Quantifying precision and accuracy of measurements of dissolved inorganic carbon stable isotopic composition using continuous-flow isotope-ratio mass spectrometry

RATIONALE: We describe an analytical procedure that allows sample collection and measurement of carbon isotopic composition (δ(13)C(V-PDB) value) and dissolved inorganic carbon concentration, [DIC], in aqueous samples without further manipulation post field collection. By comparing outputs from two...

Descripción completa

Detalles Bibliográficos
Autores principales: Waldron, Susan, Marian Scott, E, Vihermaa, Leena E, Newton, Jason
Formato: Online Artículo Texto
Lenguaje:English
Publicado: BlackWell Publishing Ltd 2014
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4312894/
https://www.ncbi.nlm.nih.gov/pubmed/24711275
http://dx.doi.org/10.1002/rcm.6873
_version_ 1782355172539760640
author Waldron, Susan
Marian Scott, E
Vihermaa, Leena E
Newton, Jason
author_facet Waldron, Susan
Marian Scott, E
Vihermaa, Leena E
Newton, Jason
author_sort Waldron, Susan
collection PubMed
description RATIONALE: We describe an analytical procedure that allows sample collection and measurement of carbon isotopic composition (δ(13)C(V-PDB) value) and dissolved inorganic carbon concentration, [DIC], in aqueous samples without further manipulation post field collection. By comparing outputs from two different mass spectrometers, we quantify with the statistical rigour uncertainty associated with the estimation of an unknown measurement. This is rarely undertaken, but it is needed to understand the significance of field data and to interpret quality assurance exercises. METHODS: Immediate acidification of field samples during collection in evacuated, pre-acidified vials removed the need for toxic chemicals to inhibit continued bacterial activity that might compromise isotopic and concentration measurements. Aqueous standards mimicked the sample matrix and avoided headspace fractionation corrections. Samples were analysed using continuous-flow isotope-ratio mass spectrometry, but for low DIC concentration the mass spectrometer response could be non-linear. This had to be corrected for. RESULTS: Mass spectrometer non-linearity exists. Rather than estimating precision as the repeat analysis of an internal standard, we have adopted inverse linear calibrations to quantify the precision and 95% confidence intervals (CI) of the δ(13)C(DIC) values. The response for [DIC] estimation was always linear. For 0.05–0.5 mM DIC internal standards, however, changes in mass spectrometer linearity resulted in estimations of the precision in the δ(13)C(VPDB) value of an unknown ranging from ± 0.44‰ to ± 1.33‰ (mean values) and a mean 95% CI half-width of ±1.1–3.1‰. CONCLUSIONS: Mass spectrometer non-linearity should be considered in estimating uncertainty in measurement. Similarly, statistically robust estimates of precision and accuracy should also be adopted. Such estimations do not inhibit research advances: our consideration of small-scale spatial variability at two points on a small order river system demonstrates field data ranges larger than the precision and uncertainties. However, without such statistical quantification, exercises such as inter-lab calibrations are less meaningful.
format Online
Article
Text
id pubmed-4312894
institution National Center for Biotechnology Information
language English
publishDate 2014
publisher BlackWell Publishing Ltd
record_format MEDLINE/PubMed
spelling pubmed-43128942015-02-10 Quantifying precision and accuracy of measurements of dissolved inorganic carbon stable isotopic composition using continuous-flow isotope-ratio mass spectrometry Waldron, Susan Marian Scott, E Vihermaa, Leena E Newton, Jason Rapid Commun Mass Spectrom Research Article RATIONALE: We describe an analytical procedure that allows sample collection and measurement of carbon isotopic composition (δ(13)C(V-PDB) value) and dissolved inorganic carbon concentration, [DIC], in aqueous samples without further manipulation post field collection. By comparing outputs from two different mass spectrometers, we quantify with the statistical rigour uncertainty associated with the estimation of an unknown measurement. This is rarely undertaken, but it is needed to understand the significance of field data and to interpret quality assurance exercises. METHODS: Immediate acidification of field samples during collection in evacuated, pre-acidified vials removed the need for toxic chemicals to inhibit continued bacterial activity that might compromise isotopic and concentration measurements. Aqueous standards mimicked the sample matrix and avoided headspace fractionation corrections. Samples were analysed using continuous-flow isotope-ratio mass spectrometry, but for low DIC concentration the mass spectrometer response could be non-linear. This had to be corrected for. RESULTS: Mass spectrometer non-linearity exists. Rather than estimating precision as the repeat analysis of an internal standard, we have adopted inverse linear calibrations to quantify the precision and 95% confidence intervals (CI) of the δ(13)C(DIC) values. The response for [DIC] estimation was always linear. For 0.05–0.5 mM DIC internal standards, however, changes in mass spectrometer linearity resulted in estimations of the precision in the δ(13)C(VPDB) value of an unknown ranging from ± 0.44‰ to ± 1.33‰ (mean values) and a mean 95% CI half-width of ±1.1–3.1‰. CONCLUSIONS: Mass spectrometer non-linearity should be considered in estimating uncertainty in measurement. Similarly, statistically robust estimates of precision and accuracy should also be adopted. Such estimations do not inhibit research advances: our consideration of small-scale spatial variability at two points on a small order river system demonstrates field data ranges larger than the precision and uncertainties. However, without such statistical quantification, exercises such as inter-lab calibrations are less meaningful. BlackWell Publishing Ltd 2014-05-30 2014-04-01 /pmc/articles/PMC4312894/ /pubmed/24711275 http://dx.doi.org/10.1002/rcm.6873 Text en © 2014 The Authors. Rapid Communications in Mass Spectrometry Published by John Wiley & Sons Ltd. http://creativecommons.org/licenses/by/3.0/ This is an open access article under the terms of the Creative Commons Attribution License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Research Article
Waldron, Susan
Marian Scott, E
Vihermaa, Leena E
Newton, Jason
Quantifying precision and accuracy of measurements of dissolved inorganic carbon stable isotopic composition using continuous-flow isotope-ratio mass spectrometry
title Quantifying precision and accuracy of measurements of dissolved inorganic carbon stable isotopic composition using continuous-flow isotope-ratio mass spectrometry
title_full Quantifying precision and accuracy of measurements of dissolved inorganic carbon stable isotopic composition using continuous-flow isotope-ratio mass spectrometry
title_fullStr Quantifying precision and accuracy of measurements of dissolved inorganic carbon stable isotopic composition using continuous-flow isotope-ratio mass spectrometry
title_full_unstemmed Quantifying precision and accuracy of measurements of dissolved inorganic carbon stable isotopic composition using continuous-flow isotope-ratio mass spectrometry
title_short Quantifying precision and accuracy of measurements of dissolved inorganic carbon stable isotopic composition using continuous-flow isotope-ratio mass spectrometry
title_sort quantifying precision and accuracy of measurements of dissolved inorganic carbon stable isotopic composition using continuous-flow isotope-ratio mass spectrometry
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4312894/
https://www.ncbi.nlm.nih.gov/pubmed/24711275
http://dx.doi.org/10.1002/rcm.6873
work_keys_str_mv AT waldronsusan quantifyingprecisionandaccuracyofmeasurementsofdissolvedinorganiccarbonstableisotopiccompositionusingcontinuousflowisotoperatiomassspectrometry
AT marianscotte quantifyingprecisionandaccuracyofmeasurementsofdissolvedinorganiccarbonstableisotopiccompositionusingcontinuousflowisotoperatiomassspectrometry
AT vihermaaleenae quantifyingprecisionandaccuracyofmeasurementsofdissolvedinorganiccarbonstableisotopiccompositionusingcontinuousflowisotoperatiomassspectrometry
AT newtonjason quantifyingprecisionandaccuracyofmeasurementsofdissolvedinorganiccarbonstableisotopiccompositionusingcontinuousflowisotoperatiomassspectrometry