Cargando…

Reliable and Accurate CD4+ T Cell Count and Percent by the Portable Flow Cytometer CyFlow MiniPOC and “CD4 Easy Count Kit-Dry”, as Revealed by the Comparison with the Gold Standard Dual Platform Technology

BACKGROUND: An accurate and affordable CD4+ T cells count is an essential tool in the fight against HIV/AIDS. Flow cytometry (FCM) is the “gold standard” for counting such cells, but this technique is expensive and requires sophisticated equipment, temperature-sensitive monoclonal antibodies (mAbs)...

Descripción completa

Detalles Bibliográficos
Autores principales: Nasi, Milena, De Biasi, Sara, Bianchini, Elena, Gibellini, Lara, Pinti, Marcello, Scacchetti, Tiziana, Trenti, Tommaso, Borghi, Vanni, Mussini, Cristina, Cossarizza, Andrea
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Public Library of Science 2015
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4306486/
https://www.ncbi.nlm.nih.gov/pubmed/25622041
http://dx.doi.org/10.1371/journal.pone.0116848
Descripción
Sumario:BACKGROUND: An accurate and affordable CD4+ T cells count is an essential tool in the fight against HIV/AIDS. Flow cytometry (FCM) is the “gold standard” for counting such cells, but this technique is expensive and requires sophisticated equipment, temperature-sensitive monoclonal antibodies (mAbs) and trained personnel. The lack of access to technical support and quality assurance programs thus limits the use of FCM in resource-constrained countries. We have tested the accuracy, the precision and the carry-over contamination of Partec CyFlow MiniPOC, a portable and economically affordable flow cytometer designed for CD4+ count and percentage, used along with the “CD4% Count Kit-Dry”. MATERIALS AND METHODS: Venous blood from 59 adult HIV+ patients (age: 25–58 years; 43 males and 16 females) was collected and stained with the “MiniPOC CD4% Count Kit-Dry”. CD4+ count and percentage were then determined in triplicate by the CyFlow MiniPOC. In parallel, CD4 count was performed using mAbs and a CyFlow Counter, or by a dual platform system (from Beckman Coulter) based upon Cytomic FC500 (“Cytostat tetrachrome kit” for mAbs) and Coulter HmX Hematology Analyzer (for absolute cell count). RESULTS: The accuracy of CyFlow MiniPOC against Cytomic FC500 showed a correlation coefficient (CC) of 0.98 and 0.97 for CD4+ count and percentage, respectively. The accuracy of CyFlow MiniPOC against CyFlow Counter showed a CC of 0.99 and 0.99 for CD4 T cell count and percentage, respectively. CyFlow MiniPOC showed an excellent repeatability: CD4+ cell count and percentage were analyzed on two instruments, with an intra-assay precision below ±5% deviation. Finally, there was no carry-over contamination for samples at all CD4 values, regardless of their position in the sequence of analysis. CONCLUSION: The cost-effective CyFlow MiniPOC produces rapid, reliable and accurate results that are fully comparable with those from highly expensive dual platform systems.