97.4% (94.9% to 98.7%) and the summary specificity was 42.4% (31.2% to 54.5%). This means that if 21 of 100 consecutive patients have the target condition (21%, the median prevalence across the studies), 2 (95% confidence interval 2 to 3) of 21 patients with acute myocardial infarction will be missed (false negatives) if 14 ng/L is used as a cut-off value and 18 (13 to 25) of 79 patients without acute myocardial infarction will test positive (false positives). If the 3-5 ng/L cut-off value is used, <1 (0 to 1) patient with acute myocardial infarction will be missed and 46 (36 to 54) patients without acute myocardial infarction will test positive. Conclusions The results indicate that a single baseline measurement of the Elecsys Troponin T high-sensitive assay could be used to rule out acute myocardial infarction if lower cut-off values such as 3 ng/L or 5 ng/L are used. However, this method should be part of a comprehensive triage strategy and may not be appropriate for patients who present less than three hours after symptom onset. Care must also be exercised because of the higher imprecision of the evaluated assay and the greater effect of lot-to-lot reagent variation at low troponin concentrations. Systematic review registration PROSPERO registration number CRD42013003926. |
format | Online Article Text |
id | pubmed-4300724 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2015 |
publisher | BMJ Publishing Group Ltd. |
record_format | MEDLINE/PubMed |
spelling | pubmed-43007242015-01-21 Diagnostic accuracy of single baseline measurement of Elecsys Troponin T high-sensitive assay for diagnosis of acute myocardial infarction in emergency department: systematic review and meta-analysis Zhelev, Zhivko Hyde, Christopher Youngman, Emily Rogers, Morwenna Fleming, Simon Slade, Toby Coelho, Helen Jones-Hughes, Tracey Nikolaou, Vasilis BMJ Research Objective To obtain summary estimates of the accuracy of a single baseline measurement of the Elecsys Troponin T high-sensitive assay (Roche Diagnostics) for the diagnosis of acute myocardial infarction in patients presenting to the emergency department. Design Systematic review and meta-analysis of diagnostic test accuracy studies. Data sources Medline, Embase, and other relevant electronic databases were searched for papers published between January 2006 and December 2013. Study selection Studies were included if they evaluated the diagnostic accuracy of a single baseline measurement of Elecsys Troponin T high-sensitive assay for the diagnosis of acute myocardial infarction in patients presenting to the emergency department with suspected acute coronary syndrome. Study appraisal and data synthesis The first author screened all titles and abstracts identified through the searches and selected all potentially relevant papers. The screening of the full texts, the data extraction, and the methodological quality assessment, using the adapted QUADAS-2 tool, were conducted independently by two reviewers with disagreements being resolved through discussion or arbitration. If appropriate, meta-analysis was conducted using the hierarchical bivariate model. Results Twenty three studies reported the performance of the evaluated assay at presentation. The results for 14 ng/L and 3-5 ng/L cut-off values were pooled separately. At 14 ng/L (20 papers), the summary sensitivity was 89.5% (95% confidence interval 86.3% to 92.1%) and the summary specificity was 77.1% (68.7% to 83.7%). At 3-5 ng/L (six papers), the summary sensitivity was 97.4% (94.9% to 98.7%) and the summary specificity was 42.4% (31.2% to 54.5%). This means that if 21 of 100 consecutive patients have the target condition (21%, the median prevalence across the studies), 2 (95% confidence interval 2 to 3) of 21 patients with acute myocardial infarction will be missed (false negatives) if 14 ng/L is used as a cut-off value and 18 (13 to 25) of 79 patients without acute myocardial infarction will test positive (false positives). If the 3-5 ng/L cut-off value is used, <1 (0 to 1) patient with acute myocardial infarction will be missed and 46 (36 to 54) patients without acute myocardial infarction will test positive. Conclusions The results indicate that a single baseline measurement of the Elecsys Troponin T high-sensitive assay could be used to rule out acute myocardial infarction if lower cut-off values such as 3 ng/L or 5 ng/L are used. However, this method should be part of a comprehensive triage strategy and may not be appropriate for patients who present less than three hours after symptom onset. Care must also be exercised because of the higher imprecision of the evaluated assay and the greater effect of lot-to-lot reagent variation at low troponin concentrations. Systematic review registration PROSPERO registration number CRD42013003926. BMJ Publishing Group Ltd. 2015-01-21 /pmc/articles/PMC4300724/ /pubmed/25646632 http://dx.doi.org/10.1136/bmj.h15 Text en © Zhelev et al 2015 http://creativecommons.org/licenses/by-nc/4.0/ This is an Open Access article distributed in accordance with the Creative Commons Attribution Non Commercial (CC BY-NC 4.0) license, which permits others to distribute, remix, adapt, build upon this work non-commercially, and license their derivative works on different terms, provided the original work is properly cited and the use is non-commercial. See: http://creativecommons.org/licenses/by-nc/4.0/. |
spellingShingle | Research Zhelev, Zhivko Hyde, Christopher Youngman, Emily Rogers, Morwenna Fleming, Simon Slade, Toby Coelho, Helen Jones-Hughes, Tracey Nikolaou, Vasilis Diagnostic accuracy of single baseline measurement of Elecsys Troponin T high-sensitive assay for diagnosis of acute myocardial infarction in emergency department: systematic review and meta-analysis |
title | Diagnostic accuracy of single baseline measurement of Elecsys Troponin T high-sensitive assay for diagnosis of acute myocardial infarction in emergency department: systematic review and meta-analysis |
title_full | Diagnostic accuracy of single baseline measurement of Elecsys Troponin T high-sensitive assay for diagnosis of acute myocardial infarction in emergency department: systematic review and meta-analysis |
title_fullStr | Diagnostic accuracy of single baseline measurement of Elecsys Troponin T high-sensitive assay for diagnosis of acute myocardial infarction in emergency department: systematic review and meta-analysis |
title_full_unstemmed | Diagnostic accuracy of single baseline measurement of Elecsys Troponin T high-sensitive assay for diagnosis of acute myocardial infarction in emergency department: systematic review and meta-analysis |
title_short | Diagnostic accuracy of single baseline measurement of Elecsys Troponin T high-sensitive assay for diagnosis of acute myocardial infarction in emergency department: systematic review and meta-analysis |
title_sort | diagnostic accuracy of single baseline measurement of elecsys troponin t high-sensitive assay for diagnosis of acute myocardial infarction in emergency department: systematic review and meta-analysis |
topic | Research |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4300724/ https://www.ncbi.nlm.nih.gov/pubmed/25646632 http://dx.doi.org/10.1136/bmj.h15 |
work_keys_str_mv | AT zhelevzhivko diagnosticaccuracyofsinglebaselinemeasurementofelecsystroponinthighsensitiveassayfordiagnosisofacutemyocardialinfarctioninemergencydepartmentsystematicreviewandmetaanalysis AT hydechristopher diagnosticaccuracyofsinglebaselinemeasurementofelecsystroponinthighsensitiveassayfordiagnosisofacutemyocardialinfarctioninemergencydepartmentsystematicreviewandmetaanalysis AT youngmanemily diagnosticaccuracyofsinglebaselinemeasurementofelecsystroponinthighsensitiveassayfordiagnosisofacutemyocardialinfarctioninemergencydepartmentsystematicreviewandmetaanalysis AT rogersmorwenna diagnosticaccuracyofsinglebaselinemeasurementofelecsystroponinthighsensitiveassayfordiagnosisofacutemyocardialinfarctioninemergencydepartmentsystematicreviewandmetaanalysis AT flemingsimon diagnosticaccuracyofsinglebaselinemeasurementofelecsystroponinthighsensitiveassayfordiagnosisofacutemyocardialinfarctioninemergencydepartmentsystematicreviewandmetaanalysis AT sladetoby diagnosticaccuracyofsinglebaselinemeasurementofelecsystroponinthighsensitiveassayfordiagnosisofacutemyocardialinfarctioninemergencydepartmentsystematicreviewandmetaanalysis AT coelhohelen diagnosticaccuracyofsinglebaselinemeasurementofelecsystroponinthighsensitiveassayfordiagnosisofacutemyocardialinfarctioninemergencydepartmentsystematicreviewandmetaanalysis AT joneshughestracey diagnosticaccuracyofsinglebaselinemeasurementofelecsystroponinthighsensitiveassayfordiagnosisofacutemyocardialinfarctioninemergencydepartmentsystematicreviewandmetaanalysis AT nikolaouvasilis diagnosticaccuracyofsinglebaselinemeasurementofelecsystroponinthighsensitiveassayfordiagnosisofacutemyocardialinfarctioninemergencydepartmentsystematicreviewandmetaanalysis |